VOCSN recallZ-1902-2021 · Ventec Life Systems, Inc. · initiated May 10, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ventec Life Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to a component failure, ventilators may unexpectedly shut down or have a loss of blower function while in patient use. The shutdown potentially prevents the ventilator from providing ventilation therapy.
FDA root cause category: Device Design
What the company is doing
During the week of May 10, 2021, the firm sent an URGENT: MEDICAL DEVICE RECALL NOTIFICATION letter via FedEx or overnight mail to customers informing them that due to malfunctions of the Control Board there is the potential of unexpected shut downs and/or lost of ventilator blower function. As a result, the ventilators may fail to provide ventilation therapy to the patients. Customers are instructed to complete the following clinical action until a replacement unit can be provided: Follow your procedures to transfer the patient to an alternative source of ventilation, if available. If no other ventilation method is available, you may continue to use the affected product adhering to the guidelines below: 1. Use only when the patient is actively monitored to ensure timely response to a potential malfunction. 2. Operate the ventilator as defined in the Instructions for Use and promptly respond to all alarms activated by the ventilator. 3. Use an external SpO2 monitor and set the alarm thresholds appropriately if the patient is oxygen dependent. 4. Connect the ventilator to an external monitoring system, wherever possible. In addition, Customers are to acknowledge receipt of the Customer Notification Letter by: 1. Confirm receipt of this notification letter by returning the enclosed Customer Response Form to Ventec within 7 days. 2. Upon receipt of the Customer Response Form, Ventec will arrange shipment of replacement device(s) or arrange repair of existing device(s). 3. Return the affected product to Ventec per instructions and shipping label provided with replacement device(s). Contact Ventec Customer Support at 844-MYVOCSN (844-698-6276) to discuss alternative arrangements, if necessary. If customers are no longer the owner of any of these devices, they need to complete an ownership transfer form at VentecLife.com/tracking so that the Recalling Firm can follow up with the current owner of the impacted device(s). If customers have any question
Recall details
- Recall number
- Z-1902-2021
- Status
- Open, Classified
- Initiated by the company
- May 10, 2021
- Posted by the FDA
- Jun 21, 2021
- Recalling firm
- Ventec Life Systems, Inc.
- Quantity
- 1,934 devices (415 device that have not been updated)
- Product code
- CBK
- FDA submissions
- K162877
- Distribution
- Distribution US nationwide, Hong Kong, and Japan
Product description
Ventec Life Systems VOCSN Multi-Function Ventilator Model/Version/Catalog Number: V+O+C+S+N+Pro/English/PRT-00490-001; V+O+C+S+N+Pro/Japan/PRT-00490-100; V+C+Pro/English/PRT-00900-001; V+C+Pro/Japan/PRT-00900-100; V+C/English/PRT-01098-000; V+O+C+S+N/English/PRT-01100-000; V+O+C+S+N/English/PRT-01100-001;
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
Catalog Number: PRT-00490-001 GTIN (Primary ID): 00855573007792/00855573007006 Serial Numbers: 114303; 114304; 113728; 113729; 113697; 114296; 113587; 113684; 113706; 113003; 112957; 112961; 112962; 112965; 112972; 112974; 112994; 112995; 112999; 113006; 113008; 113010; 113520; 113522; 113525; 113533; 113534; 113536; 113547; 113550; 113555; 113556; 113557; 113561; 113563; 113689; 113724; 114021; 114101; 114105; 114106; 114107; 114109; 114110; 114111; 114116; 114117; 114310; 114333; 114108; 112792; 112798; 114497; 114324; 114325; 114330; 114336; 114493; 114537; 113046; 113662; 113667; 113682; 113591; 113594; 113654; 113655; 113740; 114020; 114030; 114044; 114046; 114048; 114096; 114114; 114019; 114025; 114027; 114033; 114039; 114040; 114045; 114051; 114028; 113704; 113746; 113748; 113753; 113760; 113061; 113725; 113586; 113691; 114307; 114309; 113028; 113066; 113539; 113515; 113709; 114047; 112814; 112824; 112829; 112830; 112844; 112848; 112865; 112866; 112807; 112822; 112823; 112834; 112835; 112847; 112854; 112876; 122848; 113758; 114508; 114951; 114342; 114343; 114346; 114351; 114360; 114500; 114345; 114499; 114505; 112976; 113037; 113040; 113524; 114322; 114498; 114510; 114636; 114640; 114641; 114676; 114685; 114695; 114701; 114820; 114827; 114841; 114900; 114901; 114674; 114725; 114800; 114810; 114812; 114822; 114825; 113057; 113063; 112821; 112839; 112842; 112867; 112826; 112837; 112825; 112982; 112987; 113510; 113511; 113546; 113584; 113592; 113652; 113686; 113710; 113723; 113756; 113671; 113675; 113715; 114009; 114013; 114489; 114494; 113549; 113744; 114637; 114947; 112765; 112766; 112767; 112768; 112770; 112774; 112775; 112776; 112778; 112779; 112780; 112787; 112815; 112841; 112862; 112968; 112977; 113013; 113014; 113016; 113017; 113018; 113027; 113032; 113039; 113050; 113535; 113562; 113588; 113658; 113755; 114032; 114488; 114842; AARC0001; AARC002; AARC003; AARC004; AARC005; AARC006; AARC007; AARC008; AARC009; 114095; 113590; Catalog Number: PRT-00490-100 GTIN (Primary ID): 00855573007761/00855573007204 Serial Numbers: 112929; 113598; 113233; 113238; 113239; 113241; 113243; 113244; 113245; 113603; 114193; 114201; 114255; 114266; Catalog Number: PRT-00900-001 GTIN (Primary ID): 00855573007778/00855573007433 Serial Numbers: 113365; 113405; 113906; 113962; 113407; 113414; 113430; 113390; 113437; 113262; 113380; 113458; 113384; 113398; 113442; 113858; 113441; 113443; 113445; 113915; 113817; 113898; 113899; 113378; 113395; 113923; 113972; 113937; 113945; 113957; 113961; 113968; 113368; 113373; 113379; 113264; 113351; 113389; 113370; 113377; 113454; 113468; 113804; 113823; 113830; 113837; 113838; 113881; 113882; 113888; 113892; 113902; 113927; 113967; 113399; Catalog Number: PRT-00900-100 GTIN (Primary ID): 00855573007785 Serial Numbers: 113096; 113300; 112923; 113107; 112877; 110392; 113143; 113183; 113274; 113323; Catalog Number: PRT-01098-000 GTIN (Primary ID): 00855573007730 Serial Numbers: 114444; 114451; 114615; 114226; 114442; 114450; 114452; 114441; 114453; Catalog Number: PRT-01100-000 GTIN (Primary ID): 00855573007747 Serial Numbers: 114556; 114591; 114634; 114771; 114779; 114603; 114606; 114667; 114710; 114713; 114737; 114739; 114559; 114783; 114869; 114576; 114587; 114364; 114369; 114374; 114387; 114391; 114560; 114568; 114764; 114776; 114787; 114849; 114549; 114569; 114705; 114562; 114355; 114372; 114405; 114550; 114594; 114601; 114652; 114659; 114661; 114662; 114358; 114383; 114386; 114389; 114390; 114530; 114540; 114401; 114422; 114424; 114548; 114670; 114714; 114850; 114414; 114417; 114554; 114561; 114861; 114862; 114543; 114786; 114789; 114912; 114917; 114851; 114628; 114211; 114542; 114528; 114531; 114483; 114750; 114643; 114644; 114748; 114926; 114991; 114891; 114892; 114894; 114357; 114409; 114604; 114413; 114653; 114654; 114734; 114754; Catalog Number: PRT-01100-001 GTIN (Primary ID): 00855573007747 Serial Numbers: 114965; 114966; 114967;
Devices named in this recall
6 devices in the FDA device registry whose identifier appears in the recall's code information.
- VOCSN V+O+C+S+N+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit3 recalls366 reports
- VOCSN V+O+C+S+N, EnglishVentec Life Systems, Inc. · Ventilator, Continuous, Facility Use1 recall286 reports
- VOCSN V+C+Pro, JapaneseVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 recall33 reports
- VOCSN V+O+C+S+N+Pro, JapaneseVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 recall32 reports
- VOCSN V+C, EnglishVentec Life Systems, Inc. · Ventilator, Continuous, Facility Use1 recall27 reports
- VOCSN V+C+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 recall25 reports
Product lines
Other devices under the same FDA submission
17 devices cleared under K162877 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- VOCSN V+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 recall570 reports
- VOCSN V+Pro Emergency, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit558 reports
- VOCSN V+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit2 recalls2 deaths325 reports
- VOCSN V*Home, EnglishVentec Life Systems, Inc. · Heater, Breathing System W/Wo Controller (Not Humidifier Or Nebulizer4 deaths318 reports
- VOCSN V+C+S+N+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit185 reports
- VOCSN VOCSN, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 death96 reports
- VOCSN Patient Breathing Package Pediatric, Active, Oxygen, BlueVentec Life Systems, Inc. · Device, Positive Pressure Breathing, Intermittent1 recall25 reports
- VOCSN V+C+Pro, JapaneseVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit41 reports
- VOCSN V+C, JapaneseVentec Life Systems, Inc. · Ventilator, Continuous, Facility Use30 reports
- VOCSN-VC VOCSN-VC Pro, EnglishVentec Life Systems, Inc. · Ventilator, Continuous, Facility Use21 reports
- VOCSN Patient Breathing Package Adult, Active, Heated, Oxygen, BlueVentec Life Systems, Inc. · Heater, Breathing System W/Wo Controller (Not Humidifier Or Nebulizer19 reports
- VOCSN V+O+C+S+N, JapaneseVentec Life Systems, Inc. · Ventilator, Continuous, Facility Use14 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗