Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887761977860), received between May 6, 2024 and Apr 3, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (3 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 8, 2024Class IIOpen, Classified

    There is a potential for the top cover of the hospital bed mattress to delaminate as a result of improper cleaning and lack of adequate details on the appropriate use of cleaning agents containing bleach.

    Z-2151-2024 · Root cause: Error in labeling · Recall details · FDA record ↗

  2. Mar 31, 2023Class IIOpen, Classified

    There is a potential for mattress cover damage (delamination) which may result in fluid absorption in the mattress (fluid ingress).

    Z-1482-2023 · Root cause: Process change control · Recall details · FDA record ↗

  3. Jun 24, 2022Class IIOpen, Classified

    The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of pressure ulcer development.

    Z-1413-2022 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Split, Cut or Torn3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 3, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT PRO+ VERSACARE MRS WITH X-RAY (PRODUCT CODE P3255A01, SERIAL NUMBER (B)(6), HAD POWER CORD DAMAGED WITH EXPOSE COPPER WIRE. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21765108 ↗
  • MalfunctionMay 13, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19297941 ↗
  • MalfunctionMay 6, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE POWER CORD WAS DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT (B)(4) .

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19253897 ↗

FDA clearance and classification

The FDA regulates product code IOQ (bed, flotation therapy, powered) under 21 CFR 890.5170 as a Class II device. 21 CFR 890.5170 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the pro+ P3255A01

Has the pro+ P3255A01 been recalled?

Yes. 3 FDA recall records name this device's identifier: 3 still open. The most recent, Z-2151-2024 (Class II), began on May 8, 2024: There is a potential for the top cover of the hospital bed mattress to delaminate as a result of improper cleaning and lack of adequate details on the appropriate use of cleaning agents containing bleach. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the pro+ P3255A01?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between May 6, 2024 and Apr 3, 2025. In those reports, the most common specific device problem is “Material Split, Cut or Torn” (3). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the pro+ P3255A01?

None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the pro+ P3255A01 FDA approved?

Devices of this type (product code IOQ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Hill-Rom, Inc.

Who makes the pro+ P3255A01, and is it still sold?

The FDA's device database lists Hill-Rom, Inc. as the company responsible for it (brand pro+, model P3255A01). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.