pro+ P7924A03Recalls & reported problems

Hill-Rom, Inc. · Hospital Bed Surface · Class 2

4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
4

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887761977884), received between Dec 1, 2023 and Dec 19, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (3 reports).

4 FDA recall records name this device's identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 8, 2024Open, Classified

    There is a potential for the top cover of the hospital bed mattress to delaminate as a result of improper cleaning and lack of adequate details on the appropriate use of cleaning agents containing bleach.

    Z-2151-2024 · Root cause: Error in labeling · FDA record ↗

  2. Mar 5, 2024Open, Classified

    Inconsistencies were identified with service records associated with corrections performed for a previous Medical Device Correction issued by Baxter (ref number: FA-2022-026). Your devices have been identified as potentially impacted by these service inconsistencies and therefore, Baxter will need to reassess the impacted mattresses to confirm they have been corrected as identified by the service order, and to confirm the correct serial numbers have been identified.

    Z-1400-2024 · Root cause: Under Investigation by firm · FDA record ↗

  3. Mar 31, 2023Open, Classified

    There is a potential for mattress cover damage (delamination) which may result in fluid absorption in the mattress (fluid ingress).

    Z-1482-2023 · Root cause: Process change control · FDA record ↗

  4. Jun 24, 2022Open, Classified

    The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of pressure ulcer development.

    Z-1412-2022 · Root cause: Nonconforming Material/Component · FDA record ↗

Most reported problems

Device problems

  • Material Split, Cut or Torn3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 19, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    ON 04-DEC-2025, A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT PRO+ SURFACE 36" WIDE W/O XRAY (PRODUCT CODE: P7924A03, SERIAL NUMBER: (B)(6), HAD POWER CORD DAMAGED WITH COPPER WIRES EXPOSED. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23854854 ↗
  • MalfunctionMar 25, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    BAXTER RECEIVED A REPORT STATING THAT PRO+ SURFACE POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT AND OCCURRED BEFORE USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21690494 ↗
  • MalfunctionDec 1, 2023Material Split, Cut or Torn; Material Split, Cut or Torn

    BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE MATTRESS HAD POWER CORD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18251213 ↗

Other BED, FLOTATION THERAPY, POWERED devices

All BED, FLOTATION THERAPY, POWERED devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.