Philips Allura (Model Numbers 722010, 722012, 722026… recallZ-0874-2022 · Philips North America Llc · initiated Feb 2, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient.
FDA root cause category: Under Investigation by firm
What the company is doing
Urgent Medical Device Correction notification letters dated 2/2/22 were distributed to customers. 1. What is the problem and under what circumstances it can occur Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient. When the problem occurs, the area of the boosted images remains black, and the small X-ray viewer in the upper right corner will show a single static image of the previous run. The correct patient information is shown on the Philips Allura/ Azurion system. This defect is intermittent and is caused by a race condition when the software does not properly handle all the "process requests". This problem has been identified through the investigation of 2 (two) customer complaints. 2. What is the hazard/harm assoclated with this issue The incorrect image displayed to the user could lead to incorrect treatment. If the problem occurs, the Stent8oost Live application will have to be restarted (see action 4), causing a delay in the procedure. To date, Philips has not received any reports of harm associated with this problem. 3. Affected products and how to identify them The Philips StentBoost Live R2.0 used with Philips Allura and Azurion systems is affected by this issue. The software version of the Philips StentBoost Live application can be seen by clicking on the "About" box displayed when the application is opened (see Fig. 1 and 2).
Recall details
- Recall number
- Z-0874-2022
- Status
- Open, Classified
- Initiated by the company
- Feb 2, 2022
- Recalling firm
- Philips North America Llc
- Quantity
- 338
- Product code
- LLZ
- FDA submissions
- K141979, K170144, K181830, K172822
- Distribution
- Global Distribution. US Nationwide.
Product description
Philips Allura (Model Numbers 722010, 722012, 722026, 722027, 722028, 722029, 722035) and Philips Azurion (Model Numbers 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228) running StentBoost Live (Model Number 459801781661) Software Version Number: R2.0
Code information (models, lots, serial numbers)
Show the full code information (10k characters)
UDI: (01)00884838059030(21)839 (01)00884838085275(21)131 (01)00884838059054(21)577 (01)00884838085350(21)81 (01)00884838085367(21)271 (01)00884838054202(21)879 (01)00884838054202(21)1783 (01)00884838054202(21)1447 (01)00884838099241(21)282 (01)00884838099258(21)512 (01)00884838099258(21)360 (01)00884838099258(21)370 (01)00884838099258(21)584 (01)00884838099272(21)239 (01)00884838099258(21)425 (01)00884838099258(21)232 (01)00884838099258(21)591 (01)00884838099258(21)580 (01)00884838099258(21)39 (01)00884838099241(21)198 (01)00884838099258(21)589 (01)00884838099258(21)362 (01)00884838085268(21)954 (01)00884838099234(21)10 (01)00884838099241(21)214 (01)00884838099272(21)183 (01)00884838099258(21)526 (01)00884838099258(21)441 (01)00884838099272(21)121 (01)00884838099258(21)80 (01)00884838099258(21)326 (01)00884838099258(21)518 (01)00884838099258(21)404 (01)00884838099241(21)252 (01)00884838099258(21)453 (01)00884838099258(21)472 (01)00884838099258(21)465 (01)00884838099272(21)129 (01)00884838099272(21)174 (01)00884838099258(21)553 (01)00884838099258(21)517 (01)00884838099258(21)400 (01)00884838099258(21)342 (01)00884838099258(21)582 (01)00884838099241(21)196 (01)00884838085268(21)1391 (01)00884838099258(21)485 (01)00884838099258(21)262 (01)00884838099258(21)277 (01)00884838099272(21)120 (01)00884838085268(21)703826 (01)00884838085367(21)38 (01)00884838085367(21)36 (01)00884838085275(21)139 (01)00884838085268(21)1199 (01)00884838085251(21)940 (01)00884838085251(21)945 (01)00884838085251(21)947 (01)00884838085268(21)703680 (01)00884838085251(21)964 (01)00884838085268(21)703724 (01)00884838085268(21)703541 (01)00884838085268(21)703815 (01)00884838085251(21)991 (01)00884838085251(21)1013 (01)00884838085251(21)943 (01)00884838085251(21)1025 (01)00884838085282(21)417 (01)00884838054202(21)511 (01)00884838085268(21)559 (01)00884838085268(21)560 (01)00884838085268(21)340 (01)00884838085350(21)45 (01)00884838085268(21)703560 (01)00884838054202(21)810 (01)00884838085350(21)272 (01)00884838085350(21)256 (01)00884838054189(21)364 (01)00884838054189(21)164 (01)00884838054189(21)1250 (01)00884838059030(21)226 (01)00884838059030(21)860 (01)00884838054189(21)1105 (01)00884838085251(21)464 (01)00884838085268(21)703523 (01)00884838085268(21)703524 (01)00884838085268(21)1418 (01)00884838085268(21)1209 (01)00884838085268(21)1427 (01)00884838085268(21)703537 (01)00884838085268(21)703538 (01)00884838085268(21)703517 (01)00884838085268(21)1426 (01)00884838054202(21)305 (01)00884838054196(21)139 (01)00884838085251(21)1003 (01)00884838085251(21)207 (01)00884838085251(21)209 (01)00884838085251(21)211 (01)00884838085251(21)447 (01)00884838085268(21)232 (01)00884838054189(21)412 (01)00884838054202(21)1547 (01)00884838054219(21)59 (01)00884838054240(21)92 (01)00884838085251(21)1005 (01)00884838085251(21)220 (01)00884838085251(21)799 (01)00884838099258(21)221 (01)00884838099227(21)62 (01)00884838099241(21)201 (01)00884838099241(21)185 (01)00884838099258(21)145 (01)00884838099258(21)257 (01)00884838099258(21)352 (01)00884838099258(21)353 (01)00884838099272(21)141 (01)00884838099258(21)368 (01)00884838099241(21)132 (01)00884838099241(21)165 (01)00884838099241(21)117 (01)00884838099241(21)195 (01)00884838099258(21)555 (01)00884838099241(21)267 (01)00884838099241(21)268 (01)00884838099241(21)271 (01)00884838099241(21)270 (01)00884838099241(21)255 (01)00884838099241(21)224 (01)00884838099227(21)56 (01)00884838099241(21)287 (01)00884838099241(21)114 (01)00884838099258(21)67 (01)00884838099241(21)239 (01)00884838099241(21)249 (01)00884838099241(21)141 (01)00884838099258(21)783 (01)00884838099258(21)531 (01)00884838099227(21)28 (01)00884838099241(21)142 (01)00884838099241(21)235 (01)00884838099241(21)227 (01)00884838085268(21)703959 (01)00884838099258(21)600 (01)00884838099241(21)348 (01)00884838099241(21)176 (01)00884838099258(21)634 (01)00884838099241(21)321 (01)00884838099241(21)123 (01)00884838099227(21)32 (01)00884838099227(21)72 (01)00884838099203(21)8 (01)00884838099227(21)64 (01)00884838099234(21)46 (01)00884838099241(21)91 (01)00884838099227(21)2 (01)00884838099265(21)4 (01)00884838099241(21)95 (01)00884838099241(21)283 (01)00884838099234(21)21 (01)00884838099234(21)22 (01)00884838099241(21)295 (01)00884838099272(21)140 (01)00884838099227(21)5 (01)00884838099227(21)6 (01)00884838099227(21)7 (01)00884838099227(21)74 (01)00884838099241(21)218 (01)00884838099265(21)38 (01)00884838099227(21)40 (01)00884838099227(21)15 (01)00884838099272(21)166 (01)00884838099227(21)47 (01)00884838099227(21)48 (01)00884838099227(21)13 (01)00884838099241(21)140 (01)00884838099241(21)9 (01)00884838099241(21)49 (01)00884838099241(21)20 (01)00884838099241(21)157 (01)00884838099258(21)597 (01)00884838099258(21)810 (01)00884838099272(21)171 (01)00884838099258(21)413 (01)00884838099241(21)76 (01)00884838085268(21)304 (01)00884838099203(21)48 (01)00884838099210(21)27 (01)00884838099258(21)270 (01)00884838099227(21)52 (01)00884838099258(21)416 (01)00884838099203(21)45 (01)00884838099241(21)291 (01)00884838099258(21)535 (01)00884838099272(21)86 (01)00884838099258(21)576 (01)00884838099258(21)345 (01)00884838099258(21)198 (01)00884838099258(21)126 (01)00884838099258(21)457 (01)00884838099258(21)117 (01)00884838099258(21)241 (01)00884838099258(21)510 (01)00884838099241(21)86 (01)00884838099265(21)60 (01)00884838099265(21)76 (01)00884838099258(21)168 (01)00884838099272(21)157 (01)00884838099258(21)137 (01)00884838085268(21)893 (01)00884838099258(21)473 (01)00884838099265(21)77 (01)00884838099265(21)20 (01)00884838099258(21)397 (01)00884838099272(21)73 (01)00884838099241(21)107 (01)00884838099258(21)392 (01)00884838099234(21)30 (01)00884838099272(21)137 (01)00884838099203(21)11 (01)00884838099265(21)67 (01)00884838099272(21)210 (01)00884838099265(21)64 (01)00884838099272(21)77 (01)00884838099258(21)142 (01)00884838099265(21)80 (01)00884838099241(21)166 (01)00884838099265(21)70 (01)00884838099272(21)28 (01)00884838099265(21)75 (01)00884838099272(21)192 (01)00884838099272(21)193 (01)00884838099234(21)27 (01)00884838099265(21)57 (01)00884838099258(21)692 (01)00884838099272(21)110 (01)00884838099265(21)34 (01)00884838099272(21)91 (01)00884838099241(21)200 (01)00884838099272(21)43 (01)00884838099272(21)161 (01)00884838099265(21)69 (01)00884838099265(21)61 (01)00884838099265(21)63 (01)00884838099258(21)157 (01)00884838099241(21)184 (01)00884838099265(21)13 (01)00884838099272(21)46 (01)00884838099265(21)59 (01)00884838099265(21)117 (01)00884838099258(21)197 (01)00884838099265(21)58 (01)00884838099265(21)18 (01)00884838099265(21)92 (01)00884838099258(21)144 (01)00884838099258(21)212 (01)00884838099265(21)42 (01)00884838099272(21)283 (01)00884838099241(21)82 (01)00884838085268(21)703810 (01)00884838099258(21)375 (01)00884838099241(21)170 (01)00884838099241(21)4 (01)00884838099272(21)118 (01)00884838099241(21)1534 (01)00884838099241(21)118 (01)00884838099258(21)358 (01)00884838099258(21)508 (01)00884838099258(21)499 (01)00884838099258(21)530 (01)00884838099272(21)203 (01)00884838099210(21)37 (01)00884838099258(21)549 (01)00884838099203(21)15 (01)00884838099258(21)348 (01)00884838085268(21)703461 (01)00884838099241(21)263 (01)00884838099241(21)263 (01)00884838099241(21)159 (01)00884838099241(21)160 (01)00884838099258(21)211 (01)00884838099258(21)474 (01)00884838099227(21)35 (01)00884838099241(21)309 (01)00884838099241(21)315 (01)00884838099241(21)110 (01)00884838099258(21)243 (01)00884838099258(21)610 (01)00884838099258(21)542 (01)00884838099241(21)305 (01)00884838099272(21)233 (01)00884838099272(21)88 (01)00884838099258(21)203 (01)00884838099258(21)604 (01)00884838099241(21)179 (01)00884838099210(21)73 (01)00884838099258(21)698 (01)00884838099258(21)703 (01)00884838099258(21)743 (01)00884838099234(21)65 (01)00884838099241(21)276 (01)00884838099210(21)87 (01)00884838099258(21)751 (01)00884838099272(21)276 (01)00884838099258(21)705 (01)00884838099258(21)701 (01)00884838099265(21)8 (01)00884838099272(21)21 (01)00884838099258(21)537 (01)00884838099258(21)779 (01)00884838099258(21)629 (01)00884838099258(21)631 (01)00884838099272(21)168 (01)00884838099258(21)423 (01)00884838099258(21)456 (01)00884838099258(21)302 (01)00884838099258(21)305 (01)00884838099258(21)310 (01)00884838099241(21)264 (01)00884838099258(21)249 (01)00884838099258(21)386 (01)00884838099258(21)506 (01)00884838099258(21)259 (01)00884838099258(21)248 (01)00884838099203(21)42 (01)00884838099241(21)302 (01)00884838099227(21)8 (01)00884838099227(21)9 (01)00884838099265(21)31 (01)00884838099265(21)87 (01)00884838099272(21)207 (01)00884838099265(21)140 (01)00884838099258(21)551 (01)00884838099203(21)23 (01)00884838099203(21)26 (01)00884838099272(21)178 (01)00884838099272(21)179 (01)00884838099241(21)115; Serial Numbers: 839 131 577 81 271 879 1783 1447 282 512 360 370 584 239 425 232 591 580 39 198 589 362 954 10 214 183 526 441 121 80 326 518 404 252 453 472 465 129 174 553 517 400 342 582 196 1391 485 262 277 120 703826 38 36 139 1199 940 945 947 703680 964 703724 703541 703815 991 1013 943 1025 417 511 559 560 340 45 703560 810 272 256 364 164 1250 226 860 1105 464 703523 703524 1418 1209 1427 703537 703538 703517 1426 395 139 1003 207 209 211 447 232 412 1547 59 92 1005 220 799 221 62 201 185 145 257 352 353 141 368 132 165 117 195 555 267 268 271 270 255 224 56 287 114 67 239 249 141 783 531 28 142 235 227 703959 600 348 176 634 321 123 32 72 8 64 46 91 2 4 95 283 21 22 295 140 5 6 7 74 218 38 40 15 166 47 48 13 140 9 49 20 157 597 810 171 413 76 304 48 27 270 52 416 45 291 535 86 576 345 198 126 457 117 241 510 86 60 76 168 157 137 893 473 77 20 397 73 107 392 30 137 11 67 210 64 77 142 80 166 70 28 75 192 193 27 57 692 110 34 91 200 43 161 69 61 63 157 184 13 46 59 117 197 58 18 92 144 212 42 283 82 703810 375 170 4 118 1534 118 358 508 499 530 203 37 549 15 348 703461 263 263 159 160 211 474 35 309 315 110 243 610 542 305 233 88 203 604 179 73 698 703 743 65 276 87 751 276 705 701 8 21 537 779 629 631 168 423 456 302 305 310 264 249 386 506 259 248 42 302 8 9 31 87 207 140 551 23 26 178 179 115
Devices named in this recall
21 devices in the FDA device registry whose identifier appears in the recall's code information.
- Allura Xper FD Allura Xper FD20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System25 recalls2 deaths3,984 reports
- Allura Xper FD Allura Xper FD20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls2 deaths3,614 reports
- Azurion Azurion 7 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System32 recalls2 deaths2,308 reports
- Azurion Azurion 7 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System28 recalls1 death2,176 reports
- Allura Xper FD Allura Xper FD10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls2 deaths2,015 reports
- Allura Xper FD Allura Xper FD10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls2 deaths1,837 reports
- Azurion Azurion 7 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls2 deaths1,087 reports
- Azurion Azurion 7 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls2 deaths835 reports
- Azurion Azurion 7 B20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls1 death634 reports
- Azurion Azurion 7 B20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls594 reports
- Allura Xper FD Allura Xper FD10/10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System19 recalls603 reports
- Azurion Azurion 5 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls201 reports
- Azurion Azurion 7 B12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls232 reports
- Allura Xper FD Allura Xper FD20 OR TablePhilips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System18 recalls411 reports
- Azurion Azurion 7 B12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System21 recalls222 reports
- Azurion Azurion 5 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System21 recalls1 death197 reports
- Azurion Azurion 3 M15Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls2 deaths205 reports
- Azurion Azurion 3 M15Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System21 recalls104 reports
- Allura Xper FD Allura Xper FD20/10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System19 recalls172 reports
- Azurion Azurion 3 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls121 reports
- Azurion Azurion 3 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls73 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page