VOCSN recallZ-2525-2026 · Ventec Life Systems, Inc. · initiated May 8, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ventec Life Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.

FDA root cause category: Vendor change control

What the company is doing

On 5/8/2026, recall notices were mailed to customers who were asked to do the following: 1) Immediately discontinue use of affected devices. 2) Inspect current stock, quarantine/segregate any unused affected products, remove affected devices from clinical service and segregate to prevent further use. 3) Use an unaffected ventilator for patient support. 4) If you have distributed or transferred affected devices to other facilities, caregivers, or end users, you must promptly notify them of this recall and provide a copy of this notification. 5) Confirm receipt of this notification letter by returning the enclosed Customer Response Form via email to [email protected] Upon receipt of the Customer Response Form, firm will arrange for repair of existing device(s). Return the affected product to React Health If you have any questions, please contact firm at 844-698-6276.

Recall details

Recall number
Z-2525-2026
Status
Open, Classified
Initiated by the company
May 8, 2026
Posted by the FDA
Jun 23, 2026
Recalling firm
Ventec Life Systems, Inc.
Quantity
1
Product code
CBK
FDA submissions
K162877
Distribution
US Nationwide distribution in the state of UT.

Product description

VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001

Code information (models, lots, serial numbers)

REF/UDI-DI/Serial Number/Lot: PRT-00853-000/00855573007839/155391P/807138, PRT-00490-001/00855573007792/155391/804613

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

22 devices cleared under K162877 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.