Dexcom G7 Receiver STK-AT-011Recalls & reported problems

Dexcom, Inc. · G7 US DME Receiver · Class 2

All models of Dexcom G7 (32) · Dexcom G7 Receiver (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Dexcom G7 receiver displays readings from the G7 monitoring system. Its 2,886 adverse event reports often describe unexpected resets and shutdowns, alarms that do not sound, and display problems. The reports include no deaths and 17 injuries. One open recall names this device, begun in May 2025, over a speaker that can lose contact with the circuit board and miss low or high blood sugar alerts. Reports fell to 573 over the last year, from 1,560 the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
2,939
Deaths reported
0
Injuries reported
17
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2,939 adverse event reports that list this exact device (DI 00386270002198), received between Mar 28, 2023 and Sep 30, 2026 — classified as 17 injuries, 2,922 malfunctions.

The most frequently reported specific device problem is “Inappropriate or Unexpected Reset” (928 reports).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K213919); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 41 reportsDec 2023: 31 reportsJan 2024: 43 reportsFeb 2024: 24 reportsMar 2024: 23 reportsApr 2024: 65 reportsMay 2024: 63 reportsJun 2024: 82 reportsJul 2024: 56 reportsAug 2024: 68 reportsSep 2024: 83 reportsOct 2024: 91 reportsNov 2024: 81 reportsDec 2024: 52 reportsJan 2025: 101 reportsFeb 2025: 57 reportsMar 2025: 68 reportsApr 2025: 40 reportsMay 2025: 87 reportsJun 2025: 76 reportsJul 2025: 183 reportsAug 2025: 439 reportsSep 2025: 270 reportsOct 2025: 106 reportsNov 2025: 71 reportsDec 2025: 71 reportsJan 2026: 61 reportsFeb 2026: 55 reportsMar 2026: 72 reportsApr 2026: 70 reportsMay 2026: 63 reportsJun 2026: 42 reportsJul 2026: 45 reportsAug 2026: 25 reportsSep 2026: 51 reportsOct 2026: 0 reports202420252026439
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 12, 2025Class IOpen, Classified

    Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.

    Z-1944-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Inappropriate or Unexpected Reset928
  • No Audible Alarm541
  • Unexpected Shutdown519
  • Defective Alarm472
  • Display or Visual Feedback Problem147
  • Wireless Communication Problem116
  • Unable to Obtain Readings103
  • Low Audible Alarm90

Patient problems

  • No Clinical Signs, Symptoms or Conditions2,920
  • Hypoglycemia12
  • Loss of consciousness9
  • Diaphoresis3
  • Hyperglycemia3
  • Confusion/ Disorientation2
  • Convulsion/Seizure2
  • Dizziness2

Recent adverse event reports

The 25 most recent of 2,939 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Inappropriate or Unexpected Reset

    It was reported that error icon display occurred frequently (b)(6) 2026. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950756 ↗
  • MalfunctionSep 30, 2026Unable to Obtain Readings

    It was reported that error icon display occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950745 ↗
  • MalfunctionSep 30, 2026Inappropriate or Unexpected Reset

    It was reported that a receiver reinitialization occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949818 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that a signal loss occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948083 ↗
  • MalfunctionSep 30, 2026Inappropriate or Unexpected Reset

    It was reported that a receiver reinitialization occurred frequently between (b)(6)2026. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947748 ↗
  • MalfunctionSep 29, 2026Inappropriate or Unexpected Reset

    It was reported that a receiver reinitialization occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937545 ↗
  • MalfunctionSep 29, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937270 ↗
  • MalfunctionSep 29, 2026Inappropriate or Unexpected Reset

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937258 ↗
  • MalfunctionSep 29, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935891 ↗
  • MalfunctionSep 28, 2026Wireless Communication Problem

    It was reported that a signal loss occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920874 ↗
  • MalfunctionSep 24, 2026Inappropriate or Unexpected Reset

    It was reported that a receiver reinitialization occurred occasionally between (b)(6)2026. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50885443 ↗
  • MalfunctionSep 23, 2026Unexpected Shutdown

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50873730 ↗
  • MalfunctionSep 23, 2026Inappropriate or Unexpected Reset

    It was reported that a receiver reinitialization occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50872273 ↗
  • MalfunctionSep 23, 2026Inappropriate or Unexpected Reset

    It was reported that a receiver reinitialization occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50872003 ↗
  • MalfunctionSep 22, 2026No Audible Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50860556 ↗
  • MalfunctionSep 22, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50858431 ↗
  • MalfunctionSep 22, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50857721 ↗
  • MalfunctionSep 21, 2026Inappropriate or Unexpected Reset

    It was reported that a receiver reinitialization occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50846623 ↗
  • MalfunctionSep 21, 2026Inappropriate or Unexpected Reset

    It was reported that a receiver reinitialization occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50846619 ↗
  • MalfunctionSep 18, 2026Inappropriate or Unexpected Reset

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50830901 ↗
  • MalfunctionSep 17, 2026Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812612 ↗
  • MalfunctionSep 17, 2026Inappropriate or Unexpected Reset

    It was reported that a receiver reinitialization occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50809070 ↗
  • MalfunctionSep 17, 2026Inappropriate or Unexpected Reset

    It was reported that a receiver reinitialization occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50805110 ↗
  • MalfunctionSep 15, 2026Defective Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50781647 ↗
  • MalfunctionSep 15, 2026Defective Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50781490 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K213919. One submission often covers many models, so check the product and lot details in each record.

  • Z-2448-2025 · Jul 24, 2025 · Open, Classified
    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: versions 2.8.0 and earlier Produc
  • Z-2446-2025 · Jul 24, 2025 · Open, Classified
    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App Model/Catalog Number: SW12299 Software Version: version 2.8.0 and earlier Product
  • Z-2447-2025 · Jul 24, 2025 · Open, Classified
    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: SW12300 Software Version: version 2.8.0 and earlier Product Des

Other Integrated Continuous Glucose Monitoring System, Factory Calibrated devices

All Integrated Continuous Glucose Monitoring System, Factory Calibrated devices →