Dexcom G7 Android App SW12299Recalls & reported problems

Dexcom, Inc. · G7 Android App · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This is the Dexcom G7 Android app. Its reports mostly describe the app shutting down unexpectedly, crashing or freezing, and alarms that do not work. There are 28,198 reports, with no deaths and 39 injuries. One recall, started in July 2025, is open: it covers older app versions. Reports rose to 24,423 in the last 12 months, from 3,212 before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
28,569
Deaths reported
0
Injuries reported
43
Recalls naming it
1

Summary

The FDA's MAUDE database contains 28,569 adverse event reports that list this exact device (DI 00386270001870), received between Dec 27, 2022 and Sep 30, 2026 — classified as 43 injuries, 28,524 malfunctions.

The most frequently reported specific device problem is “Protective Measures Problem” (19,435 reports).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K213919); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 12 reportsJul 2024: 23 reportsAug 2024: 48 reportsSep 2024: 137 reportsOct 2024: 320 reportsNov 2024: 198 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 255 reportsJun 2025: 527 reportsJul 2025: 551 reportsAug 2025: 441 reportsSep 2025: 616 reportsOct 2025: 620 reportsNov 2025: 569 reportsDec 2025: 968 reportsJan 2026: 19,118 reportsFeb 2026: 734 reportsMar 2026: 835 reportsApr 2026: 534 reportsMay 2026: 513 reportsJun 2026: 482 reportsJul 2026: 324 reportsAug 2026: 327 reportsSep 2026: 390 reportsOct 2026: 0 reports20242025202619,118
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 24, 2025Open, Classified

    The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the App. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a "transmitterFailed" error message to the App when the CGM experiences a hardware or firmware failure. Instead, the App ends the CGM sensor session, stops reporting glucose values and displays the Start Sensor screen or No active sensor message without alerting the user, which can lead to missed detection of a hyperglycemic or hypoglycemic event and a delay in treatment.

    Z-2446-2025 · Root cause: Software design · FDA record ↗

Most reported problems

Device problems

  • Protective Measures Problem19,435
  • Unintended Application Program Shut Down4,570
  • Operating System Version or Upgrade Problem1,478
  • Application Program Freezes or Fails to Launch909
  • Wireless Communication Problem808
  • Defective Alarm434
  • Display or Visual Feedback Problem416
  • No Audible Alarm394

Patient problems

  • No Clinical Signs, Symptoms or Conditions28,525
  • Hypoglycemia23
  • Loss of consciousness13
  • Hyperglycemia6
  • Hypoglycemic Shock6
  • Convulsion/Seizure5
  • Headache5
  • Shaking/Tremors5

Recent adverse event reports

The 25 most recent of 28,569 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Unintended Application Program Shut Down

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954762 ↗
  • MalfunctionSep 30, 2026Unintended Application Program Shut Down

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954756 ↗
  • MalfunctionSep 30, 2026Unintended Application Program Shut Down

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954643 ↗
  • MalfunctionSep 30, 2026Unintended Application Program Shut Down

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954642 ↗
  • MalfunctionSep 30, 2026Unintended Application Program Shut Down

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954634 ↗
  • MalfunctionSep 30, 2026Unintended Application Program Shut Down

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954631 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951951 ↗
  • MalfunctionSep 30, 2026Display or Visual Feedback Problem

    It was reported that an alert/notification settings occurred. Performance data was reviewed. The allegation was undetermined via data. However, it was found that EGVs not updating issue occurred within the investigation window. The probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951434 ↗
  • MalfunctionSep 30, 2026Display or Visual Feedback Problem

    It was reported that an alert/notification settings occurred. Performance data was reviewed. The allegation was undetermined via data. However, it was found that EGVs not updating issue occurred within the investigation window. The probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951135 ↗
  • MalfunctionSep 30, 2026Unintended Application Program Shut Down

    It was reported that an alert/notification settings occurred. Performance data was reviewed. The allegation was undetermined via data. However, it was found that an app crash occurred within the investigation window. The probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951131 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that signal loss occurred. It was determined that loss of connection was related to the Mobile Application. Performance data was reviewed. The allegation was confirmed due to the finding of signal loss over an hour within the investigation window. The probable cause was the transmitter and app were unable to establish a connection. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949447 ↗
  • MalfunctionSep 30, 2026Unintended Application Program Shut Down

    It was reported that an alert/notification settings occurred. Performance data was reviewed. The allegation was undetermined via data. However, it was found that an app crash occurred within the investigation window. The probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948111 ↗
  • MalfunctionSep 30, 2026Unintended Application Program Shut Down

    It was reported that an alert/notification settings occurred. Performance data was reviewed. The allegation was undetermined via data. However, it was found that an app crash occurred within the investigation window. The probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947308 ↗
  • MalfunctionSep 30, 2026Unintended Application Program Shut Down

    It was reported that an alert/notification settings occurred. Performance data was reviewed. The allegation was undetermined via data. However, it was found that an app crash occurred within the investigation window. The probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943599 ↗
  • MalfunctionSep 30, 2026Defective Alarm

    It was reported that no alert/notification occurred. Performance data was reviewed. The allegation was confirmed due to the finding of No Alert/Notification. The probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943598 ↗
  • MalfunctionSep 29, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941122 ↗
  • InjurySep 29, 2026No Audible Alarm

    Patient problems: Hypoglycemia

    MAUDE report 50940876 ↗
  • MalfunctionSep 29, 2026Unintended Application Program Shut Down

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940873 ↗
  • MalfunctionSep 29, 2026Unintended Application Program Shut Down

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940313 ↗
  • MalfunctionSep 29, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940303 ↗
  • MalfunctionSep 29, 2026Operating System Version or Upgrade Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50939742 ↗
  • MalfunctionSep 29, 2026Operating System Version or Upgrade Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50939740 ↗
  • MalfunctionSep 29, 2026Operating System Version or Upgrade Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50939732 ↗
  • MalfunctionSep 29, 2026Operating System Version or Upgrade Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50939022 ↗
  • MalfunctionSep 29, 2026Unintended Application Program Shut Down

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937255 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K213919. One submission often covers many models, so check the product and lot details in each record.

  • Z-2448-2025 · Jul 24, 2025 · Open, Classified
    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: versions 2.8.0 and earlier Produc
  • Z-2447-2025 · Jul 24, 2025 · Open, Classified
    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: SW12300 Software Version: version 2.8.0 and earlier Product Des
  • Z-1944-2025 · May 12, 2025 · Open, Classified
    Dexcom G7 Glucose Receiver, SKUs (part number): STK-AT-011(MT26403-0), STK-AT-012 (MT26403-0), STK-AT-013 (MT26403-0), STK-GT-001 (MT26403-1), STK-GT-008 (MT26403-5), STK-GT-013

Other Integrated Continuous Glucose Monitoring System, Factory Calibrated devices

All Integrated Continuous Glucose Monitoring System, Factory Calibrated devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.