pro+ P3255A02Recalls & reported problems

Hill-Rom, Inc. · Hospital Bed Surface · Class 2

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00887761977853), received between Apr 28, 2023 and Nov 21, 2025 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (7 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 8, 2024Open, Classified

    There is a potential for the top cover of the hospital bed mattress to delaminate as a result of improper cleaning and lack of adequate details on the appropriate use of cleaning agents containing bleach.

    Z-2151-2024 · Root cause: Error in labeling · FDA record ↗

  2. Mar 31, 2023Open, Classified

    There is a potential for mattress cover damage (delamination) which may result in fluid absorption in the mattress (fluid ingress).

    Z-1482-2023 · Root cause: Process change control · FDA record ↗

  3. Jun 24, 2022Open, Classified

    The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of pressure ulcer development.

    Z-1414-2022 · Root cause: Nonconforming Material/Component · FDA record ↗

Most reported problems

Device problems

  • Material Split, Cut or Torn7
  • Inflation Problem1
  • Naturally Worn1
  • Sparking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 21, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23622058 ↗
  • MalfunctionNov 18, 2025Sparking; Material Split, Cut or Torn; Sparking; Material Split, Cut or Torn

    THE CUSTOMER CONTACTED BAXTER CUSTOMER SUPPORT LINE TO REQUEST ASSISTANCE IS PROPER ROUTING AND SECURING THE POWER CORD OF THE PRO+ SURFACE. THE CUSTOMER STATED THAT THE CORDS ARE BECOMING DAMAGED DUE TO RUBBING AND PULLING/STRAIN AND IN ONCE INSTANCE THE CORD WAS BECAME DAMAGED WITH EXPOSE WIRES THREW SPARKS. THE CUSTOMER ALSO STATED THAT A STAFF MEMBER WAS SHOCKED WHEN CLEANING A BED WITH A DAMAGED POWER CORD. THE STAFF WAS SEEN IN THE EMERGENCY ROOM, BASED ON THE CUSTOMER KNOWLEDGE OF THE EVENT, THE CUSTOMER DOES NOT BELIEVE THAT TREATMENT WAS REQUIRED. MULTIPLE ATTEMPTS TO OBTAIN FURTHER DETAILS OF ANY INJURY OF THE SHOCK HAVE BEEN UNSUCCESSFUL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23586480 ↗
  • MalfunctionOct 15, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE PRO+ VERSACARE MRS W/O X-RAY POWER CORD HAD COPPER WIRES EXPOSED. THE BED WAS LOCATED AT THE ACCOUNT. THIS OCCURRED DURING BAXTER SERVICING/TESTING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WAS A SECOND DEVICE ALSO REPORTED TO BE INVOLVED: MODEL NUMBER: P3200 SERIAL NUMBER: (B)(6). THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23301313 ↗
  • MalfunctionSep 16, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    DURING SERVICING BY BAXTER, TECHNICAL SERVICE IDENTIFIED THAT PRO+ VERSACARE MRS W/O X-RAY (PRODUCT CODE P3255A02, SERIAL NUMBER (B)(6)), HAD POWER CORD DAMGED WITH COPPER WIRES EXPOSED. THIS OCCURRED DURING BAXTER SERVICING/TESTING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23064024 ↗
  • MalfunctionMay 16, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT PRO+ VERSACARE MRS W/O X-RAY (PRODUCT CODE P3255A02, SERIAL NUMBER (B)(6)), HAD POWER CORD DAMAGE WITH EXPOSED COPPER WIRES. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22038219 ↗
  • MalfunctionJul 25, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19833664 ↗
  • MalfunctionJun 5, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19472807 ↗
  • MalfunctionApr 28, 2023Inflation Problem; Naturally Worn; Inflation Problem; Naturally Worn

    THE CUSTOMER REPORTED THE SEAT SECTION IS WEARING DOWN ON THE PRO PLUS MATTRESS WITH USE ON A VERSACARE FRAME. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16833528 ↗

Other BED, FLOTATION THERAPY, POWERED devices

All BED, FLOTATION THERAPY, POWERED devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.