t:slim X2 Insulin Pump with Control-IQ 1009637Recalls & reported problems

Tandem Diabetes Care, Inc. · t:slim X2 Insulin Pump with Control-IQ, 7.6.x · Class 2

All models of t:slim X2 Insulin Pump (45) · t:slim X2 Insulin Pump with Control-IQ (11)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This is a Tandem t:slim X2 insulin pump with Control-IQ. Its reports mostly describe pump stoppages, communication problems, battery drain and repeated occlusion alarms, some describing high blood glucose afterward. There are 84,241 reports, including 4 deaths and 8,883 injuries. One recall, started in July 2025, is open: it covers speakers that may fail and stop insulin delivery.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
83,767
Deaths reported
4
Injuries reported
8,868
Recalls naming it
1

Summary

The FDA's MAUDE database contains 83,767 adverse event reports that list this exact device (DI 00389152000077), received between Aug 22, 2022 and Sep 30, 2026 — classified as 4 deaths, 8,868 injuries, 74,895 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (35,390 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 2,220 reportsJan 2024: 2,416 reportsFeb 2024: 2,084 reportsMar 2024: 3,022 reportsApr 2024: 3,024 reportsMay 2024: 2,436 reportsJun 2024: 2,359 reportsJul 2024: 3,060 reportsAug 2024: 2,732 reportsSep 2024: 2,985 reportsOct 2024: 3,001 reportsNov 2024: 2,083 reportsDec 2024: 1,706 reportsJan 2025: 2,826 reportsFeb 2025: 1,532 reportsMar 2025: 1,926 reportsApr 2025: 1,405 reportsMay 2025: 1,827 reportsJun 2025: 1,434 reportsJul 2025: 2,629 reportsAug 2025: 2,092 reportsSep 2025: 1,344 reportsOct 2025: 1,301 reportsNov 2025: 1,657 reportsDec 2025: 2,280 reportsJan 2026: 1,845 reportsFeb 2026: 1,771 reportsMar 2026: 1,855 reportsApr 2026: 1,741 reportsMay 2026: 993 reportsJun 2026: 840 reportsJul 2026: 372 reportsAug 2026: 246 reportsSep 2026: 78 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263,060
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 21, 2025Class IOpen, Classified

    Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization.

    Z-2343-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Pumping Stopped35,390
  • Communication or Transmission Problem7,338
  • Battery Problem6,725
  • Incorrect Measurement6,636
  • Charging Problem6,099
  • Obstruction of Flow4,431
  • Adverse Event Without Identified Device or Use Problem3,914
  • Filling Problem2,970

Patient problems

  • No Clinical Signs, Symptoms or Conditions75,013
  • Hyperglycemia6,692
  • Hypoglycemia2,010
  • Elevated Ketones/ Diabetic Ketoacidosis423
  • Insufficient Information113
  • Loss of consciousness104
  • Fall44
  • Vomiting40

Recent adverse event reports

The 25 most recent of 83,767 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026False Alarm

    It was reported that ongoing altitude alarms occurred. Reportedly, the pump was not outside the labeled altitude operating range. The customer experienced an elevated blood glucose (BG) level displayed as \"High\"; although a specific BG value was not provided. Customer performed a supply change and administered a correction injection to address the BG. The alarm ultimately cleared, and the customer continued to use the pump for insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50954985 ↗
  • InjurySep 30, 2026False Alarm

    It was reported that an altitude alarm occurred. Reportedly, the pump was not outside the labeled altitude operating range. The customer experienced an elevated blood glucose (BG) level displayed as \"High\"; although a specific BG value was not provided. Customer administered a correction injection to address the BG. Customer reverted to an alternate method of insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50951997 ↗
  • InjurySep 29, 2026False Alarm

    It was reported that an altitude alarm occurred. Reportedly, the pump was not outside the labeled altitude operating range. The customer experienced an elevated blood glucose (BG) level displayed as \"High\"; although a specific BG value was not provided. Customer administered a correction injection to address the BG. The alarm ultimately cleared, and the customer continued to use the pump for insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50941310 ↗
  • MalfunctionSep 29, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940829 ↗
  • InjurySep 29, 2026Pumping Stopped

    A malfunction was reported on the pump. The customer¿s blood glucose (BG) level displayed as \"High\"; although a specific value was not provided. A correction injection (MDI) was delivered to address the elevated BG level, and there was no need for third-party assistance. Technical Support provided information to reset the pump, assisted the caller with the reset, and confirmed that the malfunction code did not recur. The customer was advised to continue using the pump for insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50939670 ↗
  • MalfunctionSep 28, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50924159 ↗
  • MalfunctionSep 28, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923906 ↗
  • MalfunctionSep 28, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923898 ↗
  • InjurySep 28, 2026Pumping Stopped

    It was reported that pump experienced malfunction alarm related to speaker, and customer noted that glucose readings were showing high on connected app. Customer¿s blood glucose level exceeded 500 mg/dL (27.7 mmol/L) due to issue, and there was no reported assistance from third party. Customer had not taken any actions to address high blood glucose at time of report and had no backup insulin therapy available, with plan to contact healthcare provider. Technical Support confirmed pump was under warranty, provided instructions for placing pump into storage mode and returning malfunctioning pump, and arranged shipment of replacement pump, advising customer to re-enter pump settings and reconnect glucose monitor to replacement device.

    Patient problems: Hyperglycemia

    MAUDE report 50923895 ↗
  • InjurySep 28, 2026Pumping Stopped

    It was reported that a cartridge alarm occurred during insulin delivery. The customer experienced an elevated blood glucose (BG) level displayed as \"High\"; although a specific BG value was not provided. Customer administered a correction injection to address the BG. Customer loaded a new cartridge to resolve the issue.

    Patient problems: Hyperglycemia

    MAUDE report 50923893 ↗
  • InjurySep 28, 2026Pumping Stopped

    It was reported that pump malfunction alarm occurred and displayed malfunction code, and caller saw blood glucose reading marked as ¿High¿ on pump screen. Customer¿s blood glucose level due to issue was reported as above 500 mg/dL, with specific value unknown. Customer delivered correction insulin dose using pump, contacted Technical Support, received instructions to reset pump, and successfully cleared malfunction alarm without recurrence, and was advised to continue using pump and to call back if malfunction appeared again.

    Patient problems: Hyperglycemia

    MAUDE report 50923369 ↗
  • InjurySep 27, 2026Pumping Stopped

    It was reported that ongoing cartridge alarms occurred during insulin delivery. Subsequently, the customer experienced an elevated blood glucose (BG) displayed as \"High.\" A specific BG value was not provided. The customer administered a manual injection of insulin to address their BG. Customer reverted to an alternate method of insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50910383 ↗
  • InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that customer experienced a blood glucose (BG) level that ranged from 499 mg/dL to a level was displayed as ¿High¿, cause was unknown. Due to the elevated BG level the customer \"vomited\". Customer gave a manual injection and changed out infusion set to address BG level. Reportedly, customer continued to use the pump for insulin therapy.

    Patient problems: Hyperglycemia; Vomiting

    MAUDE report 50892435 ↗
  • MalfunctionSep 23, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50863097 ↗
  • InjurySep 18, 2026Obstruction of Flow

    It was reported that intermittent occlusion alarms occurred. The customer's blood glucose level was \"High\", although specific value was not provided on 9/10/2026. The elevated BG was addressed by a correction bolus via the pump. No third party assistance was required. Customer changed infusion set and cartridge and resumed insulin to resolve the issue.

    Patient problems: Hyperglycemia

    MAUDE report 50828223 ↗
  • MalfunctionSep 17, 2026False Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812608 ↗
  • MalfunctionSep 17, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812503 ↗
  • MalfunctionSep 17, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812494 ↗
  • MalfunctionSep 17, 2026Mechanical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812488 ↗
  • MalfunctionSep 17, 2026Mechanical Problem; Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812479 ↗
  • InjurySep 17, 2026Mechanical Problem; Pumping Stopped

    It was reported that multiple cartridge alarms occurred during the load sequence and insulin delivery. The customer experienced an elevated blood glucose (BG) level of 520 mg/dL. Customer administered a correction injection to address the BG. Customer reverted to an alternate method of insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50812289 ↗
  • MalfunctionSep 17, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812279 ↗
  • InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the customer experienced a blood glucose (BG) level that was displayed as ¿High¿; however, a specific value was not provided, cause was unknown. Customer gave a correction bolus via the pump to address BG level. Recommendation was made to discuss diabetes management with a health care professional.

    Patient problems: Hyperglycemia

    MAUDE report 50810191 ↗
  • MalfunctionSep 17, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50810167 ↗
  • MalfunctionSep 17, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50809733 ↗

FDA clearance and classification

The FDA regulates product code QFG (alternate controller enabled insulin infusion pump) under 21 CFR 880.5730 as a Class II device. 21 CFR 880.5730 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: An alternate controller enabled insulin infusion pump (ACE insulin pump) is a device intended for the infusion of insulin into a patient. The ACE insulin pump may include basal and bolus drug delivery at set or variable rates. ACE insulin pumps are designed to reliably and securely communicate with external devices, such as automated insulin dosing systems, to allow insulin delivery commands to be received, executed, and confirmed. ACE insulin pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of insulin delivery.

Common questions about the t:slim X2 Insulin Pump with Control-IQ 1009637

Has the t:slim X2 Insulin Pump with Control-IQ 1009637 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2343-2025 (Class I), began on Jul 21, 2025: Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the t:slim X2 Insulin Pump with Control-IQ 1009637?

The FDA's MAUDE database lists 83,767 adverse event reports for this device, received between Aug 22, 2022 and Sep 30, 2026. In those reports, the most common specific device problems are “Pumping Stopped” (35,390), “Communication or Transmission Problem” (7,338) and “Battery Problem” (6,725); the most common patient problems are “Hyperglycemia” (6,692), “Hypoglycemia” (2,010) and “Elevated Ketones/ Diabetic Ketoacidosis” (423). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the t:slim X2 Insulin Pump with Control-IQ 1009637?

4 reports are classified as deaths, 8,868 as injuries and 74,895 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the t:slim X2 Insulin Pump with Control-IQ 1009637 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the t:slim X2 Insulin Pump with Control-IQ 1009637, and is it still sold?

The FDA's device database lists Tandem Diabetes Care, Inc. as the company responsible for it (brand t:slim X2 Insulin Pump with Control-IQ, model 1009637). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Alternate Controller Enabled Insulin Infusion Pump devices

All Alternate Controller Enabled Insulin Infusion Pump devices →