t:slim X2 Insulin Pump with Control-IQ 1006408Recalls & reported problems

Tandem Diabetes Care, Inc. · t:slim X2 Insulin Pump with Control-IQ, 7.4.x, Refurbished Replacement · Class 2

All models of t:slim X2 Insulin Pump (45) · t:slim X2 Insulin Pump with Control-IQ (11)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This is a refurbished Tandem t:slim X2 insulin pump. Its reports mostly describe pump stoppages, charging trouble, incorrect measurements and battery problems. There are 92,553 reports, including 12 deaths and 8,640 injuries. Two recalls are open: one covers speakers that may fail and stop insulin delivery, the other covers memory, battery display and touchscreen problems. Reports dropped to 2,608 in the last 12 months, from 9,437 before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
92,505
Deaths reported
12
Injuries reported
8,644
Recalls naming it
2

Summary

The FDA's MAUDE database contains 92,505 adverse event reports that list this exact device (DI 00850006613793), received between Aug 19, 2020 and Sep 28, 2026 — classified as 12 deaths, 8,644 injuries, 83,849 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (27,426 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1,832 reportsJan 2024: 1,930 reportsFeb 2024: 1,598 reportsMar 2024: 1,984 reportsApr 2024: 1,970 reportsMay 2024: 1,711 reportsJun 2024: 1,537 reportsJul 2024: 1,900 reportsAug 2024: 1,559 reportsSep 2024: 1,507 reportsOct 2024: 1,611 reportsNov 2024: 1,120 reportsDec 2024: 1,001 reportsJan 2025: 1,363 reportsFeb 2025: 649 reportsMar 2025: 830 reportsApr 2025: 598 reportsMay 2025: 730 reportsJun 2025: 543 reportsJul 2025: 1,000 reportsAug 2025: 741 reportsSep 2025: 453 reportsOct 2025: 409 reportsNov 2025: 407 reportsDec 2025: 460 reportsJan 2026: 316 reportsFeb 2026: 283 reportsMar 2026: 252 reportsApr 2026: 262 reportsMay 2026: 207 reportsJun 2026: 264 reportsJul 2026: 82 reportsAug 2026: 23 reportsSep 2026: 4 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261,984
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 21, 2025Class IOpen, Classified

    Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization.

    Z-2343-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  2. May 24, 2022Class IIOpen, Classified

    Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.

    Z-1811-2022 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Pumping Stopped27,426
  • Charging Problem9,979
  • Incorrect Measurement9,500
  • Battery Problem8,934
  • Failure to Sense7,206
  • Obstruction of Flow6,139
  • Adverse Event Without Identified Device or Use Problem4,766
  • Communication or Transmission Problem4,092

Patient problems

  • No Clinical Signs, Symptoms or Conditions82,788
  • Hyperglycemia5,702
  • Hypoglycemia2,775
  • No Consequences Or Impact To Patient878
  • Elevated Ketones/ Diabetic Ketoacidosis569
  • Insufficient Information225
  • No Known Impact Or Consequence To Patient179
  • Loss of consciousness120

Recent adverse event reports

The 25 most recent of 92,505 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Fracture

    It was reported that the pump touch screen was cracked, unresponsive, and unreadable. The customer's blood glucose (BG) level was \"High\", although a specific value was not given. The customer administered a manual injection to address their BG. The customer reverted to an alternate method insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50926142 ↗
  • InjurySep 11, 2026Pumping Stopped; Charging Problem

    It was reported that the customer was unable to charge the pump with the Tandem-provided charging cable and wall adapter, leading to the pump shutting down. Customer¿s blood glucose (BG) level was displayed as \"High\", although a specific value was not given. The customer addressed their BG with a correction injection. The customer was able to successfully charge the pump with an alternate charging cable and wall adapter.

    Patient problems: Hyperglycemia

    MAUDE report 50740170 ↗
  • InjurySep 10, 2026Pumping Stopped

    It was reported that the pump experienced a malfunction after the customer changed the cartridge. The customer¿s blood glucose (BG) level displayed as \"High\"; although a specific value was not provided. A correction injection (MDI) was delivered to address the elevated BG level, and there was no need for third-party assistance. The pump was replaced to resolve the issue, and the customer will use MDI as a backup therapy until the replacement arrives.

    Patient problems: Hyperglycemia

    MAUDE report 50719746 ↗
  • InjurySep 1, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the customer went to the hospital with a blood glucose (BG) level of 900 mg/dL. Reportedly, due to the elevated BG, the customer lost consciousness. Cause was not known. Multiple attempts were made by Tandem Technical Support to follow-up with the customer on the reported issue, but no response was received.

    Patient problems: Hyperglycemia; Loss of consciousness

    MAUDE report 50542984 ↗
  • MalfunctionAug 31, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50518444 ↗
  • MalfunctionAug 24, 2026Failure to Charge

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50417188 ↗
  • MalfunctionAug 21, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50381882 ↗
  • MalfunctionAug 21, 2026Failure to Charge

    It was reported that t:slim X2 pump battery would not charge and a power source alert appeared in pump history. There was no reported impact to the customer¿s blood glucose level. Customer was instructed to leave pump connected to working wall outlet with Tandem charging cable and adapter for at least 30 minutes, and technical support attempted follow-up by phone and email before closing case, with no additional information received regarding the outcome of the event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50369283 ↗
  • MalfunctionAug 20, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50357642 ↗
  • MalfunctionAug 20, 2026Incorrect Measurement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50357630 ↗
  • MalfunctionAug 20, 2026Mechanical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50356525 ↗
  • MalfunctionAug 20, 2026Insufficient Device Problem Information

    It was reported that customer was unable to load new cartridge, so customer used injections. There was no reported impact to the customer¿s blood glucose level. Customer declined further follow-up with Technical Support, used injections as alternate insulin therapy, and was in process of pump renewal through insurance.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50354212 ↗
  • MalfunctionAug 20, 2026Premature Discharge of Battery; Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50350343 ↗
  • MalfunctionAug 20, 2026Key or Button Unresponsive/not Working

    It was reported that the pump touch screen was unresponsive. There was no adverse impact to customer¿s blood glucose level. The customer reverted to an alternate method insulin therapy.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50348551 ↗
  • MalfunctionAug 20, 2026Date/Time-Related Software Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50347387 ↗
  • MalfunctionAug 20, 2026Melted

    It was reported that charging s cord used with pump melted and they switched to an alternate cable. There was no reported impact to the customer¿s blood glucose level. Customer declined troubleshooting, and no additional information was received regarding the outcome of the event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50346076 ↗
  • MalfunctionAug 20, 2026Pumping Stopped

    It was reported that pump displayed malfunction alarm code Malf 0, and caller contacted Technical Support for assistance. There was no adverse impact to the customer¿s blood glucose level. Caller worked with Technical Support to reset pump, malfunction did not recur after reset, and caller was advised to continue using pump and to call back if any malfunction alarm occurred again.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50345726 ↗
  • MalfunctionAug 19, 2026Obstruction of Flow

    It was reported that multiple occlusion alarms occurred. There was no reported impact to the customer's blood glucose level. The customer performed a supply change to resolve the issue and resumed insulin therapy successfully.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50330167 ↗
  • MalfunctionAug 19, 2026Charging Problem

    It was reported that the customer was unable to charge the pump using the Tandem-provided USB charging cable. There was no reported adverse impact to the customer's blood glucose level. Reportedly, the customer was able to successfully charge the pump using a secondary USB cable.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50328485 ↗
  • MalfunctionAug 19, 2026Key or Button Unresponsive/not Working

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50326072 ↗
  • InjuryAug 19, 2026Use of Device Problem

    It was reported that the pump battery was depleting quickly, resulting in the pump shutting off. The customer experienced an elevated blood glucose (BG) level displayed as "High"; although a specific BG value was not provided. Customer administered a correction injection , Phenergan suppository and oral tablets to address the BG. Subsequently, the customer went to the emergency room (ER) and was hospitalized in the Intensive Care Unit (ICU) on (b)(6) 2026 due to the issue not resolving. The customer experienced a blood glucose level of over 600 mg/dL and high ketones, identified as life threatening by a healthcare provider. The customer received intravenous fluids of saline and insulin to address the BG. The customer was released from the hospital 3 days later on (b)(6) 2026 with the issue resolved and no permanent damage. Troubleshooting with Tandem Technical Support indicated that pump battery was depleting normally based on settings and customer usage. The customer continued to use the pump for insulin therapy.

    Patient problems: Elevated Ketones/ Diabetic Ketoacidosis; Hyperglycemia

    MAUDE report 50325070 ↗
  • MalfunctionAug 19, 2026Priming Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50324167 ↗
  • MalfunctionAug 18, 2026Pumping Stopped

    It was reported that a malfunction alarm occurred on pump, with momentary power loss to vibrator motor and displayed code that caller had not previously reset within prior 24 hours. There was no adverse impact to the customer¿s blood glucose level. Technical Support provided pump reset information, assisted caller with resetting pump, confirmed malfunction did not recur after reset, and advised caller to continue using pump and call back if any further malfunction alarms occurred; caller acknowledged and understood instructions.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50305155 ↗
  • MalfunctionAug 17, 2026Incorrect Measurement

    It was reported that the insulin gauge was inaccurate. There was no reported impact to the customer's blood glucose level. No additional information was provided and the customer declined troubleshooting with Tandem Technical Support.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50278320 ↗
  • MalfunctionAug 7, 2026Image Display Error/Artifact

    It was reported that pump screen showed dead pixels. There was no reported impact to the customer. Customer declined follow up with Technical Support and no further actions or outcomes were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50181710 ↗

FDA clearance and classification

The FDA regulates product code QFG (alternate controller enabled insulin infusion pump) under 21 CFR 880.5730 as a Class II device. 21 CFR 880.5730 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: An alternate controller enabled insulin infusion pump (ACE insulin pump) is a device intended for the infusion of insulin into a patient. The ACE insulin pump may include basal and bolus drug delivery at set or variable rates. ACE insulin pumps are designed to reliably and securely communicate with external devices, such as automated insulin dosing systems, to allow insulin delivery commands to be received, executed, and confirmed. ACE insulin pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of insulin delivery.

Common questions about the t:slim X2 Insulin Pump with Control-IQ 1006408

Has the t:slim X2 Insulin Pump with Control-IQ 1006408 been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-2343-2025 (Class I), began on Jul 21, 2025: Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the t:slim X2 Insulin Pump with Control-IQ 1006408?

The FDA's MAUDE database lists 92,505 adverse event reports for this device, received between Aug 19, 2020 and Sep 28, 2026. In those reports, the most common specific device problems are “Pumping Stopped” (27,426), “Charging Problem” (9,979) and “Incorrect Measurement” (9,500); the most common patient problems are “Hyperglycemia” (5,702), “Hypoglycemia” (2,775) and “Elevated Ketones/ Diabetic Ketoacidosis” (569). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the t:slim X2 Insulin Pump with Control-IQ 1006408?

12 reports are classified as deaths, 8,644 as injuries and 83,849 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the t:slim X2 Insulin Pump with Control-IQ 1006408 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the t:slim X2 Insulin Pump with Control-IQ 1006408, and is it still sold?

The FDA's device database lists Tandem Diabetes Care, Inc. as the company responsible for it (brand t:slim X2 Insulin Pump with Control-IQ, model 1006408). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Alternate Controller Enabled Insulin Infusion Pump devices

All Alternate Controller Enabled Insulin Infusion Pump devices →