pro+ P7923A03Recalls & reported problems
Hill-Rom, Inc. · Hospital Bed Surface · Class 2
All models of pro+ (18)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887761977921), received between Nov 1, 2024 and Sep 21, 2026 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Degraded” (1 report).
3 FDA recall records name this device's identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- May 8, 2024Class IIOpen, Classified
There is a potential for the top cover of the hospital bed mattress to delaminate as a result of improper cleaning and lack of adequate details on the appropriate use of cleaning agents containing bleach.
Z-2151-2024 · Root cause: Error in labeling · Recall details · FDA record ↗
- Mar 31, 2023Class IIOpen, Classified
There is a potential for mattress cover damage (delamination) which may result in fluid absorption in the mattress (fluid ingress).
Z-1482-2023 · Root cause: Process change control · Recall details · FDA record ↗
- Jun 24, 2022Class IIOpen, Classified
The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of pressure ulcer development.
Z-1416-2022 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Degraded1
- Insufficient Device Problem Information1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Pressure Sore/Ulcer1
- Unspecified Tissue Injury1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjurySep 21, 2026Insufficient Device Problem Information
The customer reported severe foam deterioration of the Pro+ mattress and that the patient developed a severe pressure injury. Follow-up was performed, and the customer reported that the patient was admitted from the PACU after undergoing a sigmoid resection with end colostomy and appendectomy, complicated by a ureteric injury that required ureteral reimplantation with a psoas hitch. The patient subsequently developed septic shock and was admitted to the ICU. The patient reportedly developed a deep tissue pressure injury to the sacrum that required daily dressing care with Inadine and Mepilex and a ¿plastics¿ consultation. At the time of this report, no additional information was available regarding the outcome of the consultation.
Patient problems: Unspecified Tissue Injury
MAUDE report 50849269 ↗MalfunctionMay 8, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE PRO+ SURFACE - 36" WIDTH POWER CORD HAD COPPER WIRES EXPOSED. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT KNOW IF THIS EVENT WAS ALREADY COMMUNICATED TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY. THE DEVICE WAS REQUESTED FOR SERVICE/INSPECTION BY BAXTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21989120 ↗InjuryNov 1, 2024Degraded; Degraded
IT WAS REPORTED THAT THE SURFACE OF THE PRO+ CENTRELLA MATTRESS PRESENTED DELAMINATION, AND A PATIENT DEVELOPED A STAGE 3 PRESSURE INJURY. THE CUSTOMER NOTED THE SURFACE HAD BEEN CLEANED USING OXIVIR PLUS, A CLEANING AGENT LISTED AS AN APPROVED CLEANER FOR THE MATTRESS. MULTIPLE ATTEMPTS WERE MADE TO OBTAIN SPECIFIC DETAILS ABOUT THE DEVELOPMENT OF THE REPORTED PRESSURE INJURY, EVENTUAL MEDICAL/SURGICAL INTERVENTION REQUIRED, AND ITS CURRENT STAGE; HOWEVER, AT THIS TIME, NO ADDITIONAL INFORMATION COULD BE OBTAINED FROM THE CUSTOMER.
Patient problems: Pressure Sore/Ulcer
MAUDE report 20587980 ↗
FDA clearance and classification
The FDA regulates product code IOQ (bed, flotation therapy, powered) under 21 CFR 890.5170 as a Class II device. 21 CFR 890.5170 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the pro+ P7923A03
Has the pro+ P7923A03 been recalled?
Yes. 3 FDA recall records name this device's identifier: 3 still open. The most recent, Z-2151-2024 (Class II), began on May 8, 2024: There is a potential for the top cover of the hospital bed mattress to delaminate as a result of improper cleaning and lack of adequate details on the appropriate use of cleaning agents containing bleach. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the pro+ P7923A03?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Nov 1, 2024 and Sep 21, 2026. In those reports, the most common specific device problems are “Degraded” (1) and “Material Split, Cut or Torn” (1); the most common patient problems are “Pressure Sore/Ulcer” (1) and “Unspecified Tissue Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the pro+ P7923A03?
None of the reports are classified as deaths, 2 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the pro+ P7923A03 FDA approved?
Devices of this type (product code IOQ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Hill-Rom, Inc.
Who makes the pro+ P7923A03, and is it still sold?
The FDA's device database lists Hill-Rom, Inc. as the company responsible for it (brand pro+, model P7923A03). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other BED, FLOTATION THERAPY, POWERED devices
- Compella™Bariatric Bed P7800AHill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED1 recall1 death362 reports
- pro+ P7924A03Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED4 recalls3 reports
- pro+ P7924ARENT01Hill-Rom, Inc. · Bed, Flotation Therapy, Powered3 recalls1 death15 reports
- pro+ P3255A02Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls8 reports
- pro+ P7924A01Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls4 reports
- pro+ P3255A01Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls3 reports
- pro+ P7923A04Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls0 reports
- pro+ P7923A01Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page