t:slim X2 Insulin Pump with Control-IQ Technology 1013198Recalls & reported problems
Tandem Diabetes Care, Inc. · Control-IQ 7.8 · Class 2
All models of t:slim X2 Insulin Pump (45) · t:slim X2 Insulin Pump with Control-IQ Technology (11)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This is a Tandem t:slim X2 insulin pump with Control-IQ. Its reports mostly describe pump stoppages, charging problems, incorrect measurements and blocked flow. There are 47,932 reports, including 8 deaths and 6,039 injuries. One recall, started in July 2025, is open: it covers speakers that may fail and stop insulin delivery.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 47,498
- Deaths reported
- 8
- Injuries reported
- 6,056
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 47,498 adverse event reports that list this exact device (DI 00389152319810), received between Feb 2, 2024 and Sep 30, 2026 — classified as 8 deaths, 6,056 injuries, 41,434 malfunctions.
The most frequently reported specific device problem is “Pumping Stopped” (17,005 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jul 21, 2025Class IOpen, Classified
Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization.
Z-2343-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Pumping Stopped17,005
- Charging Problem4,886
- Incorrect Measurement3,519
- Adverse Event Without Identified Device or Use Problem2,728
- Obstruction of Flow2,595
- Battery Problem2,300
- Failure to Charge2,257
- Communication or Transmission Problem2,011
Patient problems
- No Clinical Signs, Symptoms or Conditions41,481
- Hyperglycemia4,452
- Hypoglycemia1,541
- Elevated Ketones/ Diabetic Ketoacidosis222
- Loss of consciousness107
- Insufficient Information89
- Fall35
- Vomiting18
Recent adverse event reports
The 25 most recent of 47,498 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Key or Button Unresponsive/not Working
It was reported that pump screen went dark and would not turn on unless pump was plugged into power source. Customer¿s blood glucose (BG) level was displayed as \"High\"; however, specific value was not provided. Elevated BG was addressed by administering a correction bolus. The customer continued to use the current pump and will revert to an alternate insulin therapy if necessary.
Patient problems: Hyperglycemia
MAUDE report 50952482 ↗InjurySep 29, 2026Obstruction of Flow
It was reported intermittent occlusion alarms occurred. The customer's blood glucose level was displayed as \"High\"; a specific value was not provided. Elevated blood glucose was addressed with a manual dose injection. The customer changed supplies and resumed insulin therapy.
Patient problems: Hyperglycemia
MAUDE report 50941246 ↗InjurySep 29, 2026Fracture
It was reported that the pump touch screen was cracked, unresponsive and unreadable. The customer experienced an elevated blood glucose level of over 500 mg/dL and ketones. The customer went to the Emergency Room (ER) on (b)(6) 2026 and was subsequently hospitalized. The customer received intravenous fluids of saline and insulin to address the BG. The customer was released from the hospital 6 hours later on (b)(6) 2026 with the issue resolved and no permanent damage. The customer reverted to an alternate method of insulin therapy.
Patient problems: Hyperglycemia
MAUDE report 50941239 ↗InjurySep 29, 2026Obstruction of Flow
It was reported intermittent occlusion alarms occurred. The customer's blood glucose level was displayed as \"High\"; a specific value was not provided. Elevated blood glucose was addressed with a manual dose injection. The customer changed supplies and resumed insulin therapy.
Patient problems: Hyperglycemia
MAUDE report 50941099 ↗InjurySep 28, 2026Use of Device Problem
Reportedly, the pump ran out of insulin. Tandem Technical Support was unable to confirm the sequence of low insulin alerts and alarms in the pump history. The customer acknowledged that they forgot to change out their empty cartridge when they should have which led to elevated blood glucose (BG) displayed as \"high\"; although specific value was not provided. Reportedly, the customer took no action to address BG level. Customer loaded a new cartridge to resolve the issue.
Patient problems: Hyperglycemia
MAUDE report 50923946 ↗InjurySep 28, 2026Incorrect Measurement
It was reported that a minimum fill notification occurred after the user filled the cartridge with 80 units of insulin during the load sequence. The customer's blood glucose (BG) level was \"High\", although a specific value was not given. The customer administered a manual injection to address their BG. A new cartridge was loaded to resolve the issue.
Patient problems: Hyperglycemia
MAUDE report 50923686 ↗MalfunctionSep 28, 2026Pumping Stopped
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50923677 ↗InjurySep 28, 2026Pumping Stopped; Charging Problem; Material Twisted/Bent
It was reported that the pump's USB port was bent resulting in difficulties charging the pump, leading to the pump shutting off. Customer¿s blood glucose level was reported as ¿High,¿ with specific value unknown. Elevated BG was addressed by administering a correction bolus. The customer reverted to manual injections for insulin therapy.
Patient problems: Hyperglycemia
MAUDE report 50923674 ↗MalfunctionSep 28, 2026Mechanical Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50923650 ↗MalfunctionSep 28, 2026Pumping Stopped
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50923375 ↗InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem
It was reported that customer experienced low blood sugar earlier in the day, with cause unknown, and contacted Technical Support for troubleshooting. Lowest blood glucose level reported was 52 mg/dL. Customer was already at hospital for unrelated issue and was treated for low blood sugar with an IV and food including peanut butter and graham crackers, received assistance from hospital nurses, did not lose consciousness, and did not report any injury. Customer used pump with automated insulin adjustment feature turned on, was unable to upload pump data or return cartridge, and Technical Support advised customer to consult healthcare provider about factors affecting blood sugars, which customer understood.
Patient problems: Hypoglycemia
MAUDE report 50922323 ↗InjurySep 25, 2026Off-Label Use
It was reported that the customer experienced an elevated blood glucose (BG) level of 566 mg/dL. Reportedly the customer was using off label insulin in the pump. Customer changed supplies to address BG. Recommendation was made to consult with a healthcare provider regarding diabetes management.
Patient problems: Hyperglycemia
MAUDE report 50902511 ↗InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem
It was reported that customer experienced a blood glucose (BG) of 698 mg/dL, with ketones (size unknown); cause was unknown. Customer went to the emergency room and was admitted into the intensive care unit. Customer was treated with intravenous fluids of saline, insulin and electrolytes. At the time of the report with Tandem Technical Support, the customer was still admitted to the hospital with no known damage. Reportedly, the customer declined further assistance from Tandem Technical Support regarding the issue. No additional patient or event information was available.
Patient problems: Hyperglycemia
MAUDE report 50892392 ↗InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the customer experienced a blood glucose (BG) level of 17-22 mg/dL. Reportedly, the customer alleged the pump software did not accurately adjust the amount of insulin delivered resulting in the low BG. Customer went to the emergency room (ER) and was treated with intravenous fluids of \"glucagon.\" Customer was released the same day, (b)(6)2026 with the issue resolved and no permanent damage. Tandem Technical Support (CTS) performed a system check and performed a review of the pump data to determine a possible cause. CTS determined that the pump functioned as intended and the cause of the low BG was unknown, customer acknowledged and understood.
Patient problems: Hypoglycemia
MAUDE report 50890426 ↗InjurySep 22, 2026Obstruction of Flow
It was reported that customer experienced occlusion alarms during use. Customer¿s blood glucose level was reported displayed as ¿High,¿ with specific value unknown. Customer gave correction insulin by injection, changed infusion set and cartridge, and then resumed insulin delivery, and customer declined further assistance while technical support proceeded to close case.
Patient problems: Hyperglycemia
MAUDE report 50861791 ↗InjurySep 21, 2026Mechanical Problem
It was reported that a cartridge alarm occurred during the load sequence. The customer experienced an elevated blood glucose (BG) level displayed as \"High\"; although a specific BG value was not provided. Customer administered a correction injection to address the BG. Customer reverted to an alternate method of insulin therapy.
Patient problems: Hyperglycemia
MAUDE report 50849029 ↗InjurySep 21, 2026Pumping Stopped
It was reported that the pump experienced a malfunction. The customer could not confirm the fault ID because they reset the pump on their own prior to contacting Technical Support. Reportedly, the customer¿s blood glucose (BG) level was over 600 mg/dL with ketones. Ketone levels were identified as life threatening by health care provider. This event occurred two years ago, and the customer could not recall the date they went to the emergency room and was subsequently admitted to the Intensive Care Unit (ICU). The customer was treated with intravenous fluids of saline and insulin. Treatment resolved the issue and there was no permanent damage. The customer could not recall the exact date they were released from the ICU. The pump was replaced to resolve the issue, and the customer continued to use the pump for insulin therapy and has an alternate method of insulin therapy available, if necessary, until the replacement arrives.
Patient problems: Elevated Ketones/ Diabetic Ketoacidosis; Hyperglycemia
MAUDE report 50847439 ↗InjurySep 18, 2026Obstruction of Flow
It was reported that customer experienced multiple occlusion alarms, including malfunction alarm 2 followed by malfunction alarm 26, while using infusion set, cartridge, and Novolog / NovoRapid insulin, with events occurring over approximately last week and a half. Customer¿s blood glucose (BG) level during one event was reported as 415 mg/dL, and on another occasion BG level was displayed as ¿High¿; however, a specific value was not provided. Customer treated high BG with correction injection using multiple daily injections, changed infusion set and cartridge, then resumed insulin delivery on pump, and customer service transferred caller and requested follow-up for additional assistance; no additional information was received regarding the outcome of the event.
Patient problems: Hyperglycemia
MAUDE report 50828556 ↗InjurySep 18, 2026Obstruction of Flow
It was reported that pump displayed recurring occlusion alarms while TruSteel infusion set and cartridge were in use over the last couple of months. Customer¿s blood glucose (BG) reading reached ¿High,¿ with specific value unknown. Customer gave correction insulin by both injection and pump to address the BG level, then changed infusion set and cartridge and resumed insulin delivery without needing help from another healthcare provider, and case was closed by Technical Support.
Patient problems: Hyperglycemia
MAUDE report 50828205 ↗InjurySep 18, 2026Pumping Stopped
It was reported that the pump experienced a malfunction. The customer¿s blood glucose (BG) level displayed as \"High\"; although a specific value was not provided. A correction injection (MDI) was delivered to address the elevated BG level, and there was no need for third-party assistance. The customer did not have an alternate method of insulin therapy available. Reportedly, the customer would reach out to their HCP to obtain a backup method of insulin therapy.
Patient problems: Hyperglycemia
MAUDE report 50826914 ↗MalfunctionSep 17, 2026Key or Button Unresponsive/not Working
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50810435 ↗MalfunctionSep 17, 2026Pumping Stopped
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50810387 ↗InjurySep 17, 2026Obstruction of Flow
It was reported that ongoing intermittent occlusion alarms occurred, leading to the customer experiencing intermittent high blood glucose (BG) levels. The customer¿s BG level was displayed as ¿High¿; however, a specific value was not provided. The elevated BG was addressed by a correction injection via manual insulin therapy. No third party assistance was required. Customer changed infusion set and cartridge to resolve the issue. The customer continued to use the pump for insulin therapy and has an alternate method of insulin therapy available if necessary.
Patient problems: Hyperglycemia
MAUDE report 50809728 ↗InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the customer experienced a blood glucose (BG) level of 30 mg/dL, cause was unknown. Due to the low BG level the customer experienced kidney issues, pain spasms, seizure and loss of consciousness. Customer required assistance giving \"shots\" and consuming \"sugar.\" Customer went to the emergency room (ER) and was admitted into the intensive care unit. Customer was treated with intravenous fluids of ¿sugar drip/sugar shots¿ and ¿insulin drip¿ and was released from the hospital on (b)(6) 2026 with the issue resolved. Reportedly, the customer continued to use the pump for insulin therapy.
Patient problems: Convulsion/Seizure; Cramp(s) /Muscle Spasm(s); Hypoglycemia; Loss of consciousness; Renal Impairment
MAUDE report 50809280 ↗InjurySep 17, 2026Obstruction of Flow
It was reported that intermittent occlusion alarms occurred. The customer would change the pump supplies to address the occlusions. Reportedly, the customer had been using the same cartridge and infusion set for over 2 days using Humalog insulin. Tandem Customer Technical Support informed customer that Humalog insulin is indicated for use with Tandem supplies for 2 days. On(b)(6)2026 the customer went to the Emergency Room and was subsequently admitted to the intensive care unit with a blood glucose (BG) level displayed as ¿High¿ with ketones; however, a specific value was not provided. Customer was treated with intravenous fluids of saline and insulin to address the elevated BG. Treatment resolved the issue and there was no permanent damage. The customer was discharged on (b)(6)2026.
Patient problems: Hyperglycemia
MAUDE report 50809008 ↗
FDA clearance and classification
The FDA regulates product code QFG (alternate controller enabled insulin infusion pump) under 21 CFR 880.5730 as a Class II device. 21 CFR 880.5730 ↗
Devices of this type normally need 510(k) clearance before sale.
FDA definition of this device type: An alternate controller enabled insulin infusion pump (ACE insulin pump) is a device intended for the infusion of insulin into a patient. The ACE insulin pump may include basal and bolus drug delivery at set or variable rates. ACE insulin pumps are designed to reliably and securely communicate with external devices, such as automated insulin dosing systems, to allow insulin delivery commands to be received, executed, and confirmed. ACE insulin pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of insulin delivery.
Common questions about the t:slim X2 Insulin Pump with Control-IQ Technology 1013198
Has the t:slim X2 Insulin Pump with Control-IQ Technology 1013198 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2343-2025 (Class I), began on Jul 21, 2025: Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the t:slim X2 Insulin Pump with Control-IQ Technology 1013198?
The FDA's MAUDE database lists 47,498 adverse event reports for this device, received between Feb 2, 2024 and Sep 30, 2026. In those reports, the most common specific device problems are “Pumping Stopped” (17,005), “Charging Problem” (4,886) and “Incorrect Measurement” (3,519); the most common patient problems are “Hyperglycemia” (4,452), “Hypoglycemia” (1,541) and “Elevated Ketones/ Diabetic Ketoacidosis” (222). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the t:slim X2 Insulin Pump with Control-IQ Technology 1013198?
8 reports are classified as deaths, 6,056 as injuries and 41,434 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the t:slim X2 Insulin Pump with Control-IQ Technology 1013198 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the t:slim X2 Insulin Pump with Control-IQ Technology 1013198, and is it still sold?
The FDA's device database lists Tandem Diabetes Care, Inc. as the company responsible for it (brand t:slim X2 Insulin Pump with Control-IQ Technology, model 1013198). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Alternate Controller Enabled Insulin Infusion Pump devices
- t:slim X2 Insulin Pump with Control-IQ 1006402Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump2 recalls18 deaths159,198 reports
- t:slim X2 Insulin Pump with Control-IQ 1006408Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump2 recalls12 deaths92,505 reports
- t:slim X2 Insulin Pump with Control-IQ 1009637Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall4 deaths83,767 reports
- Omnipod 5 Pod POD-BLE-H1-520Insulet Corporation · Alternate Controller Enabled Insulin Infusion Pump1 recall74 deaths71,491 reports
- Omnipod 5 Pod POD-OMNI-I1-6720Insulet Corporation · Alternate Controller Enabled Insulin Infusion Pump2 recalls11 deaths60,668 reports
- t:slim X2 Insulin Pump with Control-IQ 1009643Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall7 deaths51,050 reports
- t:slim X2 Insulin Pump with Control-IQ 1006404Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall18 deaths43,714 reports
- t:slim X2 Insulin Pump with Control-IQ 1005011Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall3 deaths35,598 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page