pro+ P7924ARENT01Recalls & reported problems
Hill-Rom, Inc. · PRO+ 36" MRS RENTAL SURFACE WITH X-RAY · Class 2
All models of pro+ (18)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 15
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00887761976863), received between Dec 19, 2023 and Aug 11, 2026 — classified as 1 death, 14 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (14 reports).
3 FDA recall records name this device's identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- May 8, 2024Class IIOpen, Classified
There is a potential for the top cover of the hospital bed mattress to delaminate as a result of improper cleaning and lack of adequate details on the appropriate use of cleaning agents containing bleach.
Z-2151-2024 · Root cause: Error in labeling · Recall details · FDA record ↗
- Mar 31, 2023Class IIOpen, Classified
There is a potential for mattress cover damage (delamination) which may result in fluid absorption in the mattress (fluid ingress).
Z-1482-2023 · Root cause: Process change control · Recall details · FDA record ↗
- Jun 24, 2022Class IIOpen, Classified
The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of pressure ulcer development.
Z-1411-2022 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Split, Cut or Torn14
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions14
- Unspecified Tissue Injury1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 11, 2026Material Split, Cut or Torn
Baxter received a report that FLAT DECK 36" MRS W/ X-RAY SLV had power cord damaged with copper wire exposed. There was no patient/user injury, medical intervention, symptom, or adverse event reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50204258 ↗DeathMay 20, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THERE WAS NO AIR RUNNING THROUGH THE PRO+ AND THE PATIENTS¿ PRESSURE WOUNDS WERE GETTING WORSE. THE CUSTOMER REPORTED THE LOCATION OF THE PRESSURE SORES WERE ALONG HIS SPINE FROM HIS MID-UPPER BACK DOWN TO HIS COCCYX AND STATED THEY WERE UNSTAGEABLE DEEP TISSUE. THE CUSTOMER WAS UNABLE TO PROVIDE THE STAGING OF THE PRESSURE WOUNDS PRIOR TO THEM GETTING WORSE. THE CUSTOMER REPORTED THAT A TURN SCHEDULE WAS IMPLEMENTED AND THEN THEY STARTED TO USE A DRESSING, CUSTOMER DID NOT REMEMBER THE NAME BUT STATED "IT WAS LIKE AQUACEL." THE PATIENT WAS TRANSFERRED TO A HIGHER LEVEL OF CARE, AFTER THIS THE PATIENT PASSED AWAY (DATE UNKNOWN). THE CUSTOMER NOTED ON ADMISSION THE PATIENT WAS "EXTREMELY DECLINED AND WAS UNABLE TO GET OUT OF BED". IT IS UNKNOWN WHEN THE PATIENT DIED OR THE CAUSE OF DEATH. IT WAS NOT REPORTED IF AN AUTOPSY WAS PERFORMED. A BAXTER TECHNICIAN INSPECTED THE BED AND WAS UNABLE TO DUPLICATE ANY ISSUES. THE TECHNICIAN FOUND THE BED TO BE FUNCTIONING AS DESIGNED. NO FURTHER INFORMATION WAS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Unspecified Tissue Injury
MAUDE report 25232663 ↗MalfunctionMay 4, 2026Material Split, Cut or Torn
A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT FLAT DECK 36" MRS W/ X-RAY SLV HAD AN ISSUE, POWER CORD IS CUT IN HALF. USER DEFINED DESCRIPTION: WAREHOUSE REPAIR. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25070897 ↗MalfunctionApr 14, 2026Material Split, Cut or Torn
A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT FLAT DECK 36" MRS W/ X-RAY SLV HAD AN ISSUE, FRAYED POWER CORD (EXPOSED COPPER). THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24879204 ↗MalfunctionFeb 26, 2026Material Split, Cut or Torn
ON (B)(6) 2026, DURING SERVICING BY BAXTER, TECHNICAL SERVICE IDENTIFIED THAT FLAT DECK 36" MRS W/ X-RAY SLV (PRODUCT CODE P7924ARENT01, SERIAL NUMBER (B)(6)), HAD POWER CORD DAMAGED WITH COPPER WIRES EXPOSED. THIS OCCURRED DURING BAXTER SERVICING/TESTING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24463888 ↗MalfunctionJun 10, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22177846 ↗MalfunctionMay 23, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT FLAT DECK 36" MRS W/ X-RAY SLV (PRODUCT CODE P7924ARENT01, SERIAL NUMBER (B)(6)), HAD POWER CORD DAMAGE WITH EXPOSED COPPER WIRES. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22080476 ↗MalfunctionFeb 18, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT FLAT DECK 36" MRS W/ X-RAY SLV (PRODUCT CODE P7924ARENT01, SERIAL NUMBER (B)(6)), HAD POWER CORD DAMAGE WITH EXPOSED COPPER WIRES. THIS OCCURRED AFTER PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21397546 ↗MalfunctionFeb 12, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT FLAT DECK 36" MRS W/ X-RAY SLV (PRODUCT CODE P7924ARENT01, SERIAL NUMBER (B)(6), HAD THE POWER CORD DAMAGE WITH EXPOSED COPPER WIRES. THIS OCCURRED DURING BAXTER SERVICING/TESTING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21366467 ↗MalfunctionSep 20, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20267779 ↗MalfunctionAug 9, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19954961 ↗MalfunctionMay 23, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19385703 ↗MalfunctionMay 10, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19284357 ↗MalfunctionApr 9, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT (B)(4) .
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19075456 ↗MalfunctionDec 19, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE BED HAD POWER CORD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18356407 ↗
FDA clearance and classification
The FDA regulates product code IOQ (bed, flotation therapy, powered) under 21 CFR 890.5170 as a Class II device. 21 CFR 890.5170 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the pro+ P7924ARENT01
Has the pro+ P7924ARENT01 been recalled?
Yes. 3 FDA recall records name this device's identifier: 3 still open. The most recent, Z-2151-2024 (Class II), began on May 8, 2024: There is a potential for the top cover of the hospital bed mattress to delaminate as a result of improper cleaning and lack of adequate details on the appropriate use of cleaning agents containing bleach. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the pro+ P7924ARENT01?
The FDA's MAUDE database lists 15 adverse event reports for this device, received between Dec 19, 2023 and Aug 11, 2026. In those reports, the most common specific device problem is “Material Split, Cut or Torn” (14); the most common patient problem is “Unspecified Tissue Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the pro+ P7924ARENT01?
1 report is classified as a death, 0 as injuries and 14 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the pro+ P7924ARENT01 FDA approved?
Devices of this type (product code IOQ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Hill-Rom, Inc.
Who makes the pro+ P7924ARENT01, and is it still sold?
The FDA's device database lists Hill-Rom, Inc. as the company responsible for it (brand pro+, model P7924ARENT01). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Bed, Flotation Therapy, Powered devices
- Compella™Bariatric Bed P7800AHill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED1 recall1 death362 reports
- pro+ P7924A03Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED4 recalls3 reports
- pro+ P3255A02Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls8 reports
- pro+ P7924A01Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls4 reports
- pro+ P3255A01Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls3 reports
- pro+ P7923A03Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls3 reports
- pro+ P7923A04Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls0 reports
- pro+ P7923A01Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page