Skip to content

VOCSN recallZ-0131-2027 · Ventec Life Systems, Inc. · initiated Sep 17, 2026

Class I Open, Classified All recalls posted in October 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ventec Life Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

FDA root cause category: Under Investigation by firm

What the company is doing

On September 17, 2026, Ventec Life Systems Inc. dba React Health issued a Urgent: Medical Device Recall Notification via certified mail. Ventec ask consignees to take the following actions: 1.Review the customer communication and follow the applicable Actions to be Taken section. 2. Confirm receipt of this notification letter by returning the enclosed Customer Response Form to React Health via email, mail or use the online customer response form within 7 days. 3. If you have distributed or transferred affected devices to other facilities, caregivers, or end users, you must promptly notify them and provide a copy of this notification.

Recall details

Recall number
Z-0131-2027
Recall class
Class I (most serious)
Status
Open, Classified
Initiated by the company
Sep 17, 2026
Posted by the FDA
Oct 20, 2026
Recalling firm
Ventec Life Systems, Inc.
Quantity
180 units
Product code
CBK
FDA submissions
K162877
Distribution
Distribution US nationwide including Puerto Rico.

Product description

VC(Japan) Model Number: PRT-01098-100, Package PN:PRT-01125-100. used for ventilatory support.

Code information (models, lots, serial numbers)

All Serial Numbers: UDI: 00850018761116

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

22 devices cleared under K162877 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 99835).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page