Comprehensive Shoulder System recallZ-1543-2020 · Biomet, Inc. · initiated Feb 17, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A 15mm Shoulder Mini Humeral Stem was misidentified as a 9mm Shoulder Mini Humeral Stem.
FDA root cause category: Under Investigation by firm
What the company is doing
URGENT MEDICAL DEVICE FIELD SAFETY NOTICE-REMOVAL letter dated February 17, 2020 was sent via email to consignees. The letter identified affected product, problem and the action to be taken. Customers were advised to take the following actions: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the field safety notice for hospitals and ensure documentation. 4.Complete Attachment 1 Certification of Acknowledgement Form and send to [email protected] form must be returned even if you do not have affected products available to return in your facility. 5.Retain a copy of the acknowledgement form with your field action records in the event of a compliance audit of your facilitys documentation. 6. If you have further questions or concerns after reviewing this notice, please contact your Zimmer Biomet representative. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA: Med Watch Reporting: Adverse reactions or quality problems experienced with the use of this product may be reported to the FDAs Med Watch Adverse Event Reporting program either online, by mail, or by fax. Online: www.fda.gov/medwatch/report.htm Mail: Use postage paid, pre-addressed form FDA 3500, available at: www.fda.gov/MedWatch/getforms.htm Fax: 1-800-FDA-0178 Under 21 CFR 803, manufacturers are required to report any serious injuries where a product has contributed or may have contributed to the event. Please keep Zimmer Biomet informed of any adverse events associated with this product or any other Zimmer Biomet product by emailing [email protected].
Recall details
- Recall number
- Z-1543-2020
- Status
- Terminated
- Initiated by the company
- Feb 17, 2020
- Posted by the FDA
- Mar 3, 2020
- Recalling firm
- Biomet, Inc.
- Quantity
- 14 units
- Product code
- MBF
- FDA submissions
- K060692
- Distribution
- International distribution in the countries of Canada, Japan, Netherlands, South Korea.
Product description
Comprehensive Shoulder System Mini Humeral Stem 9mm,, Model No. 113629 - Product Usage: The intended use of the comprehensive humeral stem mini is to act as the humeral component of the comprehensive total shoulder system.
Code information (models, lots, serial numbers)
Lot No. 659260, UDI Number: (1)00880304462649(17)290814(10)659260
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- COMPREHENSIVE SHOULDER SYSTEM 113629Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED1 recall2 reports
Product lines
Other devices under the same FDA submission
45 devices cleared under K060692 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- COMPREHENSIVE SHOULDER SYSTEM 113633Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED2 recalls4 reports
- COMPREHENSIVE SHOULDER SYSTEM 113631Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall2 reports
- COMPREHENSIVE SHOULDER SYSTEM 113618Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall2 reports
- Comprehensive® Shoulder System 113627Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall2 reports
- COMPREHENSIVE SHOULDER SYSTEM 113635Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED1 recall1 report
- COMPREHENSIVE SHOULDER SYSTEM 113611Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED1 recall1 report
- COMPREHENSIVE SHOULDER SYSTEM 113617Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED1 recall1 report
- Comprehensive® Shoulder System 113607Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED1 recall1 report
- COMPREHENSIVE SHOULDER SYSTEM 113632Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED1 recall0 reports
- COMPREHENSIVE SHOULDER SYSTEM 113614Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall0 reports
- Comprehensive® Shoulder System 113605Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall0 reports
- Comprehensive® Shoulder System 113606Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page