t:slim X2 Insulin Pump with Control-IQ Technology 1013198Recalls & reported problems
Tandem Diabetes Care, Inc. · Control-IQ 7.8 · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This is a Tandem t:slim X2 insulin pump with Control-IQ. Its reports mostly describe pump stoppages, charging problems, incorrect measurements and blocked flow. There are 47,932 reports, including 8 deaths and 6,039 injuries. One recall, started in July 2025, is open: it covers speakers that may fail and stop insulin delivery.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 47,637
- Deaths reported
- 8
- Injuries reported
- 6,045
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 47,637 adverse event reports that list this exact device (DI 00389152319810), received between Feb 2, 2024 and Sep 30, 2026 — classified as 8 deaths, 6,045 injuries, 41,584 malfunctions.
The most frequently reported specific device problem is “Pumping Stopped” (17,037 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jul 21, 2025Open, Classified
Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization.
Z-2343-2025 · Root cause: Nonconforming Material/Component · FDA record ↗
Most reported problems
Device problems
- Pumping Stopped17,037
- Charging Problem4,893
- Incorrect Measurement3,529
- Adverse Event Without Identified Device or Use Problem2,727
- Obstruction of Flow2,593
- Battery Problem2,300
- Failure to Charge2,269
- Communication or Transmission Problem2,014
Patient problems
- No Clinical Signs, Symptoms or Conditions41,631
- Hyperglycemia4,439
- Hypoglycemia1,543
- Elevated Ketones/ Diabetic Ketoacidosis219
- Loss of consciousness106
- Insufficient Information89
- Fall35
- Bruise/Contusion18
Recent adverse event reports
The 25 most recent of 47,637 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Mechanical Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50955079 ↗MalfunctionSep 30, 2026False Alarm
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954978 ↗MalfunctionSep 30, 2026Failure to Pump
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954869 ↗MalfunctionSep 30, 2026False Alarm
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954864 ↗InjurySep 30, 2026Key or Button Unresponsive/not Working
Patient problems: Hyperglycemia
MAUDE report 50952482 ↗MalfunctionSep 30, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951991 ↗MalfunctionSep 30, 2026Break
It was reported that a cartridge did not fit onto the pump. Reportedly, there was an O-ring from a previous cartridge on the pneumatic tap. The customer removed the O-ring and successfully loaded the cartridge to address the event. There was no reported adverse impact to the customer.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951711 ↗MalfunctionSep 30, 2026False Alarm
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951450 ↗InjurySep 30, 2026Break; Pumping Stopped; Charging Problem
It was reported that the pump's USB port was damaged resulting in difficulties charging the pump leading to the pump shutting down. The customer's blood glucose (BG) level was "High", although a specific value was not given. The customer administered a correction bolus via pump to address BG level. The customer reverted to an alternate method insulin therapy.
Patient problems: Hyperglycemia
MAUDE report 50951185 ↗MalfunctionSep 30, 2026Pumping Stopped
A malfunction was reported on the pump. There was no reported adverse impact to the customer's blood glucose level. Technical Support provided information to reset the pump, assisted the caller with the reset, and confirmed that the malfunction code did not recur. The customer was advised to continue using the pump for insulin therapy.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50950697 ↗InjurySep 30, 2026False Alarm
It was reported that pump repeatedly displayed altitude alarm and suspended insulin delivery over past month while within normal operating altitude. Customer¿s blood glucose level exceeded 500 mg/dL and meter displayed ¿High¿; specific value was unknown. Customer initially took no action for high blood glucose. Technical Support guided customer to clean speaker holes with soft brush and lint-free cloth, remove and reconnect cartridge, and wait for insulin to resume, after which insulin delivery restarted and altitude alarm did not recur. Pump resumed normal operation, insulin remained active, and customer received instructions on preventing future altitude alarms.
Patient problems: Hyperglycemia
MAUDE report 50950360 ↗MalfunctionSep 30, 2026Audible Prompt/Feedback Problem; Tactile Prompts/Feedback
It was reported that pump speaker and vibration function were too quiet, and pump did not make sound when an alert was triggered. There was no reported impact to the customer¿s blood glucose level. The customer continued to use the current pump and will revert to an alternate insulin therapy if necessary.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50950243 ↗MalfunctionSep 30, 2026Pumping Stopped
A malfunction was reported on the pump. There was no reported adverse impact to the customer's blood glucose level. Technical Support provided information to reset the pump, assisted the caller with the reset, and confirmed that the malfunction code did not recur. The customer was advised to continue using the pump for insulin therapy.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949806 ↗MalfunctionSep 30, 2026Pumping Stopped
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949790 ↗MalfunctionSep 30, 2026Fitting Problem
It was reported that customer sometimes could not get cartridge to snap into pump and had to use extra force to keep it in place. There was no reported impact to the customer¿s blood glucose level. Customer declined troubleshooting, and Technical Support created and closed case based on email, documented issue, and took no further action at that time.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949507 ↗MalfunctionSep 30, 2026Pumping Stopped
A malfunction was reported on the pump. There was no reported adverse impact to the customer. Technical Support provided information to reset the pump, assisted the caller with the reset, and confirmed that the malfunction code did not recur. The customer was advised to continue using the pump for insulin therapy.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949481 ↗MalfunctionSep 30, 2026Loose or Intermittent Connection; Charging Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949185 ↗MalfunctionSep 30, 2026Key or Button Unresponsive/not Working
It was reported that the wake button on the pump was sticking. There was no adverse impact to the customer's blood glucose level. Customer declined to troubleshoot the issue with Tandem Technical Support, therefore no additional information was obtained.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949176 ↗MalfunctionSep 30, 2026Failure to Charge
It was reported that the pump battery could not be charged. There was no adverse impact to the customer's blood glucose. Reportedly, the customer was sent new charging supplies. No additional patient or event information was available.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949168 ↗MalfunctionSep 30, 2026Loose or Intermittent Connection; Charging Problem
It was reported that the pump's USB port was loose, and the pump battery intermittently could not be charged properly without the customer maneuvering the cable into the charging port at a certain angle. There was no adverse impact to the customer¿s blood glucose level. The customer declined follow-up from Tandem Technical Support regarding the issue, therefore no additional information was obtained.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948868 ↗MalfunctionSep 30, 2026Failure to Charge; Pumping Stopped
It was reported that the pump battery could not be charged which resulted in the pump shutting off. There was no adverse impact to the customer's blood glucose. Reportedly, the pump successfully began charging after leaving the pump plugged into the power source.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948851 ↗MalfunctionSep 30, 2026Key or Button Unresponsive/not Working
It was reported that pump touchscreen was delayed in responding to touch inputs and required multiple touches to register interactions. There was no reported impact to the customer¿s blood glucose level. Customer declined follow-up, and no additional troubleshooting or corrective actions were completed by Technical Support.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948822 ↗InjurySep 30, 2026Pumping Stopped
It was reported that the pump experienced a malfunction. The customer¿s blood glucose (BG) level displayed as "High"; although a specific value was not provided. A correction injection (MDI) was delivered to address the elevated BG level, and there was no need for third-party assistance. The pump was replaced to resolve the issue, and the customer continued to use the pump for insulin therapy and has an alternate method of insulin therapy available, if necessary, until the replacement arrives.
Patient problems: Hyperglycemia
MAUDE report 50948786 ↗MalfunctionSep 30, 2026Charging Problem
It was reported that the customer was unable to charge the pump due to damage to the Tandem-provided USB charging cable. There was no reported adverse impact to the customer's blood glucose level. A new charging cable was sent to the customer.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948490 ↗MalfunctionSep 30, 2026Key or Button Unresponsive/not Working
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948467 ↗
Other Alternate Controller Enabled Insulin Infusion Pump devices
- t:slim X2 Insulin Pump with Control-IQ 1006402Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump2 recalls18 deaths159,247 reports
- t:slim X2 Insulin Pump with Control-IQ 1006408Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump2 recalls12 deaths92,526 reports
- t:slim X2 Insulin Pump with Control-IQ 1009637Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall4 deaths83,884 reports
- Omnipod 5 Pod POD-BLE-H1-520Insulet Corporation · Alternate Controller Enabled Insulin Infusion Pump1 recall75 deaths71,459 reports
- Omnipod 5 Pod POD-OMNI-I1-6720Insulet Corporation · Alternate Controller Enabled Insulin Infusion Pump2 recalls11 deaths61,198 reports
- t:slim X2 Insulin Pump with Control-IQ 1009643Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall7 deaths51,110 reports
- t:slim X2 Insulin Pump with Control-IQ 1006404Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall18 deaths43,718 reports
- t:slim X2 Insulin Pump with Control-IQ 1005011Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall3 deaths35,600 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.