pro+ P3255A02Recalls & reported problems
Hill-Rom, Inc. · Hospital Bed Surface · Class 2
All models of pro+ (18)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00887761977853), received between Apr 28, 2023 and Nov 21, 2025 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (7 reports).
3 FDA recall records name this device's identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- May 8, 2024Class IIOpen, Classified
There is a potential for the top cover of the hospital bed mattress to delaminate as a result of improper cleaning and lack of adequate details on the appropriate use of cleaning agents containing bleach.
Z-2151-2024 · Root cause: Error in labeling · Recall details · FDA record ↗
- Mar 31, 2023Class IIOpen, Classified
There is a potential for mattress cover damage (delamination) which may result in fluid absorption in the mattress (fluid ingress).
Z-1482-2023 · Root cause: Process change control · Recall details · FDA record ↗
- Jun 24, 2022Class IIOpen, Classified
The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of pressure ulcer development.
Z-1414-2022 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Split, Cut or Torn7
- Inflation Problem1
- Naturally Worn1
- Sparking1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 21, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23622058 ↗MalfunctionNov 18, 2025Sparking; Material Split, Cut or Torn; Sparking; Material Split, Cut or Torn
THE CUSTOMER CONTACTED BAXTER CUSTOMER SUPPORT LINE TO REQUEST ASSISTANCE IS PROPER ROUTING AND SECURING THE POWER CORD OF THE PRO+ SURFACE. THE CUSTOMER STATED THAT THE CORDS ARE BECOMING DAMAGED DUE TO RUBBING AND PULLING/STRAIN AND IN ONCE INSTANCE THE CORD WAS BECAME DAMAGED WITH EXPOSE WIRES THREW SPARKS. THE CUSTOMER ALSO STATED THAT A STAFF MEMBER WAS SHOCKED WHEN CLEANING A BED WITH A DAMAGED POWER CORD. THE STAFF WAS SEEN IN THE EMERGENCY ROOM, BASED ON THE CUSTOMER KNOWLEDGE OF THE EVENT, THE CUSTOMER DOES NOT BELIEVE THAT TREATMENT WAS REQUIRED. MULTIPLE ATTEMPTS TO OBTAIN FURTHER DETAILS OF ANY INJURY OF THE SHOCK HAVE BEEN UNSUCCESSFUL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23586480 ↗MalfunctionOct 15, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE PRO+ VERSACARE MRS W/O X-RAY POWER CORD HAD COPPER WIRES EXPOSED. THE BED WAS LOCATED AT THE ACCOUNT. THIS OCCURRED DURING BAXTER SERVICING/TESTING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WAS A SECOND DEVICE ALSO REPORTED TO BE INVOLVED: MODEL NUMBER: P3200 SERIAL NUMBER: (B)(6). THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23301313 ↗MalfunctionSep 16, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
DURING SERVICING BY BAXTER, TECHNICAL SERVICE IDENTIFIED THAT PRO+ VERSACARE MRS W/O X-RAY (PRODUCT CODE P3255A02, SERIAL NUMBER (B)(6)), HAD POWER CORD DAMGED WITH COPPER WIRES EXPOSED. THIS OCCURRED DURING BAXTER SERVICING/TESTING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23064024 ↗MalfunctionMay 16, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT PRO+ VERSACARE MRS W/O X-RAY (PRODUCT CODE P3255A02, SERIAL NUMBER (B)(6)), HAD POWER CORD DAMAGE WITH EXPOSED COPPER WIRES. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22038219 ↗MalfunctionJul 25, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19833664 ↗MalfunctionJun 5, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19472807 ↗MalfunctionApr 28, 2023Inflation Problem; Naturally Worn; Inflation Problem; Naturally Worn
THE CUSTOMER REPORTED THE SEAT SECTION IS WEARING DOWN ON THE PRO PLUS MATTRESS WITH USE ON A VERSACARE FRAME. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16833528 ↗
FDA clearance and classification
The FDA regulates product code IOQ (bed, flotation therapy, powered) under 21 CFR 890.5170 as a Class II device. 21 CFR 890.5170 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the pro+ P3255A02
Has the pro+ P3255A02 been recalled?
Yes. 3 FDA recall records name this device's identifier: 3 still open. The most recent, Z-2151-2024 (Class II), began on May 8, 2024: There is a potential for the top cover of the hospital bed mattress to delaminate as a result of improper cleaning and lack of adequate details on the appropriate use of cleaning agents containing bleach. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the pro+ P3255A02?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between Apr 28, 2023 and Nov 21, 2025. In those reports, the most common specific device problems are “Material Split, Cut or Torn” (7), “Inflation Problem” (1) and “Naturally Worn” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the pro+ P3255A02?
None of the reports are classified as deaths, 0 as injuries and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the pro+ P3255A02 FDA approved?
Devices of this type (product code IOQ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Hill-Rom, Inc.
Who makes the pro+ P3255A02, and is it still sold?
The FDA's device database lists Hill-Rom, Inc. as the company responsible for it (brand pro+, model P3255A02). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other BED, FLOTATION THERAPY, POWERED devices
- Compella™Bariatric Bed P7800AHill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED1 recall1 death362 reports
- pro+ P7924A03Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED4 recalls3 reports
- pro+ P7924ARENT01Hill-Rom, Inc. · Bed, Flotation Therapy, Powered3 recalls1 death15 reports
- pro+ P7924A01Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls4 reports
- pro+ P3255A01Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls3 reports
- pro+ P7923A03Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls3 reports
- pro+ P7923A04Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls0 reports
- pro+ P7923A01Hill-Rom, Inc. · BED, FLOTATION THERAPY, POWERED3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page