ARROW recallZ-2369-2026 · Arrow International, LLC · initiated Apr 21, 2026

Class I Open, Classified All recalls posted in June 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Arrow International, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On 4/21/2026, correction notices (Firm Event Number: EIF-000608, lidocaine and bupivacaine) were mailed to customers who were asked to do the following: Arrow International LLC is voluntarily recalling product due to the Huons Co. Ltd. recall of Lidocaine and Bupivacaine, which is included in impacted Arrow Kits & Sets. Medical facilities 1. Provide this correction notice to all persons who need to be aware within your organization or to any organization where the potentially affected product has been transferred. 2. Check your inventory for product within the scope of this Medical Device Notification and place a copy of this notice with all affected product. 3. Complete and return the response form virtually via: https://fca.myteleflex.com/en/recall/000608 Distributors: 1. Provide Medical Device Notification to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you as their distributor. 2. Check your inventory for affected product. Place a copy of this notification with all affected product and ensure that the notification is included with future distributions of affected product. 3. Complete and return the response form via link: https://fca.myteleflex.com/en/recall/000608 4. If you have further distributed product outside of your country, please notify Teleflex Customer Service by return email. If you have questions contact firm's Customer Service: Link: https://fca.myteleflex.com/en/recall/000608, Customer Service: Telephone: 1-866-396-2111, FAX: 1-855-419-8507 Email: [email protected] On 5/29/2026, additional amended notices (Firm Event Number: EIF-000608-01) were mailed to include additional affected devices. On 5/1/2026, additional notices (Firm Event Number: EIF-000609, sodium chloride) were mailed and emailed to customers who have 0.9% sodium chloride (10mL saline solution ampule) in kits and sets.

Recall details

Recall number
Z-2369-2026
Recall class
Class I (most serious)
Status
Open, Classified
Initiated by the company
Apr 21, 2026
Posted by the FDA
Jun 13, 2026
Recalling firm
Arrow International, LLC
Quantity
348,684
Product code
OFC
Distribution
US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.

Product description

Description/REF: PSI KIT: 7 FR X 4IN ANTIMICROBIAL/ASK-09701-UMH; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-09803-OSU4; PSI/CVC KIT: 9 FR/7 FR X 6IN (16CM)/ASK-09903-AMC1; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-CMC1; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-OSU4; PSI KIT: 9 FR X 4 1/8 IN (10CM)/ASK-09903-PCMH1; PSI KIT: 8.5 FR X 10 CM/CDC-09803-1A; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09903-ECL3; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-11242-PHP2; MULTI LUMEN/PSI KIT/ASK-11142-NSU; MULTI-LUMEN/PSI KIT/AK-21242-X; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-DMC; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-GMC1; MULTILUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-HF5; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-LMDT; 2L MAC KIT: 9 FR DISTAL X 4IN (10 CM)/ASK-21142-UCL1; 2-L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-UPMC2; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-SL1; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UMH; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UN1; MAC KIT: 2-L 9 FR DISTAL X 10 CM AGB/CDC-21142-XCN1A; MAC KIT: 2-L 9 FR DISTAL X 10 CM AGB/CDC-21142-XN1A; MAC KIT: 2-L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-XCN1A; MAC KIT: 2-L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-XN1A; 2L MAC KIT: 9 FR DISTAL X 4 IN/ASK-21142-CO1; 2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-FH3; 2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-OHSU3; MULTILUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-SG2; MULTILUMEN/MAC KIT: 9 FR X 4 IN (10 CM)/ASK-21142-UOI1; 2 LUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-WHC2; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-BID1; AGB MAC KIT: 2L 9 FR DISTAL/ASK-21242-BSWH2; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-DU1; MULTILUMEN/PSI KIT 9 FR X 11.5 CM/ASK-21242-PCMH1; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UCH2; AGB MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UHC2; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/ASK-29803-BSWH2; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-29803-HHC; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-29803-MCY1; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-29803-UPM1; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-29804-VCU; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-MCY1; PSI KIT: 9 FR/ASK-29903-MHI; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-29903-OHU4; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-29903-UCH2; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-UCL1; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-UHC4; PSI KIT: 9FR X 4IN (10 CM) ANTIMICROBIAL/ASK-29903-UPM1; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-29907-LGH; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/CDC-29803-1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/CDC-29903-1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/CDC-29903-X1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/CDC-29903-XCN1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/CDC-29903-XN1A; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-AHH; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/CDC-29803-XN1A; PSI ACCESS TRAY/ASK-04001-MM2; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-GMC; MULTI LUMEN/PSI KIT: 9 FR DISTAL X 4 1/2/ASK-21242-OSU3

Code information (models, lots, serial numbers)

Show the full code information (10k characters)

REF(Material)/UDI-DI/Lot(Batch): ASK-09701-UMH/10801902193275, 10801902139365, 10801902143874/33F24M0171, 33F25A0444, 33F25G0008, 33F25K0039, 33F25L0133, 33F26B0369, 33F23K0811; ASK-09803-OSU4/10801902212075, 10801902193275/33F24F0994, 33F24K0428, 33F25B0183, 33F25D0761; ASK-09903-AMC1/1080190219327, 10801902161830, 10801902136678/33F25A0437, 33F25E0024, 33F25G0690, 33F25L0142, 33F26C0117, 33F24B0612; ASK-09903-CMC1/10801902136678, 10801902193275/33F24D0050, 33F24D0965, 33F24F0130, 33F24J0321, 33F24J0627, 33F24L0585, 33F25A0110, 33F25C0176, 33F25C0588, 33F25E0212, 33F25E1020, 33F25F0324, 33F25G0753, 33F25L0208, 33F26A0396, 33F26A1255, 33F25E0716; ASK-09903-OSU4/10801902212006, 10801902193275/33F24G0022, 33F24L0298, 33F25D0762; ASK-09903-PCMH1/10801902193275, 20801902115205/33F25L0139, 33F26B0136, 33F24H0108; CDC-09803-1A/50801902116715, 10801902193275, 60801902116729/33F24D0892, 33F24F0766, 33F24L0274, 33F25B0147, 33F25C0497, 33F25C0498, 33F25D0553, 33F25E0647, 33F25F0209, 33F25H0574, 33F25L0121, 33F25M0410, 33F24D0504; ASK-09903-ECL3/10801902193275/33F25G0241, 33F25H0530, 33F25J0357, 33F26A0015; ASK-11242-PHP2/10801902193275, 10801902196436/33F24J0117, 33F24K0727, 33F25A0009, 33F25E0003, 33F25F0359, 33F25K0040, 33F25L0132, 33F26B0223, 33F26D0016, 33F24K0544; ASK-11142-NSU/10801902182545, 10801902193275/33F24E0326, 33F24K0354, 33F24L0244, 33F24L0300, 33F25B0189, 33F25C0228, 33F25D0467, 33F25E0591, 33F25G0272, 33F25G0405, 33F25H0629, 33F25J0363, 33F26A1572, 33F26B0816, 33F24L0840; AK-21242-X/10801902193275, 30801902122262, 80801902123127/33F24G0358, 33F24J0522, 33F24K0124, 33F24L0749, 33F25B0683, 33F25E0073, 33F25J0479, 33F25L0443, 33F26C0301, 33F24D0371; ASK-21142-DMC/10801902193275, 10801902218930, 10801902206685/33F25B0242, 33F25M0671, 33F26A0237, 33F26B0946, 33F24D0718; ASK-21142-GMC1/10801902193275, 10801902200492/33F25B0642, 33F25E0719, 33F25H0582, 33F25M0147, 33F26C0335, 33F24F0098; ASK-21142-HF5/10801902200492, 10801902193275, 10801902200492/33F24F0921, 33F24K0350, 33F24L0245, 33F25A0168, 33F25C0272, 33F25D0344, 33F25F0103, 33F25G0282, 33F25H0614, 33F25J0371, 33F26A1525, 33F26C0040, 33F24L0309; ASK-21142-LMDT/10801902193275, 10801902220728/33F25A0041, 33F25B0381, 33F25D0346, 33F25F0172, 33F25K0416, 33F25M0343, 33F26B0189, 33F26C0139, 33F25K0206; ASK-21142-UCL1/10801902193275/33F25E0541, 33F25G0391, 33F25J0102, 33F25J0373, 33F25M0235, 33F26A1375, 33F24H0098; ASK-21142-UPMC2/10801902159479, 10801902193275, 10801902225365/33F24F0879, 33F24J0161, 33F24L0310, 33F24L1023, 33F25C0276, 33F25E0780, 33F25G0348, 33F25J0209, 33F26A1371, 33F26A1547, 33R25G0348, 33F26C0490; ASK-21242-SL1/10801902193275/33F26A1347, 33F25E0314; ASK-21242-UMH/10801902129502, 10801902193275, 10801902156348/33F24E0637, 33F24F0895, 33F24L0317, 33F25B0197, 33F25C0280, 33F25D0478, 33F25F0178, 33F25H0564, 33F25L0106, 33F26A1517, 33F24F0618; ASK-21242-UN1/10801902193275, 10801902156348/33F24H0146, 33F25A0438, 33F25D0032, 33F25E0315, 33F25F0236, 33F25H0778, 33F25K0044, 33F26A0252, 33F26B0219, 33F26C0048, 33F25C0816; CDC-21142-XCN1A/10801902196498, 10801902193275/33F24E0226, 33F24E0242, 33F24E0254, 33F24F0250, 33F24G0203, 33F24H0356, 33F24H0391, 33F24J0220, 33F24L0078, 33F24M0018, 33F24M0054, 33F24M0206, 33F25A0130, 33F25A0385, 33F25B0068, 33F25B0340, 33F25C0076, 33F25C0510, 33F25D0697, 33F25D0733, 33F25E1092, 33F25E1101, 33F25F0435, 33F25G0027, 33F25H0334, 33F25H0814, 33F25J1006, 33F25K0310, 33F25L0090, 33F25L0277, 33F26A1615, 33F26A1622, 33F26A1623, 33F26A1645, 33F25B0528; CDC-21142-XN1A/10801902193275, 10801902196504, 50801902116821/33F25D0441, 33F25E0816, 33F25F0210, 33F25J0418, 33F26B0438, 33F26B0809, 33F26C0018, 33F24D0429; CDC-21242-XCN1A/10801902196511, 10801902193275, 10801902196511/33F24D0427, 33F24D0493, 33F24D0502, 33F24E0257, 33F24E0833, 33F24G0147, 33F24G0162, 33F24G0202, 33F24G0249, 33F24G0620, 33F24H0067, 33F24H0330, 33F24H0355, 33F24H0390, 33F24H0423, 33F24H0449, 33F24J0252, 33F24K0013, 33F24K0041, 33F24K0629, 33F24L0158.33F24M0056, 33F24M0066, 33F24M0091, 33F25A0132, 33F25A0373, 33F25B0028, 33F25B0097, 33F25B0135, 33F25C0217, 33F25C0529, 33F25C0692, 33F25D0690,33F25D0692, 33F25D0698, 33F25D0707, 33F25D0721, 33F25D0734, 33F25D0754, 33F25E0415, 33F25E1102, 33F25E1123, 33F25E1152, 33F25F0030, 33F25F0039, 33F25F0484, 33F25F0525, 33F25H0335, 33F25H0872, 33F25J0747, 33F25J0774, 33F25K0292, 33F25K0353, 33F25L0092, 33F25L0236, 33F25L0298, 33F26A1617, 33F26A1626, 33F26A1632, 33F26A1635, 33F26B0621, 33F26B0622, 33F26B1039, 33F26D0229, 33F26D0230, 33X25J0002, 33F24H0395; CDC-21242-XN1A/10801902193275, 10801902196528, 20801902123484/33F25A0378, 33F25B0029, 33F25B0098, 33F25B0341, 33F25C0420, 33F25C0530, 33F25E1093, 33F25E1124, 33F25H0286, 33F25H0362, 33F25J0775, 33F25J1040, 33F26A1618, 33F26A1627, 33F26C0453, 33F26C0881, 33X25J0001, 33F24F0996; ASK-21142-CO1/10801902214291, 10801902193275/33F24F0767, 33F24K0423, 33F25B0548, 33F25D0266, 33F25F0170, 33F25H0592, 33F25K0136, 33F26A1450, 33F26C0489, 33F24J0113; ASK-21142-FH3/10801902206685, 10801902193275/33F24E0615, 33F24F0988, 33F24K0455, 33F24L0308, 33F24L0896, 33F25A0244, 33F25B0549, 33F25E0539, 33F25F0171, 33F25G0254, 33F25H0607, 33F26A1574; ASK-21142-OHSU3/10801902193275, 10801902188899, 10801902129953/33F25E1194, 33F25G0692, 33F25K0041, 33F25K0422, 33F26B0907, 33F24F0049; ASK-21142-SG2/10801902193275/33F25B0550, 33F25C0274, 33F25E0540, 33F25F0174, 33F25H0533, 33F25J0208, 33F26B0198, 33F26B0786, 33F25G0695; ASK-21142-UOI1/10801902193275, 10801902158878, 10801902159479/33F24J0014, 33F25A0013, 33F25B0425, 33F25E0012, 33F25G0553, 33F25H0773, 33F25K0042, 33F25L0259, 33F26A0236, 33F26C0115, 33F24D0800; ASK-21142-WHC2/10801902193275, 10801902134964/33F25B0643, 33F25D0473, 33F25E0857, 33F25F0114, 33F25J0210, 33F26C0338, 33F24H0148; ASK-21242-BID1/10801902193275, 10801902219135, 10801902220339/33F24L1014, 33F25B0644, 33F25E0594, 33F25F0091, 33F25F0346, 33F25H0454, 33F25J0211, 33F26C0376, 33F24F0067; ASK-21242-BSWH2/10801902220339, 10801902193275, 10801902160987/33F24F0965, 33F24K0451, 33F24L0311, 33F25A0298,33F25B0645, 33F25D0347, 33F25E0542, 33F25E0858, 33F25G0206, 33F25H0515, 33F25J0212, 33F25K0094, 33F25M0326, 33F26A1445, 33F26C0152, 33F24D0321; ASK-21242-DU1/10801902160987, 10801902193275, 10801902118407/33F24D0797, 33F24F0859, 33F24J0149, 33F24L1085, 33F25B0440, 33F25C0277, 33F25D0400, 33F25E0859, 33F25H0557, 33F25J0213, 33F25K0095, 33F24E0539; ASK-21242-PCMH1/10801902160987, 10801902193275, 10801902118407/33F24D0797, 33F24F0859, 33F24J0149, 33F24L1085, 33F25B0440, 33F25C0277, 33F25D0400, 33F25E0859, 33F25H0557, 33F25J0213, 33F25K0095, 33F24E0539; ASK-21242-UCH2/10801902219036, 10801902193275/33F24F0756, 33F24L1051, 33F25C0343, 33F25D0474, 33F25F0177, 33F25H0586, 33F25J0375, 33F26A1495, 33F26C0180, 33F26C0502, 33F25D0224; ASK-21242-UHC2/10801902193275, 10801902159639, 10801902114065/33F25D0349, 33F25F0115, 33F25G0349, 33F25H0634, 33F25J0377, 33F26C0173, 33F24D0662; ASK-29803-BSWH2/10801902220346, 10801902193275/33F24F1006, 33F24K0461, 33F24L0407, 33F25B0472, 33F25B0648, 33F25C0344, 33F25F0092, 33F25F0142, 33F25H0514, 33F25J0382, 33F25M0186, 33F26A1428, 33F26B0834, 33F26A0014; ASK-29803-HHC/10801902193275, 10801902211443/33F25A0213, 33F25B0442, 33F25F0181, 33F25G0247, 33F25J0152, 33F26B0174, 33F25E0548; ASK-29803-MCY1/10801902193275/33F24H0203, 33F24K0386, 33F24L0409, 33F25B0649, 33F25E0720, 33F25G0200, 33F25E0549; ASK-29803-UPM1/10801902193275, 10801902225136/33F24L0227, 33F24L0876, 33F25C0284, 33F25E0782, 33F25G0392, 33F25K0097, 33F25L0117, 33F26A1584, 33F26C0510; ASK-29804-VCU/10801902193275, 10801902112986/33F25E0551, 33F25F0360, 33F26B0225, 33F24L0889; ASK-29903-MCY1/10801902193275, 10801902221039, 10801902217087/33F24L0322, 33F25C0348, 33F25D0481, 33F25F0182, 33F25G0199, 33F25J0264, 33F26A1571, 33F26B0869, 33F24D0590; ASK-29903-MHI/10801902217087, 10801902193275/33F24F0808, 33F24L0831, 33F25B0474, 33F25E0555, 33F25H0458, 33F25J0265, 33F26C0341, 33F25F0008; ASK-29903-OHU4/10801902193275/33F25E0449, 33F25C0289; ASK-29903-UCH2/10801902193275/33F24K0489, 33F24L0410, 33F24L1024, 33F25C0350, 33F25F0183, 33F25H0587, 33F26A1458, 33F25H0438; ASK-29903-UCL1/10801902193275, 10801902144512/33F25J0385, 33F26B0713, 33F25B0200; ASK-29903-UHC4/10801902193275, 10801902159622/33F25F0184, 33F25H0470, 33F25J0154, 33F26B0123, 33F26C0025, 33F24H0118; ASK-29903-UPM1/10801902193275, 10801902159530, 10801902225174/33F24L0326, 33F25B0555, 33F25E0674, 33F25F0185, 33F25G0263, 33F25G0347, 33F25J0217, 33F25M0113, 33F26B0178, 33F26C0511; ASK-29907-LGH/10801902193275/33F25A0008, 33F25D0033, 33F25G0684, 33F25L0260, 33F25G0459; CDC-29803-1A/50801902117019, 10801902193275/33F24F0800, 33F24J0177, 33F24L0119, 33F24L1142, 33F25B0678, 33F25D0509, 33F25E0750, 33F25G0277, 33F25G0388, 33F25J0181, 33F26A1385, 33F26B0435, 33F25E0796; CDC-29903-1A/60801902117030, 10801902193275/33F24F0742, 33F24K0372, 33F24L0121, 33F24L1146, 33F25A0289, 33F25B0531, 33F25F0086, 33F25G0177, 33F25G0319, 33F25H0575, 33F25J0253 ,33F25M0106, 33F26A1491, 33F25J0671; CDC-29903-X1A/10801902193275, 20801902117049, 10801902193190/33F25J0679, 33F25J0726, 33F25K0187, 33F26A1553, 33F26C0352, 36F25B0022, 36F25G0002, 36F25G0019, 36F25G0025, 36F25H0007, 33F24E0229; CDC-29903-XCN1A/10801902193190, 10801902193275/33F24F0025, 33F24F0160, 33F24G0207, 33F24G0248, 33F24H0075, 33F24H0432, 33F24J0254, 33F24K0042, 33F24K0692, 33F24M0076, 33F24M0135, 33F25A0133, 33F25B0020, 33F25B0069, 33F25B0119, 33F25B0318, 33F25C0512, 33F25D0691, 33F25D0699, 33F25D0723, 33F25D0755, 33F25E1125, 33F25F0040, 33F25F0402, 33F25G0488, 33F25J0656, 33F25J0777, 33F25K0312, 33F25L0012, 33F25L0315, 33F25M0045, 33F26A1629, 33F26B0527, 33F26B0926, 33F26C0073, 33F26C0086, 33F26C0100, 33F26C0550, 33F25A0314; CDC-29903-XN1A/10801902193275, 10801902193206/33F25B0532, 33F25D0511, 33F25E0701, 33F25E0894, 33F25H0423, 33F25H0623, 33F25J0326, 33F26A1475, 33F26B0734, 33F26C0019, 33F25C0701; ASK-29903-AHH/10801902193275/33F25B0553, 33F25F0143, 33F25G0384, 33F25H0489, 33F25M0239, 33F24L0166; CDC-29803-XN1A/10801902193275, 10801902193121, 60801902117030/33F25D0510, 33F25E0700, 33F25E0893, 33F25G0337, 33F25J0252, 33F25M0400, 33F26B0152, 33F26B0733, 33F26B0820, 33F26C0009, 33F26C0582, 33F24D0888; ASK-04001-MM2/10801902193275, 10801902160086, 10801902114881/33F25B0428, 33F26B0220, 33F24D0538; ASK-21142-GMC/10801902189674/33F24K0062, 33F24K0670, 33F24M0044, 33F25A0105; ASK-21242-OSU3/10801902211986/33F24F0720, 33F24K0338, 33F24L0312;

Devices named in this recall

52 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 98793).