Skip to content

EVIS EXERA COLONOVIDEOSCOPE OLYMPUS CF TYPE 2T160IRecalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of EVIS EXERA COLONOVIDEOSCOPE (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04953170307645), received between Feb 23, 2022 and Feb 23, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Microbial Contamination of Device” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0329-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Microbial Contamination of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 23, 2022Microbial Contamination of Device; Microbial Contamination of Device

    AS REPORTED, AFTER A MICROBIOLOGICAL ROUTINE CONTROL TEST ON THE SUBJECT MEDICAL DEVICE AS REQUIRED BY FRENCH REGULATION, THE USER DETECTED AN UNEXPECTED CONTAMINATION. THE ISSUE FOUND DURING REPROCESSING. THE USER THEN SENT THE DEVICE TO THE OFR (OLYMPUS FRANCE) FRENCH OLYMPUS SUBSIDIARY FOR HMI (HYGIENE MICROBIOLOGICAL INVESTIGATION) FOR FURTHER INVESTIGATION. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY PATIENT INJURY OR PATIENT INFECTION TO WHICH THIS MEDICAL DEVICE COULD HAVE BEEN A CONTRIBUTORY CAUSE. NO USER INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13589153 ↗

Other Endoscopic video imaging system/component, gastroenterology-urology devices

All Endoscopic video imaging system/component, gastroenterology-urology devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page