t:slim X2 Insulin Pump with Control-IQ 1005011Recalls & reported problems

Tandem Diabetes Care, Inc. · t:slim X2 Insulin Pump with Control-IQ, 7.x · Class 2

All models of t:slim X2 Insulin Pump (45) · t:slim X2 Insulin Pump with Control-IQ (11)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This is a Tandem t:slim X2 insulin pump with Control-IQ. Reports mostly describe pumps that will not charge, batteries that drain quickly, cartridge loading errors and unresponsive screens. There are 35,602 reports, including 3 deaths and 3,220 injuries. One recall, started in May 2022, is open over memory faults, an inaccurate battery display and a touchscreen that stays on. Reports dropped to 100 in the last 12 months, from 984 before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
35,598
Deaths reported
3
Injuries reported
3,220
Recalls naming it
1

Summary

The FDA's MAUDE database contains 35,598 adverse event reports that list this exact device (DI 00850006613205), received between Jan 23, 2020 and Jul 8, 2026 — classified as 3 deaths, 3,220 injuries, 32,375 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (7,570 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 285 reportsJan 2024: 369 reportsFeb 2024: 512 reportsMar 2024: 810 reportsApr 2024: 904 reportsMay 2024: 782 reportsJun 2024: 688 reportsJul 2024: 817 reportsAug 2024: 554 reportsSep 2024: 355 reportsOct 2024: 296 reportsNov 2024: 160 reportsDec 2024: 144 reportsJan 2025: 167 reportsFeb 2025: 65 reportsMar 2025: 89 reportsApr 2025: 77 reportsMay 2025: 55 reportsJun 2025: 34 reportsJul 2025: 71 reportsAug 2025: 69 reportsSep 2025: 36 reportsOct 2025: 17 reportsNov 2025: 22 reportsDec 2025: 16 reportsJan 2026: 18 reportsFeb 2026: 5 reportsMar 2026: 10 reportsApr 2026: 8 reportsMay 2026: 6 reportsJun 2026: 9 reportsJul 2026: 2 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports202420252026904
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 24, 2022Class IIOpen, Classified

    Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.

    Z-1811-2022 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Pumping Stopped7,570
  • Charging Problem4,347
  • Incorrect Measurement4,333
  • Battery Problem3,530
  • Obstruction of Flow2,773
  • Failure to Sense2,392
  • Adverse Event Without Identified Device or Use Problem2,151
  • Filling Problem1,525

Patient problems

  • No Clinical Signs, Symptoms or Conditions20,665
  • No Consequences Or Impact To Patient10,057
  • No Known Impact Or Consequence To Patient1,752
  • Hyperglycemia1,633
  • Hypoglycemia1,418
  • Elevated Ketones/ Diabetic Ketoacidosis264
  • Insufficient Information86
  • Loss of consciousness30

Recent adverse event reports

The 25 most recent of 35,598 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 8, 2026Failure to Charge

    IT WAS REPORTED THAT THE PUMP BATTERY COULD NOT BE CHARGED. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS PROVIDED AND THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25753214 ↗
  • MalfunctionJul 7, 2026Premature Discharge of Battery

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS PROVIDED AND THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25742747 ↗
  • MalfunctionJun 26, 2026Premature Discharge of Battery

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS PROVIDED AND THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25641397 ↗
  • MalfunctionJun 24, 2026Key or Button Unresponsive/not Working

    IT WAS REPORTED THAT THE CUSTOMER WOULD RECEIVED "BUTTON ERRORS" ON THE PUMP, REQUIRING A PUMP RESET IN ORDER TO FUNCTION AGAIN. NO INFORMATION IF CUSTOMER WAS REFERRING TO THE WAKE BUTTON ON THE PUMP OR THE BUTTONS ON THE PUMP TOUCHSCREEN. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TO TROUBLESHOOT THE ISSUE WITH TANDEM TECHNICAL SUPPORT, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25611979 ↗
  • MalfunctionJun 17, 2026Key or Button Unresponsive/not Working

    IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS UNRESPONSIVE. CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25538168 ↗
  • MalfunctionJun 17, 2026Failure to Charge

    IT WAS REPORTED THAT THE PUMP BATTERY COULD NOT BE CHARGED. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE. REPORTEDLY, THE PUMP SUCCESSFULLY BEGAN CHARGING AFTER LEAVING THE PUMP PLUGGED INTO THE POWER SOURCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25534814 ↗
  • MalfunctionJun 15, 2026Communication or Transmission Problem

    IT WAS REPORTED THAT THE CUSTOMER RECEIVED A CONTINUOUS GLUCOSE MONITOR ERROR 42. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. THE CUSTOMER WAS INSTRUCTED TO RESET THE PUMP TO RESOLVE THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25511760 ↗
  • MalfunctionJun 15, 2026False Alarm

    IT WAS REPORTED THAT AN ALTITUDE ALARM OCCURRED. REPORTEDLY, THE PUMP WAS NOT OUTSIDE THE LABELED ALTITUDE OPERATING RANGE. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. REPORTEDLY, THE CARTRIDGE WAS RELOADED, AND INSULIN DELIVERY WAS RESUMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25508835 ↗
  • MalfunctionJun 15, 2026Pumping Stopped

    IT WAS REPORTED THAT THE PUMP EXPERIENCED A MALFUNCTION. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DID NOT HAVE AN ALTERNATE METHOD OF INSULIN THERAPY AVAILABLE. REPORTEDLY, THE CUSTOMER WOULD REACH OUT TO THEIR HCP TO OBTAIN A BACKUP METHOD OF INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25505211 ↗
  • MalfunctionJun 10, 2026False Alarm

    IT WAS REPORTED THAT AN ALTITUDE ALARM OCCURRED. REPORTEDLY, THE PUMP WAS NOT OUTSIDE THE LABELED ALTITUDE OPERATING RANGE. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE. REPORTEDLY A NEW CARTRIDGE WAS LOADED, AND INSULIN DELIVERY WAS RESUMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25464795 ↗
  • MalfunctionJun 9, 2026Air/Gas in Device

    IT WAS REPORTED THAT AIR BUBBLES WERE OBSERVED WITHIN THE CARTRIDGE. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. CUSTOMER PERFORMED A SUPPLY CHANGE TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25446240 ↗
  • MalfunctionMay 28, 2026Break; Charging Problem

    IT WAS REPORTED THAT CWAS REPORTED THAT THE USB PORT WAS DAMAGED AS THE TANDEM CHARGING CABLE BECAME STUCK INSIDE THE USB PORT AND AS A RESULT THE CUSTOMER WAS UNABLE TO CHARGE THE PUMP AND THE PUMP SUBSEQUENTLY SHUTDOWN. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25314655 ↗
  • MalfunctionMay 27, 2026Break; Melted

    IT WAS REPORTED THAT PUMP HOUSING HAD PLASTIC MELTING AROUND CARTRIDGE AREA AND THERE WAS A "CRACK ON PUMP". THERE WAS NO REPORTED IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS RECEIVED REGARDING ACTIONS TAKEN BY CUSTOMER OR TECHNICAL SUPPORT IN RESPONSE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25306872 ↗
  • InjuryMay 15, 2026Excess Flow or Over-Infusion

    IT WAS REPORTED THAT THE CUSTOMER WENT TO THE EMERGENCY ROOM WITH AN UNKNOWN LOW BLOOD GLUCOSE LEVEL. THE CUSTOMER ALLEGED THAT THE PUMP DID NOT ACCURATELY ADJUST THE AMOUNT OF INSULIN DELIVERED. CUSTOMER DECLINED TO TROUBLESHOOT THE ISSUE WITH TANDEM TECHNICAL SUPPORT; THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.

    Patient problems: Hypoglycemia

    MAUDE report 25182982 ↗
  • MalfunctionMay 12, 2026Display or Visual Feedback Problem

    IT WAS REPORTED THAT SPIDERWEB CRACKS WERE OBSERVED BEHIND THE TOUCHSCREEN. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE VALUE. REPORTEDLY, THE CUSTOMER CONTACTED THE HEALTHCARE PROVIDER TO OBTAIN ALTERNATE INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25150196 ↗
  • MalfunctionMay 4, 2026Display or Visual Feedback Problem

    IT WAS REPORTED THAT PUMP WOULD NOT TURN ON AND CALLER STATED ISSUE HAD ALREADY BEEN REPORTED. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. CALLER DECLINED TROUBLESHOOTING, STATED THAT PROCESS FOR REPLACEMENT PUMP THROUGH CCS MEDICAL HAD ALREADY BEGUN, AND NO FURTHER ACTION WAS NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25068897 ↗
  • MalfunctionMay 1, 2026Key or Button Unresponsive/not Working

    IT WAS REPORTED THAT THE WAKE BUTTON ON THE PUMP HAD TO BE PRESSED MULTIPLE TIMES IN ORDER TO TURN THE PUMP SCREEN ON. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TO TROUBLESHOOT THE ISSUE WITH TANDEM TECHNICAL SUPPORT; THEREFORE, NO ADDITIONAL INFORMATION WAS OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25059014 ↗
  • MalfunctionApr 28, 2026Fracture

    IT WAS REPORTED THAT THE PUMP TOUCHSCREEN WAS CRACKED AND UNRESPONSIVE AND UNREADABLE. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25017553 ↗
  • MalfunctionApr 21, 2026Pumping Stopped

    IT WAS REPORTED THAT PUMP EXPERIENCED MALFUNCTION ALARM WITH CODE DISPLAYED AND NO NOTABLE EVENTS OCCURRED BEFOREHAND. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. CUSTOMER CONTACTED TECHNICAL SUPPORT, RECEIVED INSTRUCTIONS TO RESET PUMP, SUCCESSFULLY CLEARED MALFUNCTION WITHOUT RECURRENCE, WAS ADVISED TO CONTINUE USING PUMP, AND WAS INSTRUCTED TO CALL BACK IF MALFUNCTION RETURNED, WHICH CUSTOMER ACKNOWLEDGED AND UNDERSTOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24941857 ↗
  • MalfunctionApr 20, 2026Failure to Charge; Pumping Stopped

    IT WAS REPORTED THAT THE CUSTOMER RECEIVED A POWER SOURCE ALERT AFTER PLUGGING THE PUMP INTO A WALL OUTLET RESULTING IN THE PUMP SHUTTING DOWN. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. REPORTEDLY, THE ALERT CLEARED, AND THE PUMP SUCCESSFULLY BEGAN CHARGING AFTER LEAVING THE PUMP PLUGGED INTO THE POWER SOURCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24927990 ↗
  • MalfunctionApr 16, 2026Pumping Stopped

    IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. DURING TROUBLESHOOTING WITH TECHNICAL SUPPORT, THE MALFUNCTION ALARM WAS CLEARED, AND INSULIN DELIVERY WAS RESUMED SUCCESSFULLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24904346 ↗
  • MalfunctionApr 16, 2026Key or Button Unresponsive/not Working

    IT WAS REPORTED THAT THE WAKE BUTTON WAS NOT WORKING. CUSTOMER PRESSED THE WAKE BUTTON MULTIPLE TIMES AND THE WAKE BUTTON PERFORMED AS INTENDED. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24899728 ↗
  • MalfunctionApr 11, 2026Failure to Charge

    IT WAS REPORTED THAT THE PUMP BATTERY COULD NOT BE CHARGED. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. REPORTEDLY, THE PUMP SUCCESSFULLY BEGAN CHARGING AFTER LEAVING THE PUMP PLUGGED INTO THE POWER SOURCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24845899 ↗
  • MalfunctionApr 9, 2026Pumping Stopped

    IT WAS REPORTED THAT PUMP EXPERIENCED MALFUNCTION WITH MALFUNCTION ALARM DISPLAYED AND NO NOTABLE EVENTS OCCURRED BEFOREHAND. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. CUSTOMER CONTACTED TECHNICAL SUPPORT, RECEIVED GUIDANCE TO RESET PUMP, SUCCESSFULLY CLEARED MALFUNCTION WITHOUT RECURRENCE, AND WAS ADVISED TO CONTINUE USING PUMP AND CALL BACK IF PROBLEM RETURNED, WHICH CUSTOMER ACKNOWLEDGED AND UNDERSTOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24830296 ↗
  • MalfunctionApr 6, 2026Communication or Transmission Problem

    IT WAS REPORTED THAT CUSTOMER EXPERIENCED CGM ERROR AND NEEDED HELP TO START SENSOR SESSION. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. CUSTOMER CONTACTED SUPPORT, WAS GUIDED THROUGH COMPLETE PUMP RESET, DID NOT SEE ERROR AGAIN AFTER RESET, AND SUCCESSFULLY STARTED SENSOR SESSION, WITH NO ADDITIONAL ACTIONS REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791737 ↗

FDA clearance and classification

The FDA regulates product code QFG (alternate controller enabled insulin infusion pump) under 21 CFR 880.5730 as a Class II device. 21 CFR 880.5730 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: An alternate controller enabled insulin infusion pump (ACE insulin pump) is a device intended for the infusion of insulin into a patient. The ACE insulin pump may include basal and bolus drug delivery at set or variable rates. ACE insulin pumps are designed to reliably and securely communicate with external devices, such as automated insulin dosing systems, to allow insulin delivery commands to be received, executed, and confirmed. ACE insulin pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of insulin delivery.

Common questions about the t:slim X2 Insulin Pump with Control-IQ 1005011

Has the t:slim X2 Insulin Pump with Control-IQ 1005011 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1811-2022 (Class II), began on May 24, 2022: Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the t:slim X2 Insulin Pump with Control-IQ 1005011?

The FDA's MAUDE database lists 35,598 adverse event reports for this device, received between Jan 23, 2020 and Jul 8, 2026. In those reports, the most common specific device problems are “Pumping Stopped” (7,570), “Charging Problem” (4,347) and “Incorrect Measurement” (4,333); the most common patient problems are “Hyperglycemia” (1,633), “Hypoglycemia” (1,418) and “Elevated Ketones/ Diabetic Ketoacidosis” (264). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the t:slim X2 Insulin Pump with Control-IQ 1005011?

3 reports are classified as deaths, 3,220 as injuries and 32,375 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the t:slim X2 Insulin Pump with Control-IQ 1005011 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the t:slim X2 Insulin Pump with Control-IQ 1005011, and is it still sold?

The FDA's device database lists Tandem Diabetes Care, Inc. as the company responsible for it (brand t:slim X2 Insulin Pump with Control-IQ, model 1005011). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.