t:slim X2 Insulin Pump with Control-IQ 1009643Recalls & reported problems

Tandem Diabetes Care, Inc. · t:slim X2 Insulin Pump with Control-IQ, 7.6.x, Refurbished Replacement · Class 2

All models of t:slim X2 Insulin Pump (45) · t:slim X2 Insulin Pump with Control-IQ (11)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This is a refurbished Tandem t:slim X2 insulin pump with Control-IQ. Its reports mostly describe pump stoppages, battery and charging trouble, and communication problems. There are 51,223 reports, including 6 deaths and 5,331 injuries. One recall, started in July 2025, is open: it covers speakers that may fail and stop insulin delivery. Reports dropped to 5,992 in the last 12 months, from 13,237 before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
51,050
Deaths reported
7
Injuries reported
5,335
Recalls naming it
1

Summary

The FDA's MAUDE database contains 51,050 adverse event reports that list this exact device (DI 00389152000107), received between Aug 25, 2022 and Sep 28, 2026 — classified as 7 deaths, 5,335 injuries, 45,708 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (17,192 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1,859 reportsJan 2024: 2,056 reportsFeb 2024: 1,853 reportsMar 2024: 2,277 reportsApr 2024: 2,323 reportsMay 2024: 1,878 reportsJun 2024: 1,674 reportsJul 2024: 2,181 reportsAug 2024: 1,830 reportsSep 2024: 2,021 reportsOct 2024: 1,988 reportsNov 2024: 1,405 reportsDec 2024: 1,187 reportsJan 2025: 1,759 reportsFeb 2025: 929 reportsMar 2025: 1,170 reportsApr 2025: 829 reportsMay 2025: 1,133 reportsJun 2025: 892 reportsJul 2025: 1,424 reportsAug 2025: 1,186 reportsSep 2025: 667 reportsOct 2025: 656 reportsNov 2025: 838 reportsDec 2025: 1,052 reportsJan 2026: 758 reportsFeb 2026: 721 reportsMar 2026: 619 reportsApr 2026: 586 reportsMay 2026: 435 reportsJun 2026: 551 reportsJul 2026: 179 reportsAug 2026: 64 reportsSep 2026: 16 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262,323
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 21, 2025Class IOpen, Classified

    Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization.

    Z-2343-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Pumping Stopped17,192
  • Battery Problem5,612
  • Charging Problem5,420
  • Communication or Transmission Problem4,054
  • Incorrect Measurement3,911
  • Obstruction of Flow2,750
  • Adverse Event Without Identified Device or Use Problem2,455
  • Failure to Sense2,146

Patient problems

  • No Clinical Signs, Symptoms or Conditions45,722
  • Hyperglycemia4,017
  • Hypoglycemia1,269
  • Elevated Ketones/ Diabetic Ketoacidosis243
  • Loss of consciousness93
  • Insufficient Information80
  • Fall34
  • Vomiting24

Recent adverse event reports

The 25 most recent of 51,050 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922230 ↗
  • DeathSep 24, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that a customer had passed away on (b)(6) 2026. The cause was reported as ¿unknown¿ but reporter suspected low blood sugar. It was not known whether pump or related supplies contributed or whether customer was wearing pump at time of death. A review of pump data will be performed, and the results will be submitted in a supplemental report.

    Patient problems: Hypoglycemia

    MAUDE report 50883127 ↗
  • InjurySep 17, 2026Pumping Stopped

    It was reported that a cartridge alarm occurred during insulin delivery. The customer experienced an elevated blood glucose (BG) level displayed as \"High\"; although a specific BG value was not provided. Customer delivered a correction bolus via the pump to address the BG. Customer will continue to use the current pump for insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50812509 ↗
  • MalfunctionSep 17, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812377 ↗
  • MalfunctionSep 17, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50810180 ↗
  • InjurySep 16, 2026Pumping Stopped; Charging Problem; Material Twisted/Bent

    It was reported that pump housing and USB port pins were damaged, causing charging issues and pump shutdown. Customer¿s blood glucose level reached 600 mg/dL due to pump shutdown. Customer gave correction injection using multiple daily injections, did not require emergency help, and continued using current pump with backup insulin method if needed; Technical Support arranged shipment of replacement pump with updated software, confirmed shipping details, instructed customer to place pump in storage mode and return pump within 10 days, and advised customer to re-enter pump settings and reconnect continuous glucose monitor on replacement pump.

    Patient problems: Hyperglycemia

    MAUDE report 50795588 ↗
  • InjurySep 11, 2026Obstruction of Flow

    It was reported that customer experienced recurring occlusion alarms over several months and requested pump replacement after multiple troubleshooting attempts. Customer¿s blood glucose level reached range displayed as ¿High,¿ with specific value unknown. Customer gave correction insulin by injection to address blood glucose level. The customer continued to use the pump for insulin therapy and has an alternate method of insulin therapy available if necessary. Technical Support performed clinical troubleshooting, recommended pump replacement, arranged for new pump to be shipped, and instructed customer on how to place pump in storage mode, return pump, and contact support if occlusion alarms continued.

    Patient problems: Hyperglycemia

    MAUDE report 50740460 ↗
  • InjurySep 11, 2026Pumping Stopped

    It was reported that pump experienced a malfunction alarm, with no notable events occurring beforehand. Customer¿s blood glucose (BG) level was displayed as \"High\"; however, specific value was not provided. Elevated BG was address by a correction bolus via the pump and did not require assistance from a third party. Customer continued to use pump temporarily and planned to use alternate insulin method if needed, and Technical Support decided to replace pump under warranty, confirmed shipping details, and sent replacement pump. Customer was instructed to place problematic pump in storage mode, return it within 10 days using provided shipping materials, and then enter pump settings and reconnect continuous glucose monitor on replacement pump, which customer acknowledged.

    Patient problems: Hyperglycemia

    MAUDE report 50740442 ↗
  • MalfunctionSep 4, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50619048 ↗
  • InjurySep 3, 2026Unexpected Shutdown

    It was reported that pump battery charge dropped quickly, which led to unexpected pump shutdown. Customer¿s blood glucose level was reported as high and exceeded 500 mg/dL, with the specific value unknown. Customer delivered a correction bolus using pump, did not require outside medical assistance, and was educated by Technical Support on battery best practices, with instruction to monitor system and call back if issue persisted.

    Patient problems: Hyperglycemia

    MAUDE report 50596008 ↗
  • MalfunctionSep 3, 2026Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50590239 ↗
  • InjurySep 2, 2026Failure to Charge

    It was reported that the customer was unable to charge the pump using the Tandem-provided USB charging cable. The customer¿s blood glucose (BG) level was displayed as ¿High¿; however, a specific value was not provided. A new charging cable was sent to the customer.

    Patient problems: Hyperglycemia

    MAUDE report 50575568 ↗
  • MalfunctionSep 2, 2026False Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50564145 ↗
  • InjurySep 1, 2026Pumping Stopped

    It was reported that pump displayed malfunction alarm code Malf 12. The customer¿s blood glucose (BG) level was displayed as ¿High¿; however, a specific value was not provided. Customer gave correction insulin injection by pen or syringe and did not need help from emergency services or other healthcare providers. Technical Support walked customer through pump reset process, confirmed malfunction alarm did not recur after reset, advised customer to continue using pump, and instructed customer to call back if malfunction alarm appeared again.

    Patient problems: Hyperglycemia

    MAUDE report 50542238 ↗
  • InjurySep 1, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the customer went to the Emergency Room and was subsequently hospitalized in the Intensive Care Unit with a blood glucose (BG) level of 700 mg/dL on (b)(6) 2026. Cause was not known. Customer was treated with intravenous fluids of saline and insulin to address the elevated BG. Customer was discharged on (b)(6) 2026 with the issue resolved and no permanent damage.

    Patient problems: Hyperglycemia

    MAUDE report 50541893 ↗
  • InjurySep 1, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the customer went to the Emergency Room with a blood glucose (BG) level of 397 mg/dL with ketones of an unspecified size on (b)(6) 2026. Cause was not known. Customer was treated with intravenous fluids of saline and Tylenol to address the elevated BG. Customer was discharged on (b)(6) 2026 with the issue resolved and no permanent damage. Tandem Technical support performed a full pump system check and verified that pump was operating appropriately

    Patient problems: Hyperglycemia

    MAUDE report 50540168 ↗
  • MalfunctionAug 28, 2026Break; Charging Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50509788 ↗
  • InjuryAug 28, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the Control-IQ algorithm suspended basal delivery in response to the continuous glucose monitor (CGM) sensor reading; however, the CGM sensor reading was inaccurate. Reportedly, the CGM blood glucose (BG) reading was 47 mg/dL, and the contact was unable to provide a meter BG reading at the time of the event. As a result of the basal suspension, customer's blood glucose (BG) level rose to 350 mg/dL. Customer went to the emergency room (ER) and was treated with intravenous fluids of saline and insulin and released the same day, (b)(6) 2023 with the issue resolved and no permanent damage. System check with Tandem Technical Support did not identify any additional issues with the pump or related supplies. No additional patient or event information was available.

    Patient problems: Hyperglycemia

    MAUDE report 50507122 ↗
  • MalfunctionAug 26, 2026Failure to Power Up

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50459463 ↗
  • MalfunctionAug 25, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50438367 ↗
  • MalfunctionAug 24, 2026Failure to Charge

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50419746 ↗
  • MalfunctionAug 24, 2026False Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50408249 ↗
  • MalfunctionAug 21, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50379472 ↗
  • MalfunctionAug 21, 2026Failure to Charge

    It was reported that t:slim X2 pump battery would not charge, and charging connection was not recognized despite using working wall outlets, different wall adapters, and different charging cables, with no visible damage to USB port and no power source alerts in pump history. There was no adverse impact to the customer¿s blood glucose level. Customer planned to use injected insulin as backup therapy, and Technical Support arranged for pump replacement and discussed an out-of-warranty loaner pump option.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50369773 ↗
  • MalfunctionAug 21, 2026Premature Discharge of Battery

    It was reported that the pump battery was depleting quickly. There was no reported adverse impact to the customer. The customer declined further follow up from Tandem Technical Support. No additional information was provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50361116 ↗

FDA clearance and classification

The FDA regulates product code QFG (alternate controller enabled insulin infusion pump) under 21 CFR 880.5730 as a Class II device. 21 CFR 880.5730 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: An alternate controller enabled insulin infusion pump (ACE insulin pump) is a device intended for the infusion of insulin into a patient. The ACE insulin pump may include basal and bolus drug delivery at set or variable rates. ACE insulin pumps are designed to reliably and securely communicate with external devices, such as automated insulin dosing systems, to allow insulin delivery commands to be received, executed, and confirmed. ACE insulin pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of insulin delivery.

Common questions about the t:slim X2 Insulin Pump with Control-IQ 1009643

Has the t:slim X2 Insulin Pump with Control-IQ 1009643 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2343-2025 (Class I), began on Jul 21, 2025: Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the t:slim X2 Insulin Pump with Control-IQ 1009643?

The FDA's MAUDE database lists 51,050 adverse event reports for this device, received between Aug 25, 2022 and Sep 28, 2026. In those reports, the most common specific device problems are “Pumping Stopped” (17,192), “Battery Problem” (5,612) and “Charging Problem” (5,420); the most common patient problems are “Hyperglycemia” (4,017), “Hypoglycemia” (1,269) and “Elevated Ketones/ Diabetic Ketoacidosis” (243). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the t:slim X2 Insulin Pump with Control-IQ 1009643?

7 reports are classified as deaths, 5,335 as injuries and 45,708 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the t:slim X2 Insulin Pump with Control-IQ 1009643 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the t:slim X2 Insulin Pump with Control-IQ 1009643, and is it still sold?

The FDA's device database lists Tandem Diabetes Care, Inc. as the company responsible for it (brand t:slim X2 Insulin Pump with Control-IQ, model 1009643). Its record shows it as in commercial distribution.