Skip to content

Titan recallZ-1365-2023 · Coloplast Manufacturing US, LLC · initiated Mar 1, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Coloplast Manufacturing US, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.

FDA root cause category: Process change control

What the company is doing

The firm issued an URGENT MEDICAL DEVICE VOLUNTARY RECALL notice on 03/01/2023 by UPS next day air. The notice explained the issue and the hazard to health. The notice requests the removal and return of the affected product. Distributors are instructed to notify their customers. For further questions, please contact the Coloplast Customer Service team directly at (800) 258-3476, or by email [email protected].

Recall details

Recall number
Z-1365-2023
Status
Open, Classified
Initiated by the company
Mar 1, 2023
Posted by the FDA
Apr 12, 2023
Recalling firm
Coloplast Manufacturing US, LLC
Quantity
2 units
Product code
FHW
FDA submissions
P000006
Distribution
Worldwide distribution - US Nationwide and the countries of Turkey, Italy, Germany, Belgium, France, United Kingdom, Spain, Portugal.

Product description

Titan NB Infra Zero Ang 16cm, Catalog Number EN88162400; inflatable penile Prosthesis

Code information (models, lots, serial numbers)

UDI/DI 05708932072595, Lot Numbers: 8887620, 8939124

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

65 devices cleared under P000006 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 91800).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page