Philips AneurysmFlow. Model Number: 001015 recallZ-1805-2026 · Philips Medical Systems Nederland B.V. · initiated Mar 19, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions.

FDA root cause category: Labeling design

What the company is doing

Philips notified consignees on about 03/19/2026 via "URGENT Medical Device Correction" letter. Consignees were instructed NOT to use the MAFA ration for clinical decision making as indicated in the IFU and that AneurysmFlow may continue to be used in accordance with the Instructions for Use (IFU). Additionally, consignees were instructed to circulate the URGENT Medical Device Correction letter to all users so that they are aware of the issue and keep this URGENT Medical Device Correction letter with the documentation of the system until Philips corrects your system. Ensure the letter is in a place likely to be seen/viewed. Lastly, consignees were instructed to complete and return the response form. In case the Philips interventional X-ray system on which AneurysmFlow is installed has been transferred to another organization, consignees were requested to send a copy of the URGENT Medical Device Correction letter to that organization and inform Philips about this transfer through their local Philips representative. Philips is developing a software update to remove the MAFA ratio from AneurysmFlow. Philips expects to release this software update in December 2026. Local Philips representatives will contact consignees to schedule implementation of this software update once available.

Recall details

Recall number
Z-1805-2026
Status
Open, Classified
Initiated by the company
Mar 19, 2026
Posted by the FDA
Apr 10, 2026
Recalling firm
Philips Medical Systems Nederland B.V.
Quantity
1,033 units
Product code
LLZ
FDA submissions
K160455
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bnis insted Us Instas instes, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Egypt, Estonia, Germany, Greece, Indonesia, Iraq, Italy, Japan, Kuwait, Latvia, Macao, Maldives, Mongolia, Morocco, New Zealand, Nicaragua, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovenia, South Africa, Sri Lanka, Switzerland, TURKEY, Ukraine, Utd.Arab.Emir., Uzbekistan, Vietnam.

Product description

Philips AneurysmFlow. Model Number: 001015. AneurysmFlow is a software medical device (Interventional Tool) intended to be used in combination with a Philips interventional X-ray system and 3DRA data.

Code information (models, lots, serial numbers)

Show the full code information (28k characters)

Model Number: 001015. Software Version Number R1.x. UDI Number: (01)00884838099272(21)153, (01)00884838099272(21)67, (01)00884838099272(21)675, (01)00884838085367(21)178, (01)00884838099272(21)112, (01)00884838085367(21)335, (01)00884838116801(21)345, (01)00884838085367(21)147, (01)00884838085367(21)205, (01)00884838099272(21)113, (01)00884838099272(21)121, (01)00884838116801(21)24, (01)00884838085367(21)170, (01)00884838099258(21)1150, (01)00884838116801(21)17, (01)00884838116801(21)25, (01)00884838059061(21)639, (01)00884838085268(21)856, (01)00884838116801(21)180, (01)00884838116801(21)227, (01)00884838099272(21)239, (01)00884838099272(21)105, (01)00884838116801(21)235, (01)00884838099272(21)678, (01)00884838054202(21)373, (01)00884838059115(21)290, (01)00884838085367(21)564, (01)00884838085268(21)716, (01)00884838085367(21)440, (01)00884838116801(21)226, (01)00884838099272(21)197, (01)00884838099272(21)380, (01)00884838059115(21)274, (01)00884838085367(21)247, (01)00884838085367(21)225, (01)00884838099272(21)592, (01)00884838099272(21)593, (01)00884838099272(21)319, (01)00884838099272(21)339, (01)00884838099272(21)182, (01)00884838099258(21)1234, (01)00884838085367(21)64, (01)00884838116801(21)20, (01)00884838099272(21)272, (01)00884838099272(21)680, (01)00884838085367(21)153, (01)00884838099272(21)350, (01)00884838099272(21)280, (01)00884838116801(21)208, (01)00884838116801(21)205, (01)00884838116801(21)36, (01)00884838085367(21)85, (01)00884838085367(21)89, (01)00884838085367(21)320, (01)00884838099272(21)144, (01)00884838085367(21)109, (01)00884838085367(21)407, (01)00884838099272(21)779, (01)00884838116801(21)81, (01)00884838085367(21)505, (01)00884838059115(21)277, (01)00884838099272(21)559, (01)00884838085367(21)443, (01)00884838099272(21)98, (01)00884838099272(21)343, (01)00884838116801(21)147, (01)00884838116801(21)148, (01)00884838085367(21)104, (01)00884838099272(21)876, (01)00884838059115(21)168, (01)00884838099272(21)1021, (01)00884838099272(21)244, (01)00884838116801(21)31, (01)00884838099272(21)332, (01)00884838099272(21)781, (01)00884838099272(21)782, (01)00884838085367(21)524, (01)00884838099272(21)534, (01)00884838099272(21)129, (01)00884838099272(21)442, (01)00884838085367(21)400, (01)00884838059061(21)575, (01)00884838099272(21)189, (01)00884838099272(21)190, (01)00884838085367(21)510, (01)00884838116801(21)252, (01)00884838085367(21)110, (01)00884838099272(21)208, (01)00884838085367(21)73, (01)00884838085367(21)288, (01)00884838085268(21)281, (01)00884838116801(21)21, (01)00884838099272(21)400, (01)00884838085367(21)406, (01)00884838099272(21)326, (01)00884838085367(21)528, (01)00884838099272(21)323, (01)00884838116801(21)118, (01)00884838116801(21)50, (01)00884838085268(21)852, (01)00884838099258(21)101, (01)00884838085367(21)457, (01)00884838085367(21)470, (01)00884838085367(21)472, (01)00884838099272(21)778, (01)00884838116801(21)39, (01)00884838085268(21)704281, (01)00884838085367(21)319, (01)00884838085367(21)520, (01)00884838085367(21)490, (01)00884838085268(21)703418, (01)00884838116801(21)48, (01)00884838099272(21)482, (01)00884838116801(21)221, (01)00884838085367(21)150, (01)00884838099258(21)46, (01)00884838085367(21)172, (01)00884838085268(21)29, (01)00884838059115(21)61, (01)00884838099272(21)793, (01)00884838116801(21)68, (01)00884838085367(21)117, (01)00884838085268(21)930, (01)00884838099272(21)1017, (01)00884838085367(21)130, (01)00884838059115(21)264, (01)00884838099272(21)410, (01)00884838099272(21)508, (01)00884838059115(21)167, (01)00884838099272(21)419, (01)00884838085367(21)469, (01)00884838099272(21)472, (01)00884838059061(21)327, (01)00884838085268(21)21, (01)00884838116801(21)51, (01)00884838099272(21)383, (01)00884838085367(21)76, (01)00884838099272(21)59, (01)00884838099272(21)361, (01)00884838085367(21)477, (01)00884838099272(21)1002, (01)00884838085367(21)137, (01)00884838059061(21)580, (01)00884838116801(21)38, (01)00884838116801(21)22, (01)00884838099272(21)398, (01)00884838085367(21)107, (01)00884838116801(21)186, (01)00884838099272(21)75, (01)00884838054202(21)534, (01)00884838085268(21)255, (01)00884838116801(21)32, (01)00884838085367(21)282, (01)00884838099258(21)2557, (01)00884838099272(21)786, (01)00884838116801(21)276, (01)00884838099272(21)19, (01)00884838099272(21)1125, (01)00884838085367(21)487, (01)00884838099272(21)814, (01)00884838085367(21)213, (01)00884838099272(21)469, (01)00884838059115(21)172, (01)00884838099272(21)219, (01)00884838085367(21)82, (01)00884838099272(21)296, (01)00884838059061(21)607, (01)00884838099272(21)704, (01)00884838054226(21)102, (01)00884838059115(21)110, (01)00884838085367(21)485, (01)00884838085268(21)783, (01)00884838099258(21)2937, (01)00884838085367(21)276, (01)00884838085268(21)703565, (01)00884838116771(21)187, (01)00884838099272(21)693, (01)00884838099272(21)510, (01)00884838099272(21)882, (01)00884838099258(21)2852, (01)00884838099258(21)2931, (01)00884838099272(21)226, (01)00884838085367(21)559, (01)00884838085268(21)703844, (01)00884838085268(21)703893, (01)00884838099272(21)509, (01)00884838116801(21)289, (01)00884838116801(21)67, (01)00884838059115(21)332, (01)00884838099272(21)1050, (01)00884838099258(21)497, (01)00884838099272(21)201, (01)00884838099272(21)202, (01)00884838099272(21)303, (01)00884838099272(21)184, (01)00884838099272(21)250, (01)00884838054202(21)1914, (01)00884838099272(21)1139, (01)00884838059115(21)142, (01)00884838099272(21)368, (01)00884838099272(21)950, (01)00884838099272(21)829, (01)00884838054226(21)11, (01)00884838099272(21)249, (01)00884838085367(21)302, (01)00884838085367(21)314, (01)00884838059054(21)1144, (01)00884838099272(21)576, (01)00884838085367(21)88, (01)00884838085367(21)211, (01)00884838085367(21)468, (01)00884838085367(21)422, (01)00884838085268(21)704300, (01)00884838099272(21)665, (01)00884838099272(21)764, (01)00884838099272(21)595, (01)00884838085367(21)454, (01)00884838099258(21)799, (01)00884838099258(21)800, (01)00884838099272(21)721, (01)00884838085367(21)283, (01)00884838085367(21)231, (01)00884838085367(21)548, (01)00884838099258(21)2555, (01)00884838099272(21)1006, (01)00884838059122(21)35, (01)00884838099272(21)498, (01)00884838099272(21)60, (01)00884838085367(21)207, (01)00884838085367(21)167, (01)00884838099258(21)1976, (01)00884838099258(21)2025, (01)00884838085367(21)537, (01)00884838085268(21)703712, (01)00884838085367(21)438, (01)00884838059115(21)151, (01)00884838099272(21)698, (01)00884838099272(21)331, (01)00884838085367(21)536, (01)00884838099272(21)378, (01)00884838116771(21)160, (01)00884838085367(21)204, (01)00884838099258(21)765, (01)00884838099258(21)1470, (01)00884838099272(21)393, (01)00884838116801(21)352, (01)00884838085367(21)77, (01)00884838116801(21)300, (01)00884838099258(21)778, (01)00884838099258(21)2541, (01)00884838099272(21)718, (01)00884838099272(21)663, (01)00884838099272(21)195 (01)00884838099258(21)1338, (01)00884838116801(21)173, (01)00884838085367(21)236, (01)00884838085367(21)237, (01)00884838099272(21)775, (01)00884838099258(21)828, (01)00884838059115(21)232, (01)00884838085367(21)133, (01)00884838054202(21)1955, (01)00884838085268(21)703753, (01)00884838099272(21)282, (01)00884838116801(21)117, (01)00884838085367(21)280, (01)00884838059115(21)395, (01)00884838099258(21)2515, (01)00884838099272(21)785, (01)00884838059115(21)292, (01)00884838085367(21)307, (01)00884838099272(21)221, (01)00884838059115(21)380, (01)00884838099272(21)604, (01)00884838099272(21)415, (01)00884838099272(21)61, (01)00884838059115(21)183, (01)00884838085367(21)328, (01)00884838099258(21)720, (01)00884838099272(21)762, (01)00884838099272(21)1102, (01)00884838099272(21)306, (01)00884838085268(21)778, (01)00884838085268(21)548, (01)00884838099258(21)1581, (01)00884838099258(21)977, (01)00884838099272(21)83, (01)00884838099272(21)199, (01)00884838085268(21)1429, (01)00884838099258(21)675, (01)00884838116771(21)261, (01)00884838099272(21)898, (01)00884838085367(21)480, (01)00884838085367(21)53, (01)00884838085367(21)72, (01)00884838099272(21)899, (01)00884838085367(21)491, (01)00884838085367(21)423, (01)00884838099272(21)944, (01)00884838085367(21)509, (01)00884838099258(21)548, (01)00884838099258(21)3546, (01)00884838085367(21)246, (01)00884838059115(21)161, (01)00884838085367(21)614, (01)00884838099272(21)335, (01)00884838099272(21)1123, (01)00884838116801(21)34, (01)00884838099272(21)363, (01)00884838085367(21)376, (01)00884838054219(21)106, (01)00884838059115(21)372, (01)00884838085268(21)1239, (01)00884838054202(21)611, (01)00884838116801(21)122, (01)00884838085367(21)539, (01)00884838099258(21)771, (01)00884838059115(21)252, (01)00884838099272(21)626, (01)00884838099258(21)2592, (01)00884838099272(21)501, (01)00884838085367(21)140, (01)00884838085268(21)290, (01)00884838116801(21)374, (01)00884838085367(21)369, (01)00884838085367(21)508, (01)00884838085367(21)97, (01)00884838085268(21)464, (01)00884838099258(21)2850, (01)00884838054202(21)1540, (01)00884838085367(21)87, (01)00884838099272(21)449, (01)00884838116801(21)326, (01)00884838085367(21)303, (01)00884838085367(21)48, (01)00884838099272(21)974, (01)00884838085367(21)44, (01)00884838085367(21)301, (01)00884838085367(21)550, (01)00884838099272(21)344, (01)00884838099272(21)984, (01)00884838099272(21)487, (01)00884838059115(21)213, (01)00884838099272(21)686, (01)00884838099272(21)327, (01)00884838099272(21)424, (01)00884838099272(21)684, (01)00884838085367(21)478, (01)00884838099272(21)298, (01)00884838099272(21)1049, (01)00884838116801(21)207, (01)00884838054202(21)1893, (01)00884838099272(21)259, (01)00884838099258(21)2864, (01)00884838099272(21)140, (01)00884838059115(21)169, (01)00884838059061(21)295 (01)00884838099272(21)506, (01)00884838099258(21)3598, (01)00884838099258(21)2921, (01)00884838099258(21)2947, (01)00884838099258(21)2424, (01)00884838085268(21)703966, (01)00884838099258(21)2105, (01)00884838099258(21)2101, (01)00884838099258(21)3383, (01)00884838099272(21)263, (01)00884838099258(21)3416, (01)00884838099258(21)620, (01)00884838099258(21)2932, (01)00884838099272(21)490, (01)00884838059061(21)110, (01)00884838099272(21)633, (01)00884838099272(21)396, (01)00884838116771(21)868, (01)00884838099272(21)547, (01)00884838116801(21)311, (01)00884838085367(21)160, (01)00884838099272(21)631, (01)00884838116771(21)190, (01)00884838085268(21)704258, (01)00884838085268(21)949, (01)00884838099258(21)632, (01)00884838099258(21)1766, (01)00884838099258(21)2872, (01)00884838099272(21)514, (01)00884838085367(21)68, (01)00884838085268(21)760, (01)00884838099258(21)3570, (01)00884838085268(21)961, (01)00884838059115(21)192, (01)00884838116771(21)658, (01)00884838116801(21)57, (01)00884838099272(21)17, (01)00884838099272(21)422, (01)00884838099272(21)420, (01)00884838099272(21)247, (01)00884838099272(21)432, (01)00884838099272(21)539, (01)00884838085367(21)360, (01)00884838116801(21)169, (01)00884838099272(21)245, (01)00884838099272(21)911, (01)00884838099272(21)132, (01)00884838099272(21)133, (01)00884838099258(21)1814, (01)00884838099258(21)2134, (01)00884838099272(21)586, (01)00884838099272(21)622, (01)00884838099272(21)228, (01)00884838116801(21)44, (01)00884838099258(21)1051, (01)00884838099258(21)1052, (01)00884838099258(21)2277, (01)00884838099272(21)45, (01)00884838085367(21)179, (01)00884838085367(21)306, (01)00884838085367(21)182, (01)00884838099272(21)485, (01)00884838085268(21)493, (01)00884838085367(21)327, (01)00884838099258(21)2618, (01)00884838099272(21)448, (01)00884838099272(21)507, (01)00884838099258(21)2575, (01)00884838099272(21)97, (01)00884838054226(21)133, (01)00884838099272(21)497, (01)00884838099258(21)1700, (01)00884838099258(21)2987, (01)00884838099258(21)3212, (01)00884838099258(21)1815, (01)00884838085367(21)244, (01)00884838059115(21)170, (01)00884838099272(21)685, (01)00884838059115(21)77, (01)00884838116801(21)274, (01)00884838059115(21)273, (01)00884838099258(21)2812, (01)00884838099258(21)2835, (01)00884838099272(21)1076, (01)00884838059115(21)229, (01)00884838085268(21)703935, (01)00884838099258(21)2103, (01)00884838099272(21)543, (01)00884838099272(21)66, (01)00884838085367(21)339, (01)00884838099272(21)542, (01)00884838099272(21)550, (01)00884838099272(21)551, (01)00884838099272(21)484, (01)00884838099272(21)504, (01)00884838099272(21)18, (01)00884838099258(21)545, (01)00884838116771(21)751, (01)00884838099258(21)1030, (01)00884838099272(21)1132, (01)00884838059054(21)1090, (01)00884838099272(21)752, (01)00884838059115(21)248, (01)00884838116801(21)113, (01)00884838099272(21)408, (01)00884838116771(21)562, (01)00884838085268(21)571, (01)00884838099258(21)520, (01)00884838116771(21)702, (01)00884838099272(21)238, (01)00884838099272(21)299, (01)00884838085367(21)318, (01)00884838085367(21)518, (01)00884838099258(21)3533, (01)00884838099272(21)404, (01)00884838059115(21)258, (01)00884838059115(21)289, (01)00884838085367(21)185, (01)00884838085268(21)682, (01)00884838116801(21)292, (01)00884838099258(21)358, (01)00884838099272(21)84, (01)00884838099258(21)2224, (01)00884838085367(21)50, (01)00884838059115(21)269, (01)00884838085367(21)521, (01)00884838099272(21)924, (01)00884838099272(21)224, (01)00884838085367(21)79, (01)00884838099258(21)1984, (01)00884838059115(21)141, (01)00884838116801(21)284, (01)00884838116801(21)285, (01)00884838059115(21)391, (01)00884838116771(21)504, (01)00884838085367(21)529, (01)00884838059122(21)38, (01)00884838116801(21)193, (01)00884838099258(21)2102, (01)00884838099272(21)261, (01)00884838085367(21)394, (01)00884838085268(21)889, (01)00884838099272(21)956, (01)00884838099272(21)1077, (01)00884838116771(21)51, (01)00884838099272(21)320, (01)00884838116801(21)297, (01)00884838099258(21)635, (01)00884838099272(21)710, (01)00884838085367(21)292, (01)00884838116801(21)347, (01)00884838085367(21)497, (01)00884838116801(21)321, (01)00884838054202(21)1125, (01)00884838054226(21)135, (01)00884838059115(21)281, (01)00884838099272(21)1072, (01)00884838099258(21)1272, (01)00884838085367(21)35, (01)00884838059115(21)246, (01)00884838099272(21)235, (01)00884838099272(21)727, (01)00884838099258(21)3412, (01)00884838059115(21)256, (01)00884838099258(21)3204, (01)00884838099272(21)621, (01)00884838085268(21)703346, (01)00884838099258(21)2719, (01)00884838085268(21)521, (01)00884838059115(21)233, (01)00884838059115(21)348, (01)00884838099272(21)765, (01)00884838099272(21)194, (01)00884838085268(21)586, (01)00884838085268(21)498, (01)00884838059115(21)204, (01)00884838099272(21)451, (01)00884838059122(21)37, (01)00884838116801(21)269, (01)00884838116801(21)102, (01)00884838085268(21)967, (01)00884838099258(21)3481, (01)00884838099272(21)275, (01)00884838099272(21)352, (01)00884838099272(21)630, (01)00884838085367(21)486, (01)00884838116801(21)132, (01)00884838085367(21)543, (01)00884838059115(21)206, (01)00884838085367(21)453, (01)00884838099272(21)1047, (01)00884838099258(21)2230, (01)00884838099272(21)349, (01)00884838085367(21)174, (01)00884838085367(21)483, (01)00884838085268(21)642, (01)00884838099272(21)854, (01)00884838099272(21)888, (01)00884838116801(21)189, (01)00884838085367(21)425, (01)00884838085367(21)181, (01)00884838054202(21)1411, (01)00884838085367(21)141, (01)00884838059061(21)53, (01)00884838099272(21)670, (01)00884838099272(21)739, (01)00884838085367(21)471, (01)00884838099272(21)1115, (01)00884838099258(21)3716, (01)00884838099258(21)2061, (01)00884838099258(21)1927, (01)00884838085268(21)704058, (01)00884838085367(21)602, (01)00884838099258(21)3436, (01)00884838099272(21)574, (01)00884838099272(21)741, (01)00884838099258(21)1631, (01)00884838085367(21)593, (01)00884838085268(21)704073, (01)00884838085268(21)704291, (01)00884838099272(21)1137, (01)00884838085367(21)589, (01)00884838099272(21)632, (01)00884838099258(21)1946, (01)00884838099258(21)1990, (01)00884838085268(21)704293, (01)00884838099258(21)2216, (01)00884838099272(21)1148, (01)00884838099258(21)2491, (01)00884838099258(21)2492, (01)00884838099258(21)2169, (01)00884838099272(21)662, (01)00884838099272(21)825, (01)00884838099272(21)1075, (01)00884838099272(21)744, (01)00884838099258(21)3683, (01)00884838099258(21)2337, (01)00884838099258(21)2338, (01)00884838099258(21)2823, (01)00884838099258(21)2825, (01)00884838099258(21)2109, (01)00884838099272(21)549, (01)00884838099272(21)1023, (01)00884838085367(21)612, (01)00884838099258(21)3689, (01)00884838099258(21)772, (01)00884838085268(21)703975, (01)00884838085367(21)574, (01)00884838099272(21)1135, (01)00884838099272(21)751, (01)00884838099272(21)1036, (01)00884838085268(21)703803, (01)00884838099258(21)3494, (01)00884838059115(21)358, (01)00884838099258(21)1746, (01)00884838085367(21)616, (01)00884838099258(21)2815, (01)00884838099272(21)798, (01)00884838099258(21)2091, (01)00884838099272(21)886, (01)00884838059115(21)418, (01)00884838099272(21)1128, (01)00884838099258(21)2551, (01)00884838099272(21)979, (01)00884838054202(21)1761, (01)00884838099258(21)2203, (01)00884838110571(21)35, (01)00884838110571(21)8, (01)00884838099272(21)669, (01)00884838054226(21)162, (01)00884838116801(21)182, (01)00884838059115(21)388, (01)00884838085268(21)1034, (01)00884838099272(21)29, (01)00884838085367(21)600, (01)00884838085268(21)704159, (01)00884838085268(21)704162, (01)00884838099258(21)2251, (01)00884838099272(21)817, (01)00884838085367(21)228, (01)00884838085367(21)380, (01)00884838099272(21)743, (01)00884838085268(21)398, (01)00884838085367(21)397, (01)00884838099258(21)2015, (01)00884838085367(21)475, (01)00884838099272(21)131, (01)00884838099258(21)209, (01)00884838085367(21)180, (01)00884838085268(21)1255, (01)00884838099272(21)563, (01)00884838059061(21)378, (01)00884838085268(21)1045, (01)00884838099258(21)3734, (01)00884838099234(21)691, (01)00884838099272(21)945, (01)00884838085367(21)268, (01)00884838085367(21)269, (01)00884838085367(21)414, (01)00884838116771(21)245, (01)00884838099258(21)1579, (01)00884838099272(21)50, (01)00884838054226(21)116, (01)00884838059115(21)354, (01)00884838059115(21)355, (01)00884838059115(21)159, (01)00884838059061(21)58, (01)00884838085367(21)349, (01)00884838085268(21)941, (01)00884838099258(21)528, (01)00884838099272(21)198, (01)00884838099272(21)168 (01)00884838085268(21)704150, (01)00884838099258(21)2481, (01)00884838099272(21)580, (01)00884838099272(21)740, (01)00884838099258(21)2343, (01)00884838099258(21)3426, (01)00884838085268(21)1004, (01)00884838099272(21)1022, (01)00884838099272(21)207, (01)00884838054202(21)1703, (01)00884838085367(21)219, (01)00884838099272(21)598, (01)00884838085367(21)60, (01)00884838085268(21)546, (01)00884838099272(21)54, (01)00884838099272(21)849, (01)00884838099272(21)636, (01)00884838085367(21)398, (01)00884838116801(21)123, (01)00884838099272(21)82, (01)00884838059054(21)2406, (01)00884838099272(21)417, (01)00884838116801(21)23, (01)00884838116801(21)338, (01)00884838085367(21)96, (01)00884838116801(21)181, (01)00884838054226(21)146, (01)00884838099272(21)152, (01)00884838099272(21)385, (01)00884838099258(21)3340, (01)00884838099258(21)774, (01)00884838099258(21)3266, (01)00884838099272(21)305, (01)00884838099258(21)2164, (01)00884838099272(21)832, (01)00884838099272(21)1059, (01)00884838085268(21)72, (01)00884838099272(21)1118, (01)00884838099272(21)867, (01)00884838116801(21)130, (01)00884838054202(21)113, (01)00884838085268(21)764, (01)00884838085268(21)1188, (01)00884838099272(21)627, (01)00884838059115(21)343, (01)00884838099272(21)872, (01)00884838099272(21)1080, (01)00884838054226(21)154, (01)00884838085367(21)346, (01)00884838099272(21)6, (01)00884838116771(21)596, (01)00884838116801(21)279, (01)00884838116771(21)677, (01)00884838099272(21)1035, (01)00884838099258(21)3321, (01)00884838099272(21)791, (01)00884838059115(21)291, (01)00884838085367(21)49, (01)00884838085367(21)286, (01)00884838085268(21)775, (01)00884838085268(21)1139 (01)00884838059115(21)104, (01)00884838059115(21)105, (01)00884838099272(21)772, (01)00884838059115(21)145, (01)00884838099272(21)1093, (01)00884838099272(21)7, (01)00884838116801(21)128, (01)00884838116801(21)86, (01)00884838099258(21)1951, (01)00884838116801(21)94, (01)00884838099272(21)664, (01)00884838116771(21)58, (01)00884838099258(21)2232, (01)00884838059115(21)103, (01)00884838099272(21)52, (01)00884838099272(21)53, (01)00884838099258(21)2214, (01)00884838116771(21)127, (01)00884838054226(21)66, (01)00884838085367(21)102, (01)00884838099272(21)890 (01)00884838116771(21)360, (01)00884838085367(21)124, (01)00884838099258(21)2455, (01)00884838059115(21)188, (01)00884838116801(21)82, (01)00884838116801(21)268, (01)00884838085367(21)362, (01)00884838059115(21)350, (01)00884838116771(21)52, (01)00884838059115(21)422, (01)00884838099272(21)729, (01)00884838116801(21)277, (01)00884838099272(21)478, (01)00884838099258(21)1852, (01)00884838099272(21)650, (01)00884838099272(21)26, (01)00884838099258(21)724, (01)00884838099272(21)883, (01)00884838085367(21)145, (01)00884838099272(21)453, (01)00884838099272(21)889, (01)00884838099258(21)1431, (01)00884838085367(21)299, (01)00884838099258(21)3749, (01)00884838099258(21)2006, (01)00884838085268(21)1336, (01)00884838099258(21)2440, (01)00884838085268(21)703339, (01)00884838085268(21)1091, (01)00884838099272(21)673, (01)00884838054226(21)50, (01)00884838099272(21)844, (01)00884838054202(21)1062, (01)00884838099258(21)2342, (01)00884838085268(21)1024, (01)00884838099234(21)688, (01)00884838099258(21)2909, (01)00884838099272(21)276, (01)00884838085268(21)1092, (01)00884838085268(21)703484, (01)00884838054202(21)1557, (01)00884838099234(21)637, (01)00884838099272(21)69, (01)00884838059115(21)43, (01)00884838085367(21)187, (01)00884838085367(21)183, (01)00884838116801(21)111, (01)00884838116801(21)172, (01)00884838116801(21)112, (01)00884838099272(21)528, (01)00884838099272(21)660, (01)00884838099272(21)74, (01)00884838099272(21)855, (01)00884838059115(21)237, (01)00884838059115(21)230, (01)00884838085367(21)45, (01)00884838099272(21)286, (01)00884838099272(21)317, (01)00884838099258(21)556, (01)00884838099272(21)42, (01)00884838085367(21)331, (01)00884838116801(21)259, (01)00884838099272(21)229, (01)00884838059115(21)247, (01)00884838099272(21)908, (01)00884838085367(21)106, (01)00884838116801(21)103, (01)00884838059115(21)116, (01)00884838059115(21)242, (01)00884838099258(21)280, (01)00884838099258(21)646, (01)00884838059115(21)29, (01)00884838116801(21)52, (01)00884838085367(21)112, (01)00884838099258(21)3185, (01)00884838085367(21)501, (01)00884838099258(21)917, (01)00884838085367(21)393, (01)00884838085367(21)116, (01)00884838059115(21)276, (01)00884838099272(21)540, (01)00884838085268(21)614, (01)00884838099258(21)568, (01)00884838099272(21)73, (01)00884838099272(21)405, (01)00884838085367(21)139, (01)00884838099272(21)583, (01)00884838099272(21)784, (01)00884838099272(21)525, (01)00884838099272(21)312, (01)00884838085367(21)122, (01)00884838116801(21)80, (01)00884838116801(21)175, (01)00884838099272(21)658, (01)00884838099272(21)216, (01)00884838099272(21)95, (01)00884838059061(21)542, (01)00884838099272(21)39, (01)00884838099272(21)856, (01)00884838116801(21)305, (01)00884838099272(21)853, (01)00884838099272(21)653, (01)00884838099258(21)2642, (01)00884838099272(21)502, (01)00884838116801(21)92, (01)00884838099272(21)70, (01)00884838099272(21)128, (01)00884838099272(21)287, (01)00884838099272(21)89, (01)00884838099272(21)807, (01)00884838099272(21)265, (01)00884838059115(21)179, (01)00884838099272(21)304, (01)00884838116801(21)155, (01)00884838085367(21)390, (01)00884838099272(21)110, (01)00884838059115(21)303, (01)00884838085367(21)300, (01)00884838085367(21)67, (01)00884838099272(21)841, (01)00884838059061(21)182, (01)00884838085367(21)61, (01)00884838085367(21)157, (01)00884838099272(21)234, (01)00884838099272(21)885, (01)00884838085367(21)154, (01)00884838099272(21)64, (01)00884838085367(21)100, (01)00884838099272(21)916, (01)00884838085367(21)217, (01)00884838099272(21)192, (01)00884838099272(21)193, (01)00884838059115(21)245, (01)00884838099272(21)618, (01)00884838099272(21)32, (01)00884838099272(21)206, (01)00884838099272(21)904, (01)00884838059115(21)41, (01)00884838099272(21)49, (01)00884838099272(21)541, (01)00884838116801(21)85, (01)00884838059115(21)234, (01)00884838099272(21)646, (01)00884838099272(21)870, (01)00884838099272(21)258, (01)00884838099272(21)172, (01)00884838099272(21)529, (01)00884838059115(21)208, (01)00884838059061(21)563, (01)00884838059115(21)193, (01)00884838099272(21)524, (01)00884838116801(21)142, (01)00884838085367(21)379, (01)00884838085268(21)703357, (01)00884838099258(21)197, (01)00884838099272(21)33, (01)00884838116801(21)105, (01)00884838099272(21)656, (01)00884838059115(21)266, (01)00884838099272(21)290, (01)00884838085367(21)202, (01)00884838116771(21)20, (01)00884838085367(21)226, (01)00884838059061(21)436, (01)00884838054219(21)44, (01)00884838085367(21)464, (01)00884838116801(21)108, (01)00884838116801(21)53, (01)00884838116801(21)144, (01)00884838059061(21)406, (01)00884838099272(21)163, (01)00884838099272(21)852, (01)00884838116801(21)198, (01)00884838116801(21)124, (01)00884838099272(21)62, (01)00884838085367(21)274, (01)00884838099272(21)1034, (01)00884838085367(21)163, (01)00884838085367(21)186, (01)00884838059115(21)195, (01)00884838099272(21)806, (01)00884838099272(21)210, (01)00884838085367(21)260, (01)00884838085367(21)502, (01)00884838059115(21)330, (01)00884838059061(21)395, (01)00884838099258(21)1582, (01)00884838085367(21)498, (01)00884838059061(21)458, (01)00884838116801(21)261, (01)00884838099272(21)847, (01)00884838085367(21)156, (01)00884838116801(21)109, (01)00884838099272(21)587, (01)00884838099272(21)154, (01)00884838116801(21)83, (01)00884838099272(21)186 (01)00884838116801(21)171, (01)00884838099258(21)3451, (01)00884838054240(21)173, (01)00884838059115(21)187, (01)00884838099272(21)425, (01)00884838116801(21)76, (01)00884838085367(21)208, (01)00884838099258(21)1400, (01)00884838099272(21)500, (01)00884838059115(21)79, (01)00884838116801(21)106, (01)00884838054240(21)64, (01)00884838099272(21)537, (01)00884838085367(21)430, (01)00884838085367(21)161, (01)00884838099258(21)910, (01)00884838116801(21)49, (01)00884838099272(21)375, (01)00884838116801(21)77, (01)00884838099272(21)567, (01)00884838099272(21)479, (01)00884838059115(21)361, (01)00884838085367(21)155, (01)00884838099272(21)31, (01)00884838085268(21)294, (01)00884838099272(21)302, (01)00884838054226(21)170, (01)00884838059061(21)535, (01)00884838099272(21)406, (01)00884838099272(21)857, (01)00884838054219(21)89, (01)00884838059115(21)284, (01)00884838059115(21)21, (01)00884838085367(21)191, (01)00884838099272(21)770, (01)00884838116801(21)78, (01)00884838099272(21)56, (01)00884838059061(21)344, (01)00884838099258(21)718, (01)00884838059115(21)64, (01)00884838085367(21)94, (01)00884838099272(21)155, (01)00884838099272(21)177, (01)00884838085367(21)66, (01)00884838099272(21)96, (01)00884838099272(21)28, (01)00884838099272(21)560, (01)00884838099272(21)433, (01)00884838085367(21)108, (01)00884838085367(21)200, (01)00884838085367(21)201, (01)00884838099272(21)324, (01)00884838085367(21)189, (01)00884838116801(21)59, (01)00884838059122(21)2, (01)00884838059061(21)522, (01)00884838099272(21)588, (01)00884838085367(21)311, (01)00884838059115(21)272, (01)00884838059115(21)210, (01)00884838116801(21)104, (01)00884838059115(21)243, (01)00884838099272(21)79, (01)00884838116801(21)107, (01)00884838059061(21)394, (01)00884838099272(21)236, (01)00884838059115(21)20, (01)00884838099272(21)863, (01)00884838116801(21)152, (01)00884838099272(21)165, (01)00884838099272(21)402, (01)00884838099272(21)652, (01)00884838085367(21)297, (01)00884838116801(21)153, (01)00884838099272(21)241, (01)00884838059115(21)218. AneurysmFlow is a software medical device intended to be used in combination with the Philips interventional X-ray system (Allura and Azurion systems) and 3DRA data. The Allura and Azurion systems are not impacted by this recall. Since AneurysmFlow is always used with Allura and Azurion systems, below is information of those systems that are currently being used with the affected software to help with identification of the affected devices. The Allura and Azurion model numbers where AneurysmFlow is used are the following: 722006, 722008, 722012, 722013, 722015, 722028, 722068, 722029, 722035, 722038, 722079, 722058, 722226, 722236, 722059, 722224, 722234, 722228, 722282.