Skip to content

Coloplast Titan OTR Inflatable Penile Prosthesis,18cm… recallZ-0171-2010 · Coloplast Manufacturing US, LLC · initiated Oct 6, 2009

Terminated All recalls posted in November 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Coloplast Manufacturing US, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Coloplast determined that a portion of Titan OTR pumps were manufactured using a process that where validation activities were not completed. The Titan OTR pumps were made to our standard product specifications and have passed all in-process testing to ensure mechanical reliability of these pumps. Coloplast is not recommending product explant because there are no identified risks since the prod

FDA root cause category: Process control

What the company is doing

Coloplast issued a customer notification letter dated October 2009 describing the affected product. For further information, contact Coloplast at 1-800-533-0464.

Recall details

Recall number
Z-0171-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 6, 2009
Posted by the FDA
Nov 10, 2009
Recalling firm
Coloplast Manufacturing US, LLC
Quantity
3
Product code
FHW
FDA submissions
P000006
Distribution
Nationwide Distribution -- FL, GA, NC.

Product description

Coloplast Titan OTR Inflatable Penile Prosthesis,18cm scrotal Bioflex cylinder set w/OTR Pump, P/N QSR9181400. Sterile EO. Manufacturer: Coloplast A/S, 3050 Denmark. Distributed by: Coloplast Corp., Minneapolis, MN 55411 USA. Indicated for male patients suffering from erectile dysfunction who are considered to be candidates for implantation of a penile prosthesis.

Code information (models, lots, serial numbers)

Lot 1919244.

Other devices under the same FDA submission

66 devices cleared under P000006 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates