pro+ P7924A01Recalls & reported problems
Hill-Rom, Inc. · Hospital Bed Surface · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00887761977907), received between Oct 26, 2022 and Oct 27, 2022 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Component or Accessory Incompatibility” (4 reports).
3 FDA recall records name this device's identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- May 8, 2024Open, Classified
There is a potential for the top cover of the hospital bed mattress to delaminate as a result of improper cleaning and lack of adequate details on the appropriate use of cleaning agents containing bleach.
Z-2151-2024 · Root cause: Error in labeling · FDA record ↗
- Mar 31, 2023Open, Classified
There is a potential for mattress cover damage (delamination) which may result in fluid absorption in the mattress (fluid ingress).
Z-1482-2023 · Root cause: Process change control · FDA record ↗
- Jun 24, 2022Open, Classified
The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of pressure ulcer development.
Z-1411-2022 · Root cause: Nonconforming Material/Component · FDA record ↗
Most reported problems
Device problems
- Component or Accessory Incompatibility4
- Inflation Problem4
Patient problems
- Pressure Sore/Ulcer2
- Unspecified Tissue Injury2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryOct 27, 2022Inflation Problem; Component or Accessory Incompatibility; Inflation Problem; Component or Accessory Incompatibility
THE CUSTOMER REPORTED THAT THE PRO+ MATTRESS WAS COMPRESSED AND THAT AN INCREASE IN THE DEVELOPMENT OF PRESSURE INJURIES WAS ASSOCIATED WITH ITS USE. THE CUSTOMER REPORT COULD NOT ASSOCIATE A PATIENT WITH A SPECIFIC BED/SERIAL NUMBER BUT DID PROVIDE A ROOM NUMBER THE PATIENT OCCUPIED DURING THE HOSPITALIZATION. THE REPORTED INJURIES ASSOCIATED WITH PATIENTS 1- 7 HAVE BEEN EVALUATED SEPARATELY. THIS EVALUATION ADDRESSES PATIENT 6. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: Unspecified Tissue Injury
MAUDE report 15678861 ↗InjuryOct 26, 2022Inflation Problem; Component or Accessory Incompatibility; Inflation Problem; Component or Accessory Incompatibility
THE CUSTOMER REPORTED THAT THE PRO+ MATTRESS WAS COMPRESSED AND THAT AN INCREASE IN THE DEVELOPMENT OF PRESSURE INJURIES WAS ASSOCIATED WITH ITS USE. THE CUSTOMER REPORT COULD NOT ASSOCIATE A PATIENT WITH A SPECIFIC BED/SERIAL NUMBER BUT DID PROVIDE A ROOM NUMBER THE PATIENT OCCUPIED DURING THE HOSPITALIZATION. THE REPORTED INJURIES ASSOCIATED WITH PATIENTS 1- 7 HAVE BEEN EVALUATED SEPARATELY. THIS EVALUATION ADDRESSES PATIENT 2. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: Unspecified Tissue Injury
MAUDE report 15676853 ↗InjuryOct 26, 2022Inflation Problem; Component or Accessory Incompatibility; Inflation Problem; Component or Accessory Incompatibility
THE CUSTOMER REPORTED THAT THE PRO+ MATTRESS WAS COMPRESSED AND THAT AN INCREASE IN THE DEVELOPMENT OF PRESSURE INJURIES WAS ASSOCIATED WITH ITS USE. THE CUSTOMER REPORT COULD NOT ASSOCIATE A PATIENT WITH A SPECIFIC BED/SERIAL NUMBER BUT DID PROVIDE A ROOM NUMBER THE PATIENT OCCUPIED DURING THE HOSPITALIZATION. THE REPORTED INJURIES ASSOCIATED WITH PATIENTS 1- 7 HAVE BEEN EVALUATED SEPARATELY. THIS EVALUATION ADDRESSES PATIENT 3. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS (B)(4).
Patient problems: Pressure Sore/Ulcer
MAUDE report 15676806 ↗MalfunctionOct 26, 2022Inflation Problem; Component or Accessory Incompatibility; Inflation Problem; Component or Accessory Incompatibility
THE CUSTOMER REPORTED THAT THE PRO+ MATTRESS WAS COMPRESSED AND THAT AN INCREASE IN THE DEVELOPMENT OF PRESSURE INJURIES WAS ASSOCIATED WITH ITS USE. THE CUSTOMER REPORT COULD NOT ASSOCIATE A PATIENT WITH A SPECIFIC BED/SERIAL NUMBER BUT DID PROVIDE A ROOM NUMBER THE PATIENT OCCUPIED DURING THE HOSPITALIZATION. THE REPORTED INJURIES ASSOCIATED WITH PATIENTS 1- 7 HAVE BEEN EVALUATED SEPARATELY. THIS EVALUATION ADDRESSES PATIENT 5. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: Pressure Sore/Ulcer
MAUDE report 15674732 ↗
Other BED, FLOTATION THERAPY, POWERED devices
- Compella™Bariatric Bed P7800AHill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED1 recall1 death362 reports
- pro+ P7924A03Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED4 recalls3 reports
- pro+ P7924ARENT01Hill-Rom, Inc. · Bed, Flotation Therapy, Powered3 recalls1 death15 reports
- pro+ P3255A02Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls8 reports
- pro+ P3255A01Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls3 reports
- pro+ P7923A03Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls2 reports
- pro+ P7923A01Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls0 reports
- pro+ P7923A04Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record