Dexcom G7 recallZ-2446-2025 · Dexcom, Inc. · initiated Jul 24, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Dexcom, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the App. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a "transmitterFailed" error message to the App when the CGM experiences a hardware or firmware failure. Instead, the App ends the CGM sensor session, stops reporting glucose values and displays the Start Sensor screen or No active sensor message without alerting the user, which can lead to missed detection of a hyperglycemic or hypoglycemic event and a delay in treatment.
FDA root cause category: Software design
What the company is doing
On 07/24/2025, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION NOTIFICATION" via in-app messaging informing users that a software bug was identified with users using the following app version: -Dexcom ONE+ iOS/Android App version 1.4 and all earlier versions, -Dexcom G7 iOS/G7 Watch iOS/Android App versions 2.8 and earlier The software bug can cause an issue where, when a transmitter error occurs, the app terminates the sensor session and prompts the user to "start a new sensor" without providing a "sensor failed" alert. Customer are instructed that an upgrade from app version version the affected software versions is mandatory and users will not be able to to use the affected app versions after August 20, 2025. To update their app now to continue to use the app: 1. Tap Update App to go to the app store. 2. Install the latest Dexcom ONE+ app version or Dexcom G7 app version 3. Open the Dexcom ONE+ app or the G7 app If assistance is needed, contact Technical Support U.S.: 1-844-478-1600 Global: Find local Technical Support contact information at Dexcom.com
Recall details
- Recall number
- Z-2446-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 24, 2025
- Posted by the FDA
- Sep 4, 2025
- Recalling firm
- Dexcom, Inc.
- Quantity
- 532821
- Product code
- QBJ
- FDA submissions
- K213919
- Distribution
- SW12299 G7 Android CGM App Worldwide Distribution - US Nationwide and the countries of Andorra, United Arab Emirates, Austria, Australia, Belgium, Bahrain, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Hong Kong, Ireland, Israel, Iceland, Italy, Jordan, Japan, South Korea, Kuwait, Lebanon, Liechtenstein, Luxembourg, Malta, Netherlands, Norway, New Zealand, Oman, Poland, Portugal, Qatar, Saudi Arabia, Sweden, Slovenia, Slovakia, Turkey, South Africa SW12300 G7 iOS CGM App Worldwide Distribution - US Nationwide and the countries of United Arab Emirates, Austria, Australia, Belgium, Bahrain, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, Iceland, Italy, Jordan, Japan, South Korea, Kuwait, Lebanon, Liechtenstein, Luxembourg, Malta, Netherlands, Norway, New Zealand, Oman, Poland, Portugal, Qatar, Saudi Arabia, Sweden, Slovenia, Slovakia, Turkey, South Africa SW14245 Dexcom ONE+ Android Mobile App OUS worldwide distribution to the countries of Andorra, Australia, Belgium, Bulgaria, Estonia, Spain, France, United Kingdom, Greece, Croatia, Ireland, Israel, Italy, Lithuania, Latvia, Netherlands, New Zealand, Poland, Romania, Saudi Arabia, Sweden SW14244 D1G7 iOS Mobile App OUS worldwide distribution to the countries of Andorra, Australia, Belgium, Bulgaria, Estonia, Spain, France, United Kingdom, Greece, Croatia, Ireland, Italy, Lithuania, Latvia, Netherlands, New Zealand, Poland, Romania, Saudi Arabia, Sweden
Product description
Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App Model/Catalog Number: SW12299 Software Version: version 2.8.0 and earlier Product Description: Dexcom G7 CGM Android App SW12299 Component: Dexcom G7 Continuous Glucose Monitoring System
Code information (models, lots, serial numbers)
Lot Code: UDI: SW12299 - 00386270001870
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Dexcom G7 Android App SW12299Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall28,569 reports
Product lines
Other devices under the same FDA submission
17 devices cleared under K213919 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Dexcom G7 iOS App SW12300Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated3 recalls1 death55,219 reports
- Dexcom G7 Receiver STK-AT-012Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall3 deaths21,449 reports
- Dexcom G7 GSS 9500-151Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated12,714 reports
- Dexcom G7 Receiver STK-AT-013Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall9,330 reports
- Dexcom G7 Receiver STK-AT-011Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall2,939 reports
- Dexcom G7 GSS (155) STP-AT-012Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated406 reports
- Dexcom G7 iOS Watch App SW13355Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated2 recalls255 reports
- Dexcom G7 GSS (160) STP-AT-012Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated317 reports
- Dexcom G7 GSS (155) STP-AT-013Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated182 reports
- Dexcom G7 GSS (160) STP-AT-013Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated165 reports
- Dexcom G7 GSS STP-AT-011Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated65 reports
- Dexcom G7 GSS (160) STP-AT-011Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated51 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97262).
- Z-2447-2025 · Open, ClassifiedBrand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: SW12300 Software Version: version 2.8.0 and earlier Product Des…
- Z-2448-2025 · Open, ClassifiedBrand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: versions 2.8.0 and earlier Produc…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page