Skip to content

Dexcom G7 Receiver STK-AT-012Recalls & reported problems

Dexcom, Inc. · G7 Receiver US · Class 2

All models of Dexcom G7 (32) · Dexcom G7 Receiver (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Dexcom G7 receiver has 20,882 adverse event reports in the FDA database, including 3 deaths and 114 injuries. Reports often describe the receiver resetting or shutting down unexpectedly, alarms that do not sound, and error icons on the screen. There is 1 open recall, started in May 2025, because the receiver speaker may lose contact with its circuit board, which can cause missed alerts for low or high glucose. Reports in the last 12 months (5,464) are well below the 12 months before (10,751).

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
21,449
Deaths reported
3
Injuries reported
114
Recalls naming it
1

Summary

The FDA's MAUDE database contains 21,449 adverse event reports that list this exact device (DI 00386270002518), received between Feb 23, 2023 and Sep 30, 2026 — classified as 3 deaths, 114 injuries, 21,332 malfunctions.

The most frequently reported specific device problem is “Inappropriate or Unexpected Reset” (6,515 reports).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K213919); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 95 reportsDec 2023: 159 reportsJan 2024: 210 reportsFeb 2024: 148 reportsMar 2024: 170 reportsApr 2024: 452 reportsMay 2024: 376 reportsJun 2024: 469 reportsJul 2024: 496 reportsAug 2024: 502 reportsSep 2024: 538 reportsOct 2024: 709 reportsNov 2024: 546 reportsDec 2024: 363 reportsJan 2025: 520 reportsFeb 2025: 307 reportsMar 2025: 517 reportsApr 2025: 267 reportsMay 2025: 661 reportsJun 2025: 942 reportsJul 2025: 1,498 reportsAug 2025: 2,942 reportsSep 2025: 1,255 reportsOct 2025: 933 reportsNov 2025: 721 reportsDec 2025: 610 reportsJan 2026: 676 reportsFeb 2026: 578 reportsMar 2026: 694 reportsApr 2026: 550 reportsMay 2026: 619 reportsJun 2026: 367 reportsJul 2026: 346 reportsAug 2026: 295 reportsSep 2026: 575 reportsOct 2026: 0 reports2024202520262,942
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 12, 2025Open, Classified

    Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.

    Z-1944-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Inappropriate or Unexpected Reset6,515
  • Unexpected Shutdown4,830
  • No Audible Alarm3,527
  • Defective Alarm2,912
  • Display or Visual Feedback Problem1,114
  • Unable to Obtain Readings979
  • Low Audible Alarm710
  • Wireless Communication Problem690

Patient problems

  • No Clinical Signs, Symptoms or Conditions21,319
  • Hypoglycemia73
  • Loss of consciousness58
  • Hyperglycemia31
  • Convulsion/Seizure22
  • Vomiting18
  • Confusion/ Disorientation14
  • Dizziness13

Recent adverse event reports

The 25 most recent of 21,449 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952266 ↗
  • MalfunctionSep 30, 2026Inappropriate or Unexpected Reset

    It was reported that a receiver reinitialization occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952043 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952002 ↗
  • MalfunctionSep 30, 2026Overheating of Device

    It was reported that the receiver was overheating. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951472 ↗
  • MalfunctionSep 30, 2026Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951182 ↗
  • MalfunctionSep 30, 2026Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951178 ↗
  • MalfunctionSep 30, 2026Unable to Obtain Readings

    It was reported that error icon display occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950779 ↗
  • MalfunctionSep 30, 2026Unable to Obtain Readings

    It was reported that error icon display occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950751 ↗
  • MalfunctionSep 30, 2026Unable to Obtain Readings

    It was reported that error icon display occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950743 ↗
  • MalfunctionSep 30, 2026Unable to Obtain Readings

    It was reported that error icon display occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950733 ↗
  • MalfunctionSep 30, 2026Unable to Obtain Readings

    It was reported that error icon display occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950370 ↗
  • MalfunctionSep 30, 2026Unable to Obtain Readings

    It was reported that error icon display occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950335 ↗
  • MalfunctionSep 30, 2026Display or Visual Feedback Problem

    It was reported that a display issue occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950288 ↗
  • MalfunctionSep 30, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949845 ↗
  • MalfunctionSep 30, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949797 ↗
  • MalfunctionSep 30, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949796 ↗
  • MalfunctionSep 30, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949490 ↗
  • MalfunctionSep 30, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949489 ↗
  • MalfunctionSep 30, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949475 ↗
  • MalfunctionSep 30, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949169 ↗
  • MalfunctionSep 30, 2026Unexpected Shutdown

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949163 ↗
  • MalfunctionSep 30, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949143 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that "Signal Loss" occurred. No product or data was provided for investigation. The allegation and the probable cause cannot be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949140 ↗
  • MalfunctionSep 30, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948487 ↗
  • MalfunctionSep 30, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948466 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K213919. One submission often covers many models, so check the product and lot details in each record.

  • Z-2447-2025 · Jul 24, 2025 · Open, Classified
    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: SW12300 Software Version: version 2.8.0 and earlier Product Des
  • Z-2448-2025 · Jul 24, 2025 · Open, Classified
    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: versions 2.8.0 and earlier Produc
  • Z-2446-2025 · Jul 24, 2025 · Open, Classified
    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App Model/Catalog Number: SW12299 Software Version: version 2.8.0 and earlier Product

Other Integrated Continuous Glucose Monitoring System, Factory Calibrated devices

All Integrated Continuous Glucose Monitoring System, Factory Calibrated devices →

Sources and updates