t:slim X2 Insulin Pump with Control-IQ 1006404Recalls & reported problems

Tandem Diabetes Care, Inc. · t:slim X2 Insulin Pump with Control-IQ, 7.4.x, Medicare · Class 2

All models of t:slim X2 Insulin Pump (45) · t:slim X2 Insulin Pump with Control-IQ (11)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This is a Tandem t:slim X2 insulin pump with Control-IQ, Medicare version. Its reports mostly describe pump stoppages, incorrect measurements, charging and cartridge filling problems, and blocked flow. There are 43,740 reports, including 18 deaths and 6,399 injuries. One recall, started in May 2022, is open: memory faults, an inaccurate battery display and a touchscreen that stays on. Reports dropped to 1,465 in the last 12 months, from 5,326 before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
43,714
Deaths reported
18
Injuries reported
6,404
Recalls naming it
1

Summary

The FDA's MAUDE database contains 43,714 adverse event reports that list this exact device (DI 00850006613779), received between Aug 27, 2020 and Sep 28, 2026 — classified as 18 deaths, 6,404 injuries, 37,292 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (9,866 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 657 reportsJan 2024: 670 reportsFeb 2024: 621 reportsMar 2024: 823 reportsApr 2024: 844 reportsMay 2024: 682 reportsJun 2024: 616 reportsJul 2024: 797 reportsAug 2024: 678 reportsSep 2024: 699 reportsOct 2024: 738 reportsNov 2024: 504 reportsDec 2024: 454 reportsJan 2025: 689 reportsFeb 2025: 399 reportsMar 2025: 499 reportsApr 2025: 313 reportsMay 2025: 486 reportsJun 2025: 349 reportsJul 2025: 595 reportsAug 2025: 450 reportsSep 2025: 283 reportsOct 2025: 305 reportsNov 2025: 193 reportsDec 2025: 251 reportsJan 2026: 206 reportsFeb 2026: 162 reportsMar 2026: 183 reportsApr 2026: 141 reportsMay 2026: 115 reportsJun 2026: 114 reportsJul 2026: 46 reportsAug 2026: 20 reportsSep 2026: 8 reportsOct 2026: 0 reportsNov 2026: 0 reports202420252026844
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 24, 2022Class IIOpen, Classified

    Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.

    Z-1811-2022 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Pumping Stopped9,866
  • Incorrect Measurement5,665
  • Charging Problem4,741
  • Adverse Event Without Identified Device or Use Problem4,470
  • Filling Problem3,359
  • Obstruction of Flow2,730
  • Communication or Transmission Problem2,232
  • Battery Problem1,932

Patient problems

  • No Clinical Signs, Symptoms or Conditions36,622
  • Hyperglycemia3,821
  • Hypoglycemia2,439
  • No Consequences Or Impact To Patient609
  • Elevated Ketones/ Diabetic Ketoacidosis431
  • Insufficient Information149
  • Loss of consciousness122
  • No Known Impact Or Consequence To Patient89

Recent adverse event reports

The 25 most recent of 43,714 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Use of Device Problem

    It was reported that customer experienced a recurring issue during tubing fill in which it took significantly more insulin units to see drops at end of tubing, increasing from about 10¿12 units to 20¿25 units, and customer stated this issue had been getting worse since late June. The customer¿s blood glucose (BG) level was displayed as ¿High¿; however, a specific value was not provided. Customer gave correction insulin by injection to address high blood glucose and was instructed by Technical Support to hold tubing downward while filling, and customer was able to see insulin drops after using 20¿25 units; customer was advised to monitor issue and call back as needed, and later follow-up attempts were unsuccessful due to full voicemail, so no additional information was received regarding the outcome of the event.

    Patient problems: Hyperglycemia

    MAUDE report 50923655 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the customer had a low blood sugar, leading to emergency room visit; suspected cause of low blood sugar was unknown. Customer experienced low blood sugar that required emergency care. Customer declined further troubleshooting, and no additional information was received regarding the outcome of the event.

    Patient problems: Hypoglycemia

    MAUDE report 50922297 ↗
  • InjurySep 11, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that customer experienced high blood glucose levels, with a highest reported value of 511 mg/dL, and cause of high reading was unknown. Customer had not taken any actions before contacting support; Technical Support guided customer through checking infusion site, tubing, and insulin, found no issues, and recommended customer consult healthcare provider about what might be affecting blood glucose and replace supplies as needed before resuming insulin.

    Patient problems: Hyperglycemia

    MAUDE report 50743048 ↗
  • MalfunctionSep 11, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50740690 ↗
  • MalfunctionSep 11, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50740689 ↗
  • InjurySep 11, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that customer experienced a past high blood glucose event displaying ¿high¿; specific value was not provided, and suspected cause unknown. Customer changed infusion set and cartridge to address high blood glucose, and Technical Support advised customer to consult healthcare provider about possible factors affecting blood glucose, after which case was closed.

    Patient problems: Hyperglycemia

    MAUDE report 50740469 ↗
  • InjurySep 10, 2026Use of Incorrect Control/Treatment Settings

    It was reported that the customer experienced a low blood glucose (BG) level of below 54 mg/dL. Suspected cause was due to the customer¿s settings requiring adjustments. It is unknown what actions were taken to address BG. Customer updated their pump settings to address the event.

    Patient problems: Hypoglycemia

    MAUDE report 50722588 ↗
  • InjurySep 10, 2026Obstruction of Flow

    It was reported that intermittent occlusion alarms occurred. The customer's blood glucose level was 579 mg/dL on (B(6)2026. The elevated BG was addressed by a correction injection via manual insulin therapy. Customer changed infusion set and cartridge and resumed insulin to resolve the issue. Reportedly, the customer was using Fiasp insulin with the pump. Tandem Customer Technical Support informed customer that Fiasp insulin is off-label per the User Guide.

    Patient problems: Hyperglycemia

    MAUDE report 50722515 ↗
  • MalfunctionAug 31, 2026Key or Button Unresponsive/not Working

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50530766 ↗
  • InjuryAug 21, 2026Use of Device Problem

    It was reported that the pump displayed a minimum fill notification. Reportedly the customer initially filled the cartridge with 60 units of insulin. Technical support educated the customer of the minimum fill recommendation for the pump. The customer¿s blood glucose (BG) level was displayed as ¿High¿; however, a specific value was not provided. Customer changed out supplies and then administrated a correction bolus via pump to address BG. Customer added more insulin to the cartridge and resumed insulin delivery.

    Patient problems: Hyperglycemia

    MAUDE report 50369775 ↗
  • MalfunctionAug 20, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50358375 ↗
  • MalfunctionAug 20, 2026Failure to Charge

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50355468 ↗
  • MalfunctionAug 20, 2026Mechanical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50355444 ↗
  • MalfunctionAug 20, 2026Failure to Charge

    It was reported that the pump battery was charging slower than expected. There was no reported impact to the customer's blood glucose level. No additional information was provided and the customer declined troubleshooting with Tandem Technical Support.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50355022 ↗
  • MalfunctionAug 20, 2026Date/Time-Related Software Problem

    It was reported that the pump time was incorrect, cause was unknown. There was no adverse impact to the customer's blood glucose level. During troubleshooting with Tandem Technical Support, the customer corrected the time and resumed insulin therapy.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50352895 ↗
  • MalfunctionAug 20, 2026Date/Time-Related Software Problem

    It was reported that the pump time was incorrect, cause was unknown. There was no adverse impact to the customer's blood glucose level. During troubleshooting with Tandem Technical Support, the customer corrected the time and resumed insulin therapy.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50348545 ↗
  • MalfunctionAug 19, 2026Incorrect Measurement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50331499 ↗
  • MalfunctionAug 19, 2026Incorrect Measurement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50331485 ↗
  • MalfunctionAug 19, 2026Charging Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50330898 ↗
  • MalfunctionAug 19, 2026Incorrect Measurement

    It was reported that the insulin gauge was inaccurate. There was no reported adverse impact to the customer. The customer declined follow-up from Tandem Technical Support regarding the issue; therefore, no additional information was obtained.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50328480 ↗
  • MalfunctionAug 19, 2026Premature Discharge of Battery

    It was reported that the pump battery was depleting quickly. There was no reported adverse impact to the customer¿s blood glucose level. The customer declined assistance from Tandem Technical Support regarding the issues. No additional patient or event information was available.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50327635 ↗
  • MalfunctionAug 19, 2026Premature Discharge of Battery

    It was reported that pump battery depleted rapidly and customer declined any troubleshooting. There was no reported impact to the customer¿s blood glucose level. Customer declined further support, no follow up was needed, and no additional information was received regarding the outcome of the event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50327251 ↗
  • InjuryAug 19, 2026Use of Device Problem

    It was reported that customer received alert for incomplete cartridge change and experienced high blood glucose reading, with specific value unknown but displayed as ¿High.¿ Customer delivered correction bolus using pump, did not require third-party assistance, and was educated by Technical Support on meaning of alert and instructed to return to pump home screen and ensure pump screen was off before putting pump away to prevent accidental screen touches and incomplete alerts; customer understood and acknowledged instructions.

    Patient problems: Hyperglycemia

    MAUDE report 50325072 ↗
  • MalfunctionAug 19, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50322739 ↗
  • MalfunctionAug 18, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50303958 ↗

FDA clearance and classification

The FDA regulates product code QFG (alternate controller enabled insulin infusion pump) under 21 CFR 880.5730 as a Class II device. 21 CFR 880.5730 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: An alternate controller enabled insulin infusion pump (ACE insulin pump) is a device intended for the infusion of insulin into a patient. The ACE insulin pump may include basal and bolus drug delivery at set or variable rates. ACE insulin pumps are designed to reliably and securely communicate with external devices, such as automated insulin dosing systems, to allow insulin delivery commands to be received, executed, and confirmed. ACE insulin pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of insulin delivery.

Common questions about the t:slim X2 Insulin Pump with Control-IQ 1006404

Has the t:slim X2 Insulin Pump with Control-IQ 1006404 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1811-2022 (Class II), began on May 24, 2022: Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the t:slim X2 Insulin Pump with Control-IQ 1006404?

The FDA's MAUDE database lists 43,714 adverse event reports for this device, received between Aug 27, 2020 and Sep 28, 2026. In those reports, the most common specific device problems are “Pumping Stopped” (9,866), “Incorrect Measurement” (5,665) and “Charging Problem” (4,741); the most common patient problems are “Hyperglycemia” (3,821), “Hypoglycemia” (2,439) and “Elevated Ketones/ Diabetic Ketoacidosis” (431). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the t:slim X2 Insulin Pump with Control-IQ 1006404?

18 reports are classified as deaths, 6,404 as injuries and 37,292 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the t:slim X2 Insulin Pump with Control-IQ 1006404 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the t:slim X2 Insulin Pump with Control-IQ 1006404, and is it still sold?

The FDA's device database lists Tandem Diabetes Care, Inc. as the company responsible for it (brand t:slim X2 Insulin Pump with Control-IQ, model 1006404). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Alternate Controller Enabled Insulin Infusion Pump devices

All Alternate Controller Enabled Insulin Infusion Pump devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page