t:slim X2 Insulin Pump with Control-IQ 1006402Recalls & reported problems

Tandem Diabetes Care, Inc. · t:slim X2 Insulin Pump with Control-IQ, 7.4.x · Class 2

All models of t:slim X2 Insulin Pump (45) · t:slim X2 Insulin Pump with Control-IQ (11)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This is a Tandem t:slim X2 insulin pump. Its reports mostly describe pump stoppages, battery trouble, unresponsive touchscreens and blocked insulin flow. There are 159,360 reports, including 18 deaths and 15,792 injuries. Two recalls are open: one covers speakers that may fail and stop insulin delivery, the other covers memory, battery display and touchscreen problems. Reports dropped to 6,311 in the last 12 months, from 22,854 before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
159,198
Deaths reported
18
Injuries reported
15,794
Recalls naming it
2

Summary

The FDA's MAUDE database contains 159,198 adverse event reports that list this exact device (DI 00850006613762), received between Aug 10, 2020 and Sep 22, 2026 — classified as 18 deaths, 15,794 injuries, 143,386 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (48,278 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 2,500 reportsJan 2024: 2,561 reportsFeb 2024: 2,321 reportsMar 2024: 3,737 reportsApr 2024: 3,571 reportsMay 2024: 2,896 reportsJun 2024: 2,805 reportsJul 2024: 3,220 reportsAug 2024: 3,111 reportsSep 2024: 3,355 reportsOct 2024: 3,449 reportsNov 2024: 2,413 reportsDec 2024: 2,129 reportsJan 2025: 3,134 reportsFeb 2025: 1,550 reportsMar 2025: 2,022 reportsApr 2025: 1,439 reportsMay 2025: 1,873 reportsJun 2025: 1,454 reportsJul 2025: 2,292 reportsAug 2025: 2,059 reportsSep 2025: 1,302 reportsOct 2025: 1,187 reportsNov 2025: 974 reportsDec 2025: 1,089 reportsJan 2026: 750 reportsFeb 2026: 695 reportsMar 2026: 542 reportsApr 2026: 462 reportsMay 2026: 470 reportsJun 2026: 765 reportsJul 2026: 321 reportsAug 2026: 71 reportsSep 2026: 10 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263,737
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 21, 2025Class IOpen, Classified

    Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization.

    Z-2343-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  2. May 24, 2022Class IIOpen, Classified

    Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.

    Z-1811-2022 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Pumping Stopped48,278
  • Incorrect Measurement16,023
  • Charging Problem15,369
  • Battery Problem12,479
  • Obstruction of Flow9,765
  • Adverse Event Without Identified Device or Use Problem9,358
  • Communication or Transmission Problem8,288
  • Failure to Sense7,613

Patient problems

  • No Clinical Signs, Symptoms or Conditions141,192
  • Hyperglycemia9,846
  • Hypoglycemia5,438
  • No Consequences Or Impact To Patient1,945
  • Elevated Ketones/ Diabetic Ketoacidosis1,051
  • Insufficient Information486
  • No Known Impact Or Consequence To Patient326
  • Loss of consciousness194

Recent adverse event reports

The 25 most recent of 159,198 reports, in the reporter's own words as published by the FDA.

  • InjurySep 22, 2026Pumping Stopped

    It was reported that a cartridge alarm occurred during insulin delivery. The customer experienced an elevated blood glucose (BG) level displayed as \"High\"; although a specific BG value was not provided. Customer delivered a correction bolus via the pump to address the BG. Customer will continue to use the current pump for insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50849399 ↗
  • MalfunctionSep 22, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50849385 ↗
  • InjurySep 21, 2026Key or Button Unresponsive/not Working

    It was reported that pump had unresponsive or glitching touchscreen during bolus delivery, requiring pump to be turned off to reset. There was no specific blood glucose values were provided. Customer declined troubleshooting or call back, and no further technical support actions were taken, with no additional information received regarding outcome of event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50846916 ↗
  • MalfunctionSep 15, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50778443 ↗
  • InjurySep 11, 2026Obstruction of Flow

    It was reported that occlusion alarms occurred. Customer¿s blood glucose level was 555 mg/dL. The elevated BG was addressed by a correction injection via manual insulin therapy. No third party assistance was required. A system check was performed by Tandem Technical Support and there was debris around the pneumatic tap. The customer cleaned around the pneumatic tap, changed the pump supplies and successfully resumed insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50725250 ↗
  • MalfunctionSep 8, 2026Insufficient Flow or Under Infusion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50679502 ↗
  • InjurySep 3, 2026Use of Device Problem

    It was reported that the customer allowed the pump battery to fully deplete due to not charging the battery and the pump subsequently shut off. Customer¿s blood glucose (BG) level was displayed as "High", although a specific value was not given. The customer addressed their BG with a correction injection. Customer also had ketones which were not identified as life threatening by the healthcare provider. The customer continued using the pump after resuming insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50592073 ↗
  • InjurySep 2, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the customer experienced a blood glucose (BG) level of 30 mg/dL, cause was unknown. Customer went to the emergency room (ER) and was treated with intravenous fluids of glucagon and was released from the ER 5-6 hours later with the issue resolved and no permanent damage. Reportedly, the customer reverted to an alternate method of insulin therapy.

    Patient problems: Hypoglycemia

    MAUDE report 50571450 ↗
  • InjurySep 2, 2026Pumping Stopped

    It was reported that cartridge alarm 20 occurred and caller experienced repeated cartridge alarms with consecutive cartridges. Customer¿s blood glucose level was over 600 mg/dL due to the issue. Customer gave correction insulin by injection. The customer reverted to manual injections for insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50569444 ↗
  • InjurySep 2, 2026Pumping Stopped

    It was reported that pump triggered cartridge alarm 20 and customer experienced repeated cartridge alarms with consecutive cartridges, so pump was no longer trusted for insulin delivery. Customer¿s blood glucose reading was reported as ¿High,¿ but specific value was unknown. Customer gave correction injection using multiple daily injections, switched to multiple daily injections as backup therapy, and did not require help from another healthcare provider; Technical Support provided instructions for putting pump into storage mode, arranged return of pump, and sent replacement pump with updated software, advising customer to enter pump settings and reconnect continuous glucose monitor on replacement pump.

    Patient problems: Hyperglycemia

    MAUDE report 50562724 ↗
  • MalfunctionAug 21, 2026Premature Discharge of Battery

    It was reported that pump had low battery life, dirt in charging port, and a scratch on pump, but pump was still manageable for customer. There was no reported impact to the customer¿s blood glucose level. Customer declined transfer to Technical Support, started renewal process for pump, and no additional information was received regarding further actions or outcome of event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50369777 ↗
  • MalfunctionAug 21, 2026Premature Discharge of Battery

    It was reported that pump battery was depleting too fast, requiring charging every other day. There was no reported impact to the customer¿s blood glucose level. Customer declined support from Technical Support, and CTS was unable to confirm battery depletion allegation due to lack of pump logs, with no additional information received regarding outcome of event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50369771 ↗
  • MalfunctionAug 21, 2026Key or Button Unresponsive/not Working

    It was reported that pump quick bolus/wake button did not function as expected. There was no reported impact to the customer¿s blood glucose level. Customer outcome or actions taken were not reported, and no additional information was received regarding the outcome of the event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50368367 ↗
  • MalfunctionAug 21, 2026Premature Discharge of Battery

    It was reported that the pump battery was depleting quickly. There was no reported adverse impact to the customer. Technical Support checked source information but found that logs were not uploaded, and case was closed with no further actions taken.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50361341 ↗
  • MalfunctionAug 21, 2026Premature Discharge of Battery

    It was reported that the pump battery was depleting quickly. There was no reported adverse impact to the customer. The customer declined further follow up from Tandem Technical Support. No additional information was provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50361115 ↗
  • MalfunctionAug 21, 2026Key or Button Unresponsive/not Working

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50360425 ↗
  • MalfunctionAug 21, 2026Key or Button Unresponsive/not Working

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50360417 ↗
  • MalfunctionAug 20, 2026Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50359613 ↗
  • MalfunctionAug 20, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50358808 ↗
  • MalfunctionAug 20, 2026Unexpected Shutdown

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50358798 ↗
  • MalfunctionAug 20, 2026Obstruction of Flow

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50358373 ↗
  • MalfunctionAug 20, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50357645 ↗
  • MalfunctionAug 20, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50357643 ↗
  • MalfunctionAug 20, 2026Key or Button Unresponsive/not Working

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50357541 ↗
  • InjuryAug 20, 2026Obstruction of Flow

    It was reported that customer experienced multiple intermittent occlusion alarms. Customer's blood glucose (BG) displayed as ¿High.¿ Customer changed supplies, then did correction bolus via pump to address the high BG. Reportedly, the customer changed cartridge and infusion set to resolve the issue.

    Patient problems: Hyperglycemia

    MAUDE report 50356495 ↗

FDA clearance and classification

The FDA regulates product code QFG (alternate controller enabled insulin infusion pump) under 21 CFR 880.5730 as a Class II device. 21 CFR 880.5730 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: An alternate controller enabled insulin infusion pump (ACE insulin pump) is a device intended for the infusion of insulin into a patient. The ACE insulin pump may include basal and bolus drug delivery at set or variable rates. ACE insulin pumps are designed to reliably and securely communicate with external devices, such as automated insulin dosing systems, to allow insulin delivery commands to be received, executed, and confirmed. ACE insulin pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of insulin delivery.

Common questions about the t:slim X2 Insulin Pump with Control-IQ 1006402

Has the t:slim X2 Insulin Pump with Control-IQ 1006402 been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-2343-2025 (Class I), began on Jul 21, 2025: Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the t:slim X2 Insulin Pump with Control-IQ 1006402?

The FDA's MAUDE database lists 159,198 adverse event reports for this device, received between Aug 10, 2020 and Sep 22, 2026. In those reports, the most common specific device problems are “Pumping Stopped” (48,278), “Incorrect Measurement” (16,023) and “Charging Problem” (15,369); the most common patient problems are “Hyperglycemia” (9,846), “Hypoglycemia” (5,438) and “Elevated Ketones/ Diabetic Ketoacidosis” (1,051). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the t:slim X2 Insulin Pump with Control-IQ 1006402?

18 reports are classified as deaths, 15,794 as injuries and 143,386 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the t:slim X2 Insulin Pump with Control-IQ 1006402 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the t:slim X2 Insulin Pump with Control-IQ 1006402, and is it still sold?

The FDA's device database lists Tandem Diabetes Care, Inc. as the company responsible for it (brand t:slim X2 Insulin Pump with Control-IQ, model 1006402). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Alternate Controller Enabled Insulin Infusion Pump devices

All Alternate Controller Enabled Insulin Infusion Pump devices →