Dexcom G7 Glucose Receiver, SKUs (part number):… recallZ-1944-2025 · Dexcom, Inc. · initiated May 12, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Dexcom, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On May 12, 2025, Dexcom issued a "Urgent: Medical Device Correction" Notification to affected consignees via E-Mail. On June 9, 2025, Dexcom issued an Urgent: Medical Device Correction update Dexcom asked consignees to take the following actions: 1. Verify if your receiver is affected by this issue: Check the serial number for any receiver(s) you use: https://dexcom.com/en-us/checkreceiver 2. If your receiver is affected, contact Dexcom Technical Support at 1-844-478-1600 to coordinate return of the affected receiver and a no charge replacement. 3. Whether or not your receiver is affected by this notification, Dexcom continues to recommend regular testing of the speaker. Test speaker function every time you charge your receiver. When you charge, a reminder prompt will invite you to test speaker function. 4.If you need assistance related to this issue or notification, please call Dexcom Technical Support at 1- 844-478-1600. Technical Support is available 24 hours a day/7 days a week to assist you.
Recall details
- Recall number
- Z-1944-2025
- Status
- Open, Classified
- Initiated by the company
- May 12, 2025
- Posted by the FDA
- Jun 16, 2025
- Recalling firm
- Dexcom, Inc.
- Quantity
- 602,445 units
- Product code
- QBJ
- FDA submissions
- K213919
- Distribution
- worldwide distribution - US Nationwide and the countries of Andorra, Argentina Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Greece, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Latvia, Liechtenstein, Lithuania, Luxembourg, Malta, Netherlands, New Zealand, Norway, Oman, Poland, Romania, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom.
Product description
Dexcom G7 Glucose Receiver, SKUs (part number): STK-AT-011(MT26403-0), STK-AT-012 (MT26403-0), STK-AT-013 (MT26403-0), STK-GT-001 (MT26403-1), STK-GT-008 (MT26403-5), STK-GT-013 (MT26403-1) , STK-GT-019 (MT26403-4) , STK-GT-100 (MT26403-2), STK-GT-109 (MT26403-2) , STK-GT-113 (MT26403-3)
Code information (models, lots, serial numbers)
All Lots/ UDI: STK-AT-011 - 00386270002198 STK-AT-012 -00386270002518 STK-AT-013 - 00386270002525 STK-GT-001 - 00386270004291 STK-GT-008 - 00386270004260 STK-GT-013 - 00386270002273 STK-GT-019 - 04595986293080 STK-GT-100 - 00386270002020 STK-GT-109 - 00386270002129 STK-GT-113 - 00386270003638
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- Dexcom G7 Receiver STK-AT-012Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall3 deaths21,449 reports
- Dexcom G7 Receiver STK-AT-013Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall9,330 reports
- Dexcom G7 Receiver STK-AT-011Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall2,939 reports
Product lines
Other devices under the same FDA submission
15 devices cleared under K213919 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Dexcom G7 iOS App SW12300Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated3 recalls1 death55,219 reports
- Dexcom G7 Android App SW12299Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall28,569 reports
- Dexcom G7 GSS 9500-151Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated12,714 reports
- Dexcom G7 GSS (155) STP-AT-012Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated406 reports
- Dexcom G7 iOS Watch App SW13355Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated2 recalls255 reports
- Dexcom G7 GSS (160) STP-AT-012Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated317 reports
- Dexcom G7 GSS (155) STP-AT-013Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated182 reports
- Dexcom G7 GSS (160) STP-AT-013Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated165 reports
- Dexcom G7 GSS STP-AT-011Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated65 reports
- Dexcom G7 GSS (160) STP-AT-011Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated51 reports
- Dexcom G7 Sample Pack (155) STE-AT-030Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated34 reports
- Dexcom G7 Receiver MT26403-0Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated14 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96743).
- Z-1943-2025 · Open, ClassifiedDexcom G6 Glucose Receiver, SKUs (Part Number): STK-FM-001 (MT27408-1) & STK-FR-001 (MT27408-1)
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.