t:slim X2 Insulin Pump with Control-IQ Technology 1013198Recalls & reported problems

Tandem Diabetes Care, Inc. · Control-IQ 7.8 · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This is a Tandem t:slim X2 insulin pump with Control-IQ. Its reports mostly describe pump stoppages, charging problems, incorrect measurements and blocked flow. There are 47,932 reports, including 8 deaths and 6,039 injuries. One recall, started in July 2025, is open: it covers speakers that may fail and stop insulin delivery.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
47,637
Deaths reported
8
Injuries reported
6,045
Recalls naming it
1

Summary

The FDA's MAUDE database contains 47,637 adverse event reports that list this exact device (DI 00389152319810), received between Feb 2, 2024 and Sep 30, 2026 — classified as 8 deaths, 6,045 injuries, 41,584 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (17,037 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 117 reportsMar 2024: 294 reportsApr 2024: 566 reportsMay 2024: 632 reportsJun 2024: 861 reportsJul 2024: 1,272 reportsAug 2024: 1,360 reportsSep 2024: 1,632 reportsOct 2024: 1,864 reportsNov 2024: 1,607 reportsDec 2024: 1,684 reportsJan 2025: 2,970 reportsFeb 2025: 1,674 reportsMar 2025: 2,454 reportsApr 2025: 1,768 reportsMay 2025: 2,346 reportsJun 2025: 1,906 reportsJul 2025: 2,886 reportsAug 2025: 2,313 reportsSep 2025: 1,436 reportsOct 2025: 1,439 reportsNov 2025: 1,890 reportsDec 2025: 2,456 reportsJan 2026: 1,840 reportsFeb 2026: 1,778 reportsMar 2026: 1,713 reportsApr 2026: 1,635 reportsMay 2026: 1,231 reportsJun 2026: 1,257 reportsJul 2026: 386 reportsAug 2026: 195 reportsSep 2026: 175 reportsOct 2026: 0 reports2024202520262,970
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 21, 2025Open, Classified

    Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization.

    Z-2343-2025 · Root cause: Nonconforming Material/Component · FDA record ↗

Most reported problems

Device problems

  • Pumping Stopped17,037
  • Charging Problem4,893
  • Incorrect Measurement3,529
  • Adverse Event Without Identified Device or Use Problem2,727
  • Obstruction of Flow2,593
  • Battery Problem2,300
  • Failure to Charge2,269
  • Communication or Transmission Problem2,014

Patient problems

  • No Clinical Signs, Symptoms or Conditions41,631
  • Hyperglycemia4,439
  • Hypoglycemia1,543
  • Elevated Ketones/ Diabetic Ketoacidosis219
  • Loss of consciousness106
  • Insufficient Information89
  • Fall35
  • Bruise/Contusion18

Recent adverse event reports

The 25 most recent of 47,637 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Mechanical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50955079 ↗
  • MalfunctionSep 30, 2026False Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954978 ↗
  • MalfunctionSep 30, 2026Failure to Pump

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954869 ↗
  • MalfunctionSep 30, 2026False Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954864 ↗
  • InjurySep 30, 2026Key or Button Unresponsive/not Working

    Patient problems: Hyperglycemia

    MAUDE report 50952482 ↗
  • MalfunctionSep 30, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951991 ↗
  • MalfunctionSep 30, 2026Break

    It was reported that a cartridge did not fit onto the pump. Reportedly, there was an O-ring from a previous cartridge on the pneumatic tap. The customer removed the O-ring and successfully loaded the cartridge to address the event. There was no reported adverse impact to the customer.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951711 ↗
  • MalfunctionSep 30, 2026False Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951450 ↗
  • InjurySep 30, 2026Break; Pumping Stopped; Charging Problem

    It was reported that the pump's USB port was damaged resulting in difficulties charging the pump leading to the pump shutting down. The customer's blood glucose (BG) level was "High", although a specific value was not given. The customer administered a correction bolus via pump to address BG level. The customer reverted to an alternate method insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50951185 ↗
  • MalfunctionSep 30, 2026Pumping Stopped

    A malfunction was reported on the pump. There was no reported adverse impact to the customer's blood glucose level. Technical Support provided information to reset the pump, assisted the caller with the reset, and confirmed that the malfunction code did not recur. The customer was advised to continue using the pump for insulin therapy.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950697 ↗
  • InjurySep 30, 2026False Alarm

    It was reported that pump repeatedly displayed altitude alarm and suspended insulin delivery over past month while within normal operating altitude. Customer¿s blood glucose level exceeded 500 mg/dL and meter displayed ¿High¿; specific value was unknown. Customer initially took no action for high blood glucose. Technical Support guided customer to clean speaker holes with soft brush and lint-free cloth, remove and reconnect cartridge, and wait for insulin to resume, after which insulin delivery restarted and altitude alarm did not recur. Pump resumed normal operation, insulin remained active, and customer received instructions on preventing future altitude alarms.

    Patient problems: Hyperglycemia

    MAUDE report 50950360 ↗
  • MalfunctionSep 30, 2026Audible Prompt/Feedback Problem; Tactile Prompts/Feedback

    It was reported that pump speaker and vibration function were too quiet, and pump did not make sound when an alert was triggered. There was no reported impact to the customer¿s blood glucose level. The customer continued to use the current pump and will revert to an alternate insulin therapy if necessary.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950243 ↗
  • MalfunctionSep 30, 2026Pumping Stopped

    A malfunction was reported on the pump. There was no reported adverse impact to the customer's blood glucose level. Technical Support provided information to reset the pump, assisted the caller with the reset, and confirmed that the malfunction code did not recur. The customer was advised to continue using the pump for insulin therapy.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949806 ↗
  • MalfunctionSep 30, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949790 ↗
  • MalfunctionSep 30, 2026Fitting Problem

    It was reported that customer sometimes could not get cartridge to snap into pump and had to use extra force to keep it in place. There was no reported impact to the customer¿s blood glucose level. Customer declined troubleshooting, and Technical Support created and closed case based on email, documented issue, and took no further action at that time.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949507 ↗
  • MalfunctionSep 30, 2026Pumping Stopped

    A malfunction was reported on the pump. There was no reported adverse impact to the customer. Technical Support provided information to reset the pump, assisted the caller with the reset, and confirmed that the malfunction code did not recur. The customer was advised to continue using the pump for insulin therapy.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949481 ↗
  • MalfunctionSep 30, 2026Loose or Intermittent Connection; Charging Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949185 ↗
  • MalfunctionSep 30, 2026Key or Button Unresponsive/not Working

    It was reported that the wake button on the pump was sticking. There was no adverse impact to the customer's blood glucose level. Customer declined to troubleshoot the issue with Tandem Technical Support, therefore no additional information was obtained.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949176 ↗
  • MalfunctionSep 30, 2026Failure to Charge

    It was reported that the pump battery could not be charged. There was no adverse impact to the customer's blood glucose. Reportedly, the customer was sent new charging supplies. No additional patient or event information was available.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949168 ↗
  • MalfunctionSep 30, 2026Loose or Intermittent Connection; Charging Problem

    It was reported that the pump's USB port was loose, and the pump battery intermittently could not be charged properly without the customer maneuvering the cable into the charging port at a certain angle. There was no adverse impact to the customer¿s blood glucose level. The customer declined follow-up from Tandem Technical Support regarding the issue, therefore no additional information was obtained.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948868 ↗
  • MalfunctionSep 30, 2026Failure to Charge; Pumping Stopped

    It was reported that the pump battery could not be charged which resulted in the pump shutting off. There was no adverse impact to the customer's blood glucose. Reportedly, the pump successfully began charging after leaving the pump plugged into the power source.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948851 ↗
  • MalfunctionSep 30, 2026Key or Button Unresponsive/not Working

    It was reported that pump touchscreen was delayed in responding to touch inputs and required multiple touches to register interactions. There was no reported impact to the customer¿s blood glucose level. Customer declined follow-up, and no additional troubleshooting or corrective actions were completed by Technical Support.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948822 ↗
  • InjurySep 30, 2026Pumping Stopped

    It was reported that the pump experienced a malfunction. The customer¿s blood glucose (BG) level displayed as "High"; although a specific value was not provided. A correction injection (MDI) was delivered to address the elevated BG level, and there was no need for third-party assistance. The pump was replaced to resolve the issue, and the customer continued to use the pump for insulin therapy and has an alternate method of insulin therapy available, if necessary, until the replacement arrives.

    Patient problems: Hyperglycemia

    MAUDE report 50948786 ↗
  • MalfunctionSep 30, 2026Charging Problem

    It was reported that the customer was unable to charge the pump due to damage to the Tandem-provided USB charging cable. There was no reported adverse impact to the customer's blood glucose level. A new charging cable was sent to the customer.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948490 ↗
  • MalfunctionSep 30, 2026Key or Button Unresponsive/not Working

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948467 ↗

Other Alternate Controller Enabled Insulin Infusion Pump devices

All Alternate Controller Enabled Insulin Infusion Pump devices →