t:slim X2 Insulin Pump with Control-IQ 1009641Recalls & reported problems

Tandem Diabetes Care, Inc. · t:slim X2 Insulin Pump with Control-IQ, 7.6.x, Medicare · Class 2

All models of t:slim X2 Insulin Pump (45) · t:slim X2 Insulin Pump with Control-IQ (11)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Tandem t:slim X2 insulin pump has 25,914 adverse event reports in the FDA database, including 8 deaths and 3,836 injuries. Recent reports often describe blockage alarms, pumps that stop working, and high blood sugar readings. There is 1 open recall, started in July 2025, because some pumps have speakers that may fail, silencing alerts and stopping insulin delivery. Reports in the last 12 months (3,515) are fewer than in the 12 months before (6,327).

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
25,808
Deaths reported
8
Injuries reported
3,836
Recalls naming it
1

Summary

The FDA's MAUDE database contains 25,808 adverse event reports that list this exact device (DI 00389152000091), received between Sep 7, 2022 and Sep 30, 2026 — classified as 8 deaths, 3,836 injuries, 21,964 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (8,350 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 872 reportsJan 2024: 931 reportsFeb 2024: 794 reportsMar 2024: 818 reportsApr 2024: 875 reportsMay 2024: 689 reportsJun 2024: 735 reportsJul 2024: 840 reportsAug 2024: 776 reportsSep 2024: 843 reportsOct 2024: 880 reportsNov 2024: 603 reportsDec 2024: 528 reportsJan 2025: 844 reportsFeb 2025: 450 reportsMar 2025: 556 reportsApr 2025: 395 reportsMay 2025: 509 reportsJun 2025: 405 reportsJul 2025: 706 reportsAug 2025: 627 reportsSep 2025: 346 reportsOct 2025: 358 reportsNov 2025: 427 reportsDec 2025: 531 reportsJan 2026: 460 reportsFeb 2026: 425 reportsMar 2026: 434 reportsApr 2026: 401 reportsMay 2026: 292 reportsJun 2026: 267 reportsJul 2026: 93 reportsAug 2026: 55 reportsSep 2026: 24 reportsOct 2026: 0 reportsNov 2026: 0 reports202420252026931
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 21, 2025Class IOpen, Classified

    Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization.

    Z-2343-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Pumping Stopped8,350
  • Incorrect Measurement3,169
  • Charging Problem2,411
  • Adverse Event Without Identified Device or Use Problem2,357
  • Communication or Transmission Problem1,839
  • Filling Problem1,797
  • Obstruction of Flow1,357
  • Battery Problem1,341

Patient problems

  • No Clinical Signs, Symptoms or Conditions21,952
  • Hyperglycemia2,564
  • Hypoglycemia1,242
  • Elevated Ketones/ Diabetic Ketoacidosis197
  • Loss of consciousness77
  • Insufficient Information72
  • Fall32
  • Vomiting22

Recent adverse event reports

The 25 most recent of 25,808 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Fluid/Blood Leak

    It was reported that multiple cartridges were leaking from the pigtail. Customer¿s blood glucose (BG) level was displayed as \"High\"; however, specific value was not provided. Elevated BG was addressed by administering a correction bolus. Customer loaded a new cartridge to address the issue.

    Patient problems: Hyperglycemia

    MAUDE report 50955106 ↗
  • MalfunctionSep 28, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923902 ↗
  • InjurySep 25, 2026Use of Device Problem

    It was reported that the customer received an incomplete bolus alert as they had not completed a bolus request. The customer experienced an elevated blood glucose (BG) level displayed as \"High\"; however, specific value was not provided. The customer administered a manual injection to address the BG and continued to use the pump for insulin therapy. Tandem technical support educated the customer on the meaning of an incomplete bolus alert.

    Patient problems: Hyperglycemia

    MAUDE report 50902518 ↗
  • MalfunctionSep 22, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50862879 ↗
  • MalfunctionSep 22, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50849389 ↗
  • InjurySep 21, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that customer experienced a past low blood glucose event while using pump with automated insulin adjustment feature active, with lowest blood glucose level of 51 mg/dL and unknown cause. Customer consumed carbohydrates to treat low blood glucose and did not require help from another person. Customer was able to upload pump data, return of cartridge was arranged, and Technical Support advised customer to consult healthcare provider to discuss factors that might be affecting blood glucose levels.

    Patient problems: Hypoglycemia

    MAUDE report 50847698 ↗
  • InjurySep 17, 2026Obstruction of Flow

    It was reported that customer experienced occlusion alarms while using infusion set, cartridge, and Humalog insulin. Customer¿s blood glucose level was reported displayed as ¿High,¿ with specific value unknown. Customer delivered correction bolus via pump, changed infusion set and cartridge, and then resumed insulin, did not require help from another healthcare provider, and Tandem Technical Support closed case with no further assistance needed.

    Patient problems: Hyperglycemia

    MAUDE report 50810975 ↗
  • MalfunctionSep 14, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50764056 ↗
  • MalfunctionSep 14, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50764034 ↗
  • InjurySep 14, 2026Obstruction of Flow

    It was reported that customer experienced recurring occlusion alarms. The customer¿s blood glucose (BG) level was displayed as ¿High¿; however, a specific value was not provided. As a result the customer went to emergency room, was admitted to intensive care on (b)(6) 2026. Customer received intravenous fluids of saline and insulin. Customer was released from hospital on (b)(6) 2026 with the issue resolved and no permanent damage. Reportedly, the customer resumed insulin. Technical Support reviewed data, provided guidance to rotate infusion sites, use longer cannula, change insulin cartridge every three days, and contact support if issues continued, and case was closed.

    Patient problems: Hyperglycemia

    MAUDE report 50763159 ↗
  • InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the customer went to the Emergency Room with a blood glucose of 370 mg/dL and ketones identified as dangerous by a healthcare provider. The customer was treated with intravenous fluids of saline and insulin to address BG. The customer was released (b)(6) 2026 with the issue resolved and no permanent damage. A system check was completed and no issues with found with the pump.

    Patient problems: Elevated Ketones/ Diabetic Ketoacidosis; Hyperglycemia

    MAUDE report 50758685 ↗
  • InjurySep 11, 2026Use of Device Problem

    It was reported that the customer allowed the pump battery to fully deplete due to not charging the battery and the pump subsequently shut off. Customer¿s blood glucose (BG) level was displayed as \"High\", although a specific value was not given. The customer addressed their BG with a manual injection. Reportedly, the customer continued using the pump for insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50741526 ↗
  • MalfunctionSep 11, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50737018 ↗
  • InjurySep 8, 2026Obstruction of Flow

    It was reported occlusion alarms occurred every three days for the past couple months. Eventually resulting in the customer's blood glucose level being displayed as \"High\"; a specific value was not provided on 08/16/26. Elevated blood glucose was addressed with a correction bolus via the pump. The customer changed supplies and resumed insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50679159 ↗
  • MalfunctionSep 4, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50620572 ↗
  • InjurySep 4, 2026Use of Device Problem

    It was reported that the pump battery was depleting quickly resulting in the pump shutting off. The customer¿s blood glucose (BG) was 601 mg/dL. BG level was corrected by administering manual injection. Troubleshooting with Tandem Technical Support indicated that pump battery was depleting normally based on settings and customer usage. The customer continued to use the pump for insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50620241 ↗
  • MalfunctionSep 3, 2026Failure to Charge

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50596508 ↗
  • InjurySep 3, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the customer's BG value displayed as  "Low"; however, specific value was not provided. Cause of low BG was unknown. Customer treated low by consuming glucose tablets and candy, had already spoken with healthcare provider, and Technical Support planned to review pump data due to concern related to automated adjustment feature, with no additional information received regarding outcome of event.

    Patient problems: Hypoglycemia

    MAUDE report 50594595 ↗
  • MalfunctionSep 3, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50591405 ↗
  • MalfunctionSep 3, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50590882 ↗
  • InjurySep 2, 2026Fracture

    It was reported that the pump touchscreen was cracked and unresponsive and unreadable. Customer¿s blood glucose (BG) level was displayed as "High", although a specific value was not given. The customer addressed their BG with a correction injection. Reportedly, the customer contacted the healthcare provider to obtain an alternate method of insulin therapy until the replacement pump arrived.

    Patient problems: Hyperglycemia

    MAUDE report 50572633 ↗
  • MalfunctionSep 1, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50544361 ↗
  • InjurySep 1, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that customer experienced very low blood sugar, lost consciousness, went to ER, and was subsequently hospitalized from (b)(6) 2026 to (b)(6) 2026. Lowest recorded blood sugar related to incident was 30 mg/dL, cause was not known, and there was no injury or permanent damage identified by healthcare provider. Customer received IV treatment and fluids in hospital, although the customer could not recall the type of fluids received, and treatment resolved issue. Customer was disconnected from pump during tubing fill step, bolus history did not show more insulin than intended, and personal profile settings were entered correctly. Cartridge was not available for return and customer was unable to upload pump data. Tandem Technical Support recommended customer consult healthcare provider to discuss factors affecting blood sugar levels, and customer acknowledged and understood this guidance.

    Patient problems: Hypoglycemia; Loss of consciousness

    MAUDE report 50543012 ↗
  • InjurySep 1, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the customer went to the Emergency Room and was subsequently hospitalized with a blood glucose (BG) level of 764 mg/dL on (b)(6) 2026. Cause was not known. Customer was treated with intravenous fluids of saline and insulin as well as magnesium and potassium to address the elevated BG. Customer was discharged on (b)(6) 2026 with the issue resolved and no permanent damage.

    Patient problems: Hyperglycemia

    MAUDE report 50541079 ↗
  • MalfunctionAug 30, 2026Pumping Stopped; Charging Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50510308 ↗

FDA clearance and classification

The FDA regulates product code QFG (alternate controller enabled insulin infusion pump) under 21 CFR 880.5730 as a Class II device. 21 CFR 880.5730 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: An alternate controller enabled insulin infusion pump (ACE insulin pump) is a device intended for the infusion of insulin into a patient. The ACE insulin pump may include basal and bolus drug delivery at set or variable rates. ACE insulin pumps are designed to reliably and securely communicate with external devices, such as automated insulin dosing systems, to allow insulin delivery commands to be received, executed, and confirmed. ACE insulin pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of insulin delivery.

Common questions about the t:slim X2 Insulin Pump with Control-IQ 1009641

Has the t:slim X2 Insulin Pump with Control-IQ 1009641 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2343-2025 (Class I), began on Jul 21, 2025: Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the t:slim X2 Insulin Pump with Control-IQ 1009641?

The FDA's MAUDE database lists 25,808 adverse event reports for this device, received between Sep 7, 2022 and Sep 30, 2026. In those reports, the most common specific device problems are “Pumping Stopped” (8,350), “Incorrect Measurement” (3,169) and “Charging Problem” (2,411); the most common patient problems are “Hyperglycemia” (2,564), “Hypoglycemia” (1,242) and “Elevated Ketones/ Diabetic Ketoacidosis” (197). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the t:slim X2 Insulin Pump with Control-IQ 1009641?

8 reports are classified as deaths, 3,836 as injuries and 21,964 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the t:slim X2 Insulin Pump with Control-IQ 1009641 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the t:slim X2 Insulin Pump with Control-IQ 1009641, and is it still sold?

The FDA's device database lists Tandem Diabetes Care, Inc. as the company responsible for it (brand t:slim X2 Insulin Pump with Control-IQ, model 1009641). Its record shows it as in commercial distribution.