Azurion recallZ-1113-2026 · Philips Medical Systems Nederland B.V. · initiated Dec 12, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.
FDA root cause category: Software design
What the company is doing
On December 12, 2025, Philips sent URGENT Medical Device Correction letters to customers. Actions to be taken: - Circulate this Urgent Medical Device Correction letter to all users of the system so that they are aware of the issues. " In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. " Keep this Urgent Medical Device Correction letter with the documentation of the system until Philips corrects your system. Ensure that the letter is in a place likely to be seen/viewed. " Complete and return the response form included in this Urgent Medical Device Correction letter to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction letter and understanding of the issues and required actions to be taken. " If you experience an issue described in this letter, please report it to your local Philips representative. Your local Philips representative will contact you to schedule visits to install the software updates once available. ***Update May 8, 2026*** Previously, Philips planned to resolve issue 1 with software update 3.1.5 and issue 2 with software update 3.1.15. Philips has now updated their plan thusly: Both issues 1 and 2 will be included with software release 3.1.15. Anticipated release date of software version 3.1.15 has been moved to Q3 2026.
Recall details
- Recall number
- Z-1113-2026
- Status
- Open, Classified
- Initiated by the company
- Dec 12, 2025
- Posted by the FDA
- Jan 16, 2026
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Quantity
- 228 units (all OUS)
- Product code
- OWB
- FDA submissions
- K200917
- Distribution
- International distribution in the countries of AE, AR, AT, AZ, BE, BG, BH, BR, BS, CH, CL, CO, CR, CY, CZ, DE, DK, DZ, EC, EG, ES, GB, GR, GT, ID, IE, IL, IN, IQ, IT, JE, JP, LB, MA, MO, NI, NL, NO, PE, PL, RO, SA, SE, SV, TR, UA, VE, VN, ZA;
Product description
Philips Azurion system; Software Version Number: R3.1;
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
Software Version Number: R3.1; (1) Model Number: 722221; System Product Name: Azurion 3 M12; UDI-DI: 00884838099203; Serial Numbers: (2) Model Number: 722222; System Product Name: Azurion 3 M15; UDI-DI: 00884838099210; Serial Numbers: (3) Model Number: 722223; System Product Name: Azurion 7 M12; UDI-DI: 00884838099241; Serial Numbers: (4) Model Number: 722224; System Product Name: Azurion 7 M20; UDI-DI: 00884838099258; Serial Numbers: (5) Model Number: 722225; System Product Name: Azurion 7 B12; UDI-DI: 00884838099265; Serial Numbers: (6) Model Number: 722226; System Product Name: Azurion 7 B20; UDI-DI: 00884838099272; Serial Numbers: (7) Model Number: 722227; System Product Name: Azurion 5 M12; UDI-DI: 00884838099227; Serial Numbers: (8) Model Number: 722228; System Product Name: Azurion 5 M20; UDI-DI: 00884838099234; Serial Numbers: (9) Model Number: 722229; System Product Name: Azurion 3 M12; UDI-DI: 00884838116726; Serial Numbers: 5, 9, 32, 40, 35, 31, 33; (10) Model Number: 722230; System Product Name: Azurion 3 M15; UDI-DI: 00884838116733; Serial Numbers: 25, 54, 50, 52, 55, 30, 63, 60, 67; (11) Model Number: 722231; System Product Name: Azurion 5 M12; UDI-DI: 00884838116740; Serial Numbers: 62, 7, 8, 14, 17, 63, 6, 66, 59, 69, 56, 67, 48, 68, 80, 93, 87, 94, 81, 92, 18; (12) Model Number: 722232; System Product Name: Azurion 5 M20; UDI-DI: 00884838116757; Serial Numbers: 97, 39, 31, 100, 104, 93, 115, 113, 118, 119, 89, 54, 67, 7, 128, 86, 110, 132, 130, 131, 123, 137, 139, 117, 260; (13) Model Number: 722233; System Product Name: Azurion 7 M12; UDI-DI: 00884838116764; Serial Numbers: 239, 100, 233, 240, 176, 227, 223, 130, 161, 194, 243, 56, 193, 280, 75, 253, 156, 288, 72, 281, 304, 300, 105, 295, 312, 310, 285, 292, 290, 298, 297; (14) Model Number: 722234; System Product Name: Azurion 7 M20; UDI-DI: 00884838116771; Serial Numbers: 731, 727, 261, 473, 683, 722, 562, 750, 745, 766, 265, 701, 400, 310, 421, 314, 182, 579, 96, 763, 272, 612, 15, 844, 105, 702, 751, 384, 851, 869, 743, 337, 651, 455, 706, 312, 563, 709, 764, 871, 846, 278, 713, 802, 868, 277, 742, 184, 819, 759, 910, 928, 415, 414, 945, 961, 355, 933, 692, 920, 569, 939, 909, 553, 837, 856, 571, 756, 941, 570, 761, 981, 353, 888, 964, 820, 746, 736, 876, 599, 1017, 908, 487, 983, 495; (15) Model Number: 722235; System Product Name: Azurion 7 B12; UDI-DI: 00884838116788; Serial Numbers: 96, 79, 111, 100, 101, 38, 27, 92, 78, 118, 48, 131, 135, 22, 125, 103, 138, 139, 137; (16) Model Number: 722236; System Product Name: Azurion 7 B20; UDI-DI: 00884838116801; Serial Numbers: 241, 297, 198, 215, 305, 274, 299, 78, 279, 287, 318, 353, 324, 331, 293, 245, 314, 310, 317, 312, 77, 358, 206, 383, 380, 370, 377, 385, 405, 379, 328;
Devices named in this recall
16 devices in the FDA device registry whose identifier appears in the recall's code information.
- Azurion Azurion 7 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System32 recalls2 deaths2,308 reports
- Azurion Azurion 7 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls2 deaths835 reports
- Azurion Azurion 7 B20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls1 death634 reports
- Azurion Azurion 5 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls201 reports
- Azurion Azurion 7 B12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls232 reports
- Azurion Azurion 5 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System21 recalls1 death197 reports
- Azurion Azurion 3 M15Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System21 recalls104 reports
- Azurion Azurion 3 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls121 reports
- Azurion Azurion 7 M20Philips Medical Systems Nederland B.V. · Interventional fluoroscopic x-ray system11 recalls1 death108 reports
- Azurion Azurion 7 B12Philips Medical Systems Nederland B.V. · Interventional fluoroscopic x-ray system11 recalls5 reports
- Azurion Azurion 7 B20Philips Medical Systems Nederland B.V. · Interventional fluoroscopic x-ray system10 recalls1 death50 reports
- Azurion Azurion 7 M12Philips Medical Systems Nederland B.V. · Interventional fluoroscopic x-ray system10 recalls27 reports
- Azurion Azurion 5 M20Philips Medical Systems Nederland B.V. · System, x-ray, fluoroscopic, image-intensified9 recalls7 reports
- Azurion Azurion 5 M12Philips Medical Systems Nederland B.V. · Interventional fluoroscopic x-ray system8 recalls8 reports
- Azurion Azurion 3 M15Philips Medical Systems Nederland B.V. · System, x-ray, fluoroscopic, image-intensified7 recalls1 report
- Azurion Azurion 3 M12Philips Medical Systems Nederland B.V. · System, x-ray, fluoroscopic, image-intensified6 recalls4 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page