Dexcom G7 Receiver STK-AT-013Recalls & reported problems

Dexcom, Inc. · G7 Receiver US (Medicare/Medicaid) · Class 2

All models of Dexcom G7 (32) · Dexcom G7 Receiver (4)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Dexcom G7 receiver appears in 9,136 adverse event reports, including 29 injuries and no deaths. The reports mostly describe unexpected resets and shutdowns, missing audible alarms, and failures to obtain readings. One open recall, from May 2025, concerns a speaker that can lose contact with the circuit board, so low or high glucose alerts may not sound. Reports in the last 12 months number 2,392, down from 5,091 the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
9,131
Deaths reported
0
Injuries reported
28
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9,131 adverse event reports that list this exact device (DI 00386270002525), received between Mar 24, 2023 and Sep 30, 2026 — classified as 28 injuries, 9,103 malfunctions.

The most frequently reported specific device problem is “Inappropriate or Unexpected Reset” (2,752 reports).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K213919); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 44 reportsJan 2024: 64 reportsFeb 2024: 41 reportsMar 2024: 68 reportsApr 2024: 166 reportsMay 2024: 128 reportsJun 2024: 180 reportsJul 2024: 172 reportsAug 2024: 166 reportsSep 2024: 206 reportsOct 2024: 226 reportsNov 2024: 198 reportsDec 2024: 198 reportsJan 2025: 230 reportsFeb 2025: 157 reportsMar 2025: 208 reportsApr 2025: 131 reportsMay 2025: 243 reportsJun 2025: 295 reportsJul 2025: 650 reportsAug 2025: 1,930 reportsSep 2025: 530 reportsOct 2025: 321 reportsNov 2025: 284 reportsDec 2025: 328 reportsJan 2026: 297 reportsFeb 2026: 285 reportsMar 2026: 328 reportsApr 2026: 177 reportsMay 2026: 191 reportsJun 2026: 142 reportsJul 2026: 114 reportsAug 2026: 58 reportsSep 2026: 183 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261,930
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 12, 2025Class IOpen, Classified

    Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.

    Z-1944-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Inappropriate or Unexpected Reset2,752
  • Unexpected Shutdown1,904
  • Defective Alarm1,597
  • No Audible Alarm1,592
  • Display or Visual Feedback Problem387
  • Unable to Obtain Readings306
  • Low Audible Alarm265
  • Wireless Communication Problem239

Patient problems

  • No Clinical Signs, Symptoms or Conditions9,101
  • Hypoglycemia16
  • Loss of consciousness11
  • Shaking/Tremors6
  • Dizziness5
  • Hyperglycemia4
  • Hypoglycemic Shock4
  • Coma3

Recent adverse event reports

The 25 most recent of 9,131 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Unable to Obtain Readings

    It was reported that error icon display occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951176 ↗
  • MalfunctionSep 30, 2026Wireless Communication Problem

    It was reported that Signal Loss occurred. No product or data was provided for investigation. The allegation and the probable cause cannot be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950691 ↗
  • MalfunctionSep 30, 2026No Audible Alarm

    It was reported that an alert/notification settings issue occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950246 ↗
  • MalfunctionSep 30, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949800 ↗
  • MalfunctionSep 30, 2026Inappropriate or Unexpected Reset

    It was reported that a receiver reinitialization occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947717 ↗
  • MalfunctionSep 30, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945063 ↗
  • MalfunctionSep 30, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred frequently between (b)(6) 2026. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945060 ↗
  • MalfunctionSep 30, 2026Display or Visual Feedback Problem

    It was reported that a display issue occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943601 ↗
  • MalfunctionSep 29, 2026Display or Visual Feedback Problem

    It was reported that a display issue occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937875 ↗
  • MalfunctionSep 29, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937862 ↗
  • MalfunctionSep 29, 2026Low Audible Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937844 ↗
  • MalfunctionSep 29, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937586 ↗
  • MalfunctionSep 29, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937558 ↗
  • MalfunctionSep 29, 2026Inappropriate or Unexpected Reset

    It was reported that a receiver reinitialization occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937555 ↗
  • MalfunctionSep 29, 2026Unexpected Shutdown

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937253 ↗
  • MalfunctionSep 29, 2026Inappropriate or Unexpected Reset

    It was reported that a receiver reinitialization occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937220 ↗
  • MalfunctionSep 29, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936597 ↗
  • MalfunctionSep 29, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936592 ↗
  • MalfunctionSep 29, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936585 ↗
  • MalfunctionSep 29, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936250 ↗
  • MalfunctionSep 29, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936242 ↗
  • MalfunctionSep 29, 2026Inappropriate or Unexpected Reset

    It was reported that a receiver reinitialization occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935862 ↗
  • MalfunctionSep 29, 2026Inappropriate or Unexpected Reset

    It was reported that a receiver reinitialization occurred. Data was received but not investigated as data will not confirm the issue. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933950 ↗
  • MalfunctionSep 29, 2026Unexpected Shutdown

    It was reported that an unexpected receiver shutdown occurred. No product or data was provided for evaluation. The allegation and a probable cause could not be determined. No injury or medical intervention was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50932893 ↗
  • MalfunctionSep 28, 2026Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926177 ↗

FDA clearance and classification

Cleared through 510(k) K213919 on Dec 7, 2022 (decision: Substantially Equivalent). Submitted by Dexcom, Inc. as “Dexcom G7 Continuous Glucose Monitoring System”. The FDA review took 357 days. FDA record ↗

The FDA regulates product code QBJ (integrated continuous glucose monitoring system, factory calibrated) under 21 CFR 862.1355 as a Class II device. 21 CFR 862.1355 ↗

FDA definition of this device type: An integrated continuous glucose monitoring system (iCGM) is intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. iCGM systems are designed to reliably and securely transmit glucose measurement data to digitally connected devices, including automated insulin dosing systems, and are intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control.

Recalls under the same FDA submission

These recalls involve devices cleared under K213919. One submission often covers many models, so check the product and lot details in each record.

  • Z-2447-2025 · Jul 24, 2025 · Open, Classified
    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: SW12300 Software Version: version 2.8.0 and earlier Product Des
  • Z-2446-2025 · Jul 24, 2025 · Open, Classified
    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App Model/Catalog Number: SW12299 Software Version: version 2.8.0 and earlier Product
  • Z-2448-2025 · Jul 24, 2025 · Open, Classified
    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: versions 2.8.0 and earlier Produc

Common questions about the Dexcom G7 Receiver STK-AT-013

Has the Dexcom G7 Receiver STK-AT-013 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1944-2025 (Class I), began on May 12, 2025: Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Dexcom G7 Receiver STK-AT-013?

The FDA's MAUDE database lists 9,131 adverse event reports for this device, received between Mar 24, 2023 and Sep 30, 2026. In those reports, the most common specific device problems are “Inappropriate or Unexpected Reset” (2,752), “Unexpected Shutdown” (1,904) and “Defective Alarm” (1,597); the most common patient problems are “Hypoglycemia” (16), “Loss of consciousness” (11) and “Shaking/Tremors” (6). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Dexcom G7 Receiver STK-AT-013?

None of the reports are classified as deaths, 28 as injuries and 9,103 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Dexcom G7 Receiver STK-AT-013 FDA approved?

It was cleared through 510(k) K213919 on Dec 7, 2022. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Dexcom G7 Receiver STK-AT-013, and is it still sold?

The FDA's device database lists Dexcom, Inc. as the company responsible for it (brand Dexcom G7 Receiver, model STK-AT-013). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Integrated Continuous Glucose Monitoring System, Factory Calibrated devices

All Integrated Continuous Glucose Monitoring System, Factory Calibrated devices →