t:slim X2 Insulin Pump with Control-IQ 1009642Recalls & reported problems
Tandem Diabetes Care, Inc. · t:slim X2 Insulin Pump with Control-IQ, 7.6.x, Replacement · Class 2
All models of t:slim X2 Insulin Pump (45) · t:slim X2 Insulin Pump with Control-IQ (11)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Tandem t:slim X2 insulin pump with Control-IQ is named in 10,029 adverse event reports, including 1 death and 986 injuries. The reports mostly describe pumps that stop delivering insulin, along with battery, charging and cartridge problems. One open recall, started in July 2025, concerns speakers that can make the pump stop infusing insulin without audible alerts. Reports in the last 12 months number 868, down from 1,980 the year before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 10,005
- Deaths reported
- 1
- Injuries reported
- 988
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 10,005 adverse event reports that list this exact device (DI 00389152000084), received between Aug 18, 2022 and Sep 10, 2026 — classified as 1 death, 988 injuries, 9,016 malfunctions.
The most frequently reported specific device problem is “Pumping Stopped” (3,197 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jul 21, 2025Class IOpen, Classified
Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization.
Z-2343-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Pumping Stopped3,197
- Battery Problem1,237
- Charging Problem1,010
- Communication or Transmission Problem962
- Incorrect Measurement834
- Obstruction of Flow519
- Adverse Event Without Identified Device or Use Problem508
- Failure to Sense418
Patient problems
- No Clinical Signs, Symptoms or Conditions9,015
- Hyperglycemia700
- Hypoglycemia282
- Elevated Ketones/ Diabetic Ketoacidosis49
- Insufficient Information13
- Fall11
- Loss of consciousness11
- Bruise/Contusion5
Recent adverse event reports
The 25 most recent of 10,005 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 10, 2026Pumping Stopped
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50717906 ↗InjurySep 2, 2026Use of Device Problem
It was reported that temperature alarm occurred and pump was described as malfunctioning after exposure to extreme temperatures. Customer¿s blood glucose level was 598 mg/dL due to the issue. Customer gave correction insulin by injection using multiple daily injections, did not need emergency help, and Technical Support explained that pump was working as designed and advised avoiding activities that could expose pump to extreme temperatures, which customer understood and acknowledged.
Patient problems: Hyperglycemia
MAUDE report 50568257 ↗MalfunctionAug 31, 2026Mechanical Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50528425 ↗InjuryAug 28, 2026Pumping Stopped
It was reported that pump displayed malfunction alarm after exposure to extreme heat above 100 F, and pump showed malfunction code that could not be reset. Customer¿s blood glucose level rose to 598 mg/dL due to this issue. Customer gave correction insulin injection by pen or syringe, did not require emergency help, and planned to use multiple daily injections as backup therapy. Technical Support arranged replacement pump under warranty, confirmed shipping details, provided instructions for storage mode and pump return, and advised customer to enter pump settings and reconnect continuous glucose monitor on replacement pump.
Patient problems: Hyperglycemia
MAUDE report 50507841 ↗MalfunctionAug 20, 2026Display or Visual Feedback Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50355447 ↗MalfunctionAug 20, 2026Audible Prompt/Feedback Problem
It was reported that pump speaker did not work at times, customer could not hear alarms or alerts, and battery life was diminished. There was no reported impact to the customer¿s blood glucose level. Customer declined assistance from Technical Support, and no actions were taken other than documentation of event, with no additional information received regarding the outcome of the event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50351540 ↗MalfunctionAug 19, 2026Premature Discharge of Battery
It was reported that the pump battery was depleting quickly. There was no reported adverse impact to the customer¿s blood glucose level. The customer declined follow-up troubleshooting with Tandem Technical Support, therefore no additional information could be obtained.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50325143 ↗MalfunctionAug 18, 2026Pumping Stopped
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50305920 ↗MalfunctionAug 17, 2026Loose or Intermittent Connection; Charging Problem
It was reported that charging port on pump appeared worn out and customer had to hold charging cord in place to charge device, causing concern that pump might stop charging. There was no reported impact to the customer¿s blood glucose level. Customer declined troubleshooting and no additional information was obtained, and CTS took no further action and noted that pump in active use at time of report.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50279370 ↗MalfunctionAug 14, 2026False Alarm
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50260712 ↗InjuryAug 6, 2026Obstruction of Flow
It was reported that ongoing intermittent occlusion alarms occurred. Customer changed pump supplies to address the issue. Customer¿s blood glucose (BG) level was displayed as "High"; however, specific value was not provided. Elevated BG was addressed by a correction bolus via the pump.
Patient problems: Hyperglycemia
MAUDE report 50166857 ↗MalfunctionJul 31, 2026Break; Charging Problem
It was reported that the pump's USB port was damaged resulting in difficulties charging the pump. There was no adverse impact to the customer's blood glucose level. Reportedly, the customer continued using the pump for insulin therapy until the replacement pump arrived.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50104897 ↗InjuryJul 31, 2026Priming Problem
It was reported that insulin drips were not observed to be exiting the infusion set tubing during the load fill process. The customer¿s blood glucose (BG) level was displayed as ¿High¿; however, a specific value was not provided. BG level was addressed with a correction bolus via the pump. Reportedly, the cartridge was changed to address the issue and insulin delivery was resumed.
Patient problems: Hyperglycemia
MAUDE report 50104127 ↗MalfunctionJul 30, 2026Charging Problem; Material Twisted/Bent
It was reported that the pump's USB port was bent resulting in difficulties charging the pump. There was no adverse impact to the customer's blood glucose level. The customer reverted to an alternate method of insulin therapy.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50090077 ↗MalfunctionJul 30, 2026False Alarm
It was reported that the pump was not exposed to extreme temperatures, but temperature alarms occurred while the pump was charging. There was no adverse impact to the customer's blood glucose level. Reportedly, customer continued using pump for insulin therapy.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50083979 ↗InjuryJul 30, 2026Use of Device Problem
It was reported that the pump battery was depleting quickly resulting in the pump shutting off. The customer¿s blood glucose (BG) level was displayed as ¿High¿; however, a specific value was not provided. A correction injection was administered to address the BG. Troubleshooting with Tandem Technical Support indicated that pump battery was depleting normally based on settings and customer usage. The customer continued to use the pump for insulin therapy.
Patient problems: Hyperglycemia
MAUDE report 50083626 ↗MalfunctionJul 27, 2026Mechanical Problem
It was reported that a cartridge alarm occurred during the load sequence. Customer will continue to use the current pump for insulin therapy. There was no adverse impact to the customer¿s blood glucose level.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50047785 ↗MalfunctionJul 27, 2026Pumping Stopped
It was reported that the pump experienced a malfunction. There was no reported adverse impact to the customer¿s blood glucose level. The pump was replaced to resolve the issue, and the customer will use multiple daily injections (MDI) as a backup therapy until the replacement arrives.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50047464 ↗MalfunctionJul 27, 2026Mechanical Problem
It was reported that a cartridge alarm occurred during the load sequence. There was no adverse impact to the customer's blood glucose level. The customer reverted to an alternate method of insulin therapy.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50038264 ↗MalfunctionJul 9, 2026Key or Button Unresponsive/not Working
IT WAS REPORTED THAT POWER AND DISPLAY BUTTON WAS NOT RESPONSIVE. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TRANSFER TO TECHNICAL SUPPORT DURING CALL, AND NO ADDITIONAL INFORMATION WAS RECEIVED REGARDING FURTHER ACTIONS OR OUTCOME OF EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25771702 ↗MalfunctionJul 9, 2026Obstruction of Flow
IT WAS REPORTED THAT ONGOING INTERMITTENT OCCLUSION ALARMS OCCURRED. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TO TROUBLESHOOT THE ISSUE WITH TANDEM TECHNICAL SUPPORT, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25766324 ↗MalfunctionJul 9, 2026Failure to Charge
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A POWER SOURCE ALERT AFTER PLUGGING THE PUMP INTO A WALL OUTLET. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. REPORTEDLY, THE ALERT CLEARED, AND THE PUMP SUCCESSFULLY BEGAN CHARGING AFTER LEAVING THE PUMP PLUGGED INTO THE POWER SOURCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25766084 ↗MalfunctionJul 9, 2026Key or Button Unresponsive/not Working
IT WAS REPORTED THAT THE WAKE BUTTON WAS STICKING. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS RECEIVED REGARDING ACTIONS TAKEN BY THE CUSTOMER OR TECHNICAL SUPPORT IN RESPONSE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25765157 ↗MalfunctionJul 9, 2026Pumping Stopped
IT WAS REPORTED THAT PUMP EXPERIENCED MALFUNCTION ALARM, AND CALLER CONFIRMED NO NOTABLE EVENTS OCCURRED BEFOREHAND. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. TECHNICAL SUPPORT ASSISTED CALLER WITH RESETTING PUMP AND RECONNECTING CONTINUOUS GLUCOSE MONITOR, ADVISED THAT PUMP COULD CONTINUE TO BE USED, AND INSTRUCTED CALLER TO CALL BACK IF MALFUNCTION RECURRED, WHICH CALLER ACKNOWLEDGED AND UNDERSTOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25765113 ↗MalfunctionJul 9, 2026Pumping Stopped
IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. DURING TROUBLESHOOTING WITH TECHNICAL SUPPORT, THE MALFUNCTION ALARM WAS CLEARED, AND INSULIN DELIVERY WAS RESUMED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25764746 ↗
FDA clearance and classification
The FDA regulates product code QFG (alternate controller enabled insulin infusion pump) under 21 CFR 880.5730 as a Class II device. 21 CFR 880.5730 ↗
Devices of this type normally need 510(k) clearance before sale.
FDA definition of this device type: An alternate controller enabled insulin infusion pump (ACE insulin pump) is a device intended for the infusion of insulin into a patient. The ACE insulin pump may include basal and bolus drug delivery at set or variable rates. ACE insulin pumps are designed to reliably and securely communicate with external devices, such as automated insulin dosing systems, to allow insulin delivery commands to be received, executed, and confirmed. ACE insulin pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of insulin delivery.
Common questions about the t:slim X2 Insulin Pump with Control-IQ 1009642
Has the t:slim X2 Insulin Pump with Control-IQ 1009642 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2343-2025 (Class I), began on Jul 21, 2025: Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the t:slim X2 Insulin Pump with Control-IQ 1009642?
The FDA's MAUDE database lists 10,005 adverse event reports for this device, received between Aug 18, 2022 and Sep 10, 2026. In those reports, the most common specific device problems are “Pumping Stopped” (3,197), “Battery Problem” (1,237) and “Charging Problem” (1,010); the most common patient problems are “Hyperglycemia” (700), “Hypoglycemia” (282) and “Elevated Ketones/ Diabetic Ketoacidosis” (49). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the t:slim X2 Insulin Pump with Control-IQ 1009642?
1 report is classified as a death, 988 as injuries and 9,016 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the t:slim X2 Insulin Pump with Control-IQ 1009642 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the t:slim X2 Insulin Pump with Control-IQ 1009642, and is it still sold?
The FDA's device database lists Tandem Diabetes Care, Inc. as the company responsible for it (brand t:slim X2 Insulin Pump with Control-IQ, model 1009642). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Alternate Controller Enabled Insulin Infusion Pump devices
- t:slim X2 Insulin Pump with Control-IQ 1006402Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump2 recalls18 deaths159,198 reports
- t:slim X2 Insulin Pump with Control-IQ 1006408Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump2 recalls12 deaths92,505 reports
- t:slim X2 Insulin Pump with Control-IQ 1009637Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall4 deaths83,767 reports
- Omnipod 5 Pod POD-BLE-H1-520Insulet Corporation · Alternate Controller Enabled Insulin Infusion Pump1 recall74 deaths71,491 reports
- Omnipod 5 Pod POD-OMNI-I1-6720Insulet Corporation · Alternate Controller Enabled Insulin Infusion Pump2 recalls11 deaths60,668 reports
- t:slim X2 Insulin Pump with Control-IQ 1009643Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall7 deaths51,050 reports
- t:slim X2 Insulin Pump with Control-IQ Technology 1013198Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall8 deaths47,498 reports
- t:slim X2 Insulin Pump with Control-IQ 1006404Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall18 deaths43,714 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page