EVIS EXERA GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE 2T160 (UDI 04953170307492)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of EVIS EXERA GASTROINTESTINAL VIDEOSCOPE (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04953170307492), received between Jul 5, 2020 and Mar 30, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Device Reprocessing Problem” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Sep 25, 2023Class IIOpen, Classified
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Z-0334-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Reprocessing Problem1
- Microbial Contamination of Device1
- Poor Quality Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2023Poor Quality Image; Poor Quality Image
THE CUSTOMER REPORTED TO OLYMPUS THAT THEY OBSERVED A LEAKAGE FROM THE INSTRUMENT CHANNEL, ANGULATION PLAY IN ALL DIRECTIONS, LOOSE ANGULATION KNOB, AND WRINKLES ON THE CONNECTING TUBE. THERE WERE NO REPORTS OF PATIENT HARM ASSOCIATED WITH THIS EVENT. THE SUBJECT DEVICE WAS SENT BACK TO OLYMPUS FOR EVALUATION. DURING INSPECTION AND TESTING THE FOLLOWING WAS FOUND: THE ENDOSCOPIC IMAGE WAS BLURRED. THIS REPORT IS BEING SUBMITTED FOR THE MALFUNCTION FOUND DURING EVALUATION (THE ENDOSCOPIC IMAGE WAS BLURRED).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16642310 ↗MalfunctionDec 22, 2022Device Reprocessing Problem; Device Reprocessing Problem
COMPANY REPRESENTATIVE SENT REPAIR REQUEST REPORTED WITH AN ISSUE OF "CHEMICAL ADHESION" . THE ISSUE FOUND DURING AN UNKNOWN EVENT. NO HARM WAS REPORTED. NO PATIENT HARM, NO USER INJURY REPORTED DUE TO THE EVENT. DEVICE EVALUATION FOUND THE DEVICE CONNECTING TUBE HAS FOREIGN MATERIAL . THIS REPORT IS BEING SUBMITTED FOR FOREIGN MATTER ADHERED (PROBLEMS RELATED TO REPROCESSING, INSUFFICIENT REPROCESSING ).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16040378 ↗MalfunctionJul 5, 2020Microbial Contamination of Device; Microbial Contamination of Device
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT AS A RESULT OF MICROBIOLOGICAL TESTING BY THE USER FACILITY, FOLLOWING MICROBES WERE DETECTED FROM THE SAMPLE COLLECTED FROM THE SUBJECT DEVICE. INSTRUMENT CHANNEL: FUNGAL HYPHAE (>20 CFU/ENDOSCOPE), BACILLUS SPECIES (>20 CFU/ENDOSCOPE), COAGULASE-NEGATIVE STAPHYLOCOCCI (>20 CFU/ENDOSCOPE) THE DEVICE HAD BEEN REPROCESSED WITH AN OLYMPUS AUTOMATED ENDOSCOPE REPROCESSOR MODEL ETD3 (NOT AVAILABLE IN THE USA), USING GLUTARALDEHYDE. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10230998 ↗
Other Endoscopic video imaging system/component, gastroenterology-urology devices
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EBOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,522 reports
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EAOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,375 reports
- HD CAMERA HEAD OTV-S7ProH-HD-L08EOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1,424 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology529 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology4 deaths413 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1 recall189 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic Video Imaging System/Component, Gastroenterology-Urology158 reports
- VIDEO SYSTEM OTV-SIOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology143 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page