Skip to content

P500 Surface P005787RENT02Recalls & reported problems

Hill-Rom, Inc. · Hospital Bed Surface · Class 2

Adverse event reports
18
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00887761001022), received between Aug 21, 2023 and Sep 9, 2026 — classified as 18 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (16 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 1 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Split, Cut or Torn16
  • Sparking3
  • Material Frayed1
  • No Audible Alarm1

Patient problems

  • No Clinical Signs, Symptoms or Conditions18

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 9, 2026Sparking; Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50700011 ↗
  • MalfunctionSep 7, 2026Sparking

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50658393 ↗
  • MalfunctionAug 12, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50229761 ↗
  • MalfunctionJul 18, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25864844 ↗
  • MalfunctionJul 16, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25842723 ↗
  • MalfunctionMay 19, 2026Material Split, Cut or Torn

    A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT P500 MRS CONTROL UNIT HAD WIRES EXPOSED ON BLOWER. THIS OCCURRED BEFORE USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25219216 ↗
  • MalfunctionMar 11, 2026Material Split, Cut or Torn

    BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24572061 ↗
  • MalfunctionJan 9, 2026Material Split, Cut or Torn

    ON 13-DEC-2025, A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT P500 MRS CONTROL UNIT (PRODUCT CODE: P005787RENT02, SERIAL NUMBER: (B)(6), HAD POWER CORD DAMAGED WITH COPPER WIRES EXPOSED. THIS OCCURRED BEFORE USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24022077 ↗
  • MalfunctionAug 27, 2025Sparking; Material Split, Cut or Torn; Sparking; Material Split, Cut or Torn

    BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE P500 MRS CONTROL UNIT POWER CORD IS SPARKING WHEN THE BED IS RAISED OR LOWERED; THE POWER CORD HAD COPPER WIRES EXPOSED. THE BED WAS LOCATED AT THE ACCOUNT. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22903505 ↗
  • MalfunctionAug 19, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    BAXTER RECEIVED A REPORT STATING THAT P500 MRS CONTROL UNIT FRAME POWER CORD WAS MISSING A PRONG AND COPPER WIRES WERE SHOWING. THE BED WAS LOCATED AT THE ACCOUNT AND OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22841401 ↗
  • MalfunctionMay 29, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    BAXTER RECEIVED A REPORT STATING THAT P500 MRS CONTROL UNIT POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT AND OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22107610 ↗
  • MalfunctionJan 30, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT P500 MRS CONTROL UNIT (PRODUCT CODE P005787RENT02, SERIAL NUMBER (B)(6), HAD POWER CORD DAMAGE WITH EXPOSED COPPER WIRES. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21276019 ↗
  • MalfunctionDec 12, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT P500 MRS CONTROL UNIT POWER CORD IS FRAYED WITH EXPOSED WIRES. USER DEFINED DESCRIPTION:WAREHOUSE REPAIR. THIS OCCURRED AFTER PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20911444 ↗
  • MalfunctionOct 18, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    BAXTER RECEIVED A REPORT FROMA BAXTER TECHNICIAN TO REPORT THAT P500 MRS CONTROL UNIT POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT AND OCCURRED DURING BAXTER SERVICING/TESTING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20486784 ↗
  • MalfunctionOct 8, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THAT POWER CORD WAS EXPOSING COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT AND OCCURRED BEFORE USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT PR # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20403552 ↗
  • MalfunctionMay 1, 2024No Audible Alarm; No Audible Alarm

    BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE BED EXIT ALARM WAS NOT ALERTING THE NURSE'S STATION. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT(B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19220563 ↗
  • MalfunctionMar 27, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18986020 ↗
  • MalfunctionAug 21, 2023Material Frayed; Material Split, Cut or Torn; Material Frayed; Material Split, Cut or Torn

    HILLROM RECEIVED A REPORT FROM A HILLROM TECHNICIAN STATING THE BED HAD POWER CORD FRAYED WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17589535 ↗

Other BED, FLOTATION THERAPY, POWERED devices

All BED, FLOTATION THERAPY, POWERED devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.