t:slim X2 Insulin Pump with Control-IQ 1005013Recalls & reported problems
Tandem Diabetes Care, Inc. · t:slim X2 Insulin Pump with Control-IQ, 7.x, Replacement · Class 2
All models of t:slim X2 Insulin Pump (45) · t:slim X2 Insulin Pump with Control-IQ (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Tandem t:slim X2 insulin pump with Control-IQ delivers insulin. Reports often describe the pump stopping, charging problems, incorrect readings and battery drain. The device appears in 5,097 reports, with 442 injuries and no deaths. It has 1 open recall, started in May 2022, for memory, battery display and touchscreen faults that a software update can address. Reports dropped to 1 in the last 12 months from 27 before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 5,097
- Deaths reported
- 0
- Injuries reported
- 442
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5,097 adverse event reports that list this exact device (DI 00850006613212), received between Feb 13, 2020 and Dec 4, 2025 — classified as 442 injuries, 4,655 malfunctions.
The most frequently reported specific device problem is “Pumping Stopped” (967 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 24, 2022Class IIOpen, Classified
Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.
Z-1811-2022 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Pumping Stopped967
- Charging Problem644
- Incorrect Measurement576
- Battery Problem485
- Obstruction of Flow450
- Failure to Sense392
- Adverse Event Without Identified Device or Use Problem283
- False Alarm244
Patient problems
- No Consequences Or Impact To Patient2,747
- No Clinical Signs, Symptoms or Conditions1,330
- No Known Impact Or Consequence To Patient603
- Hypoglycemia228
- Hyperglycemia195
- Elevated Ketones/ Diabetic Ketoacidosis36
- Loss of consciousness5
- Insufficient Information4
Recent adverse event reports
The 25 most recent of 5,097 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 4, 2025Failure to Charge; Pumping Stopped; Failure to Charge; Pumping Stopped
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A POWER SOURCE ALERT AFTER PLUGGING THE PUMP INTO A WALL OUTLET, CAUSING THE PUMP TO SHUT DOWN. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. REPORTEDLY, THE ALERT CLEARED, AND THE PUMP SUCCESSFULLY BEGAN CHARGING AFTER LEAVING THE PUMP PLUGGED INTO THE POWER SOURCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23716354 ↗MalfunctionOct 15, 2025Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TO TROUBLESHOOT THE ISSUE WITH TANDEM TECHNICAL SUPPORT; THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23291957 ↗InjuryAug 18, 2025Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THAT INTERMITTENT OCCLUSION ALARMS OCCURRED. REPORTEDLY, THE CUSTOMER HAD BEEN USING THE SAME CARTRIDGE AND INFUSION SET FOR OVER 2 DAYS USING HUMALOG INSULIN. TANDEM CUSTOMER TECHNICAL SUPPORT INFORMED CUSTOMER THAT HUMALOG INSULIN IS INDICATED FOR USE WITH TANDEM SUPPLIES FOR 2 DAYS. CUSTOMER CHANGED PUMP SUPPLIES TO ADDRESS THE ISSUE. THE CUSTOMER¿S BLOOD GLUCOSE (BG) LEVEL WAS DISPLAYED AS ¿HIGH¿; HOWEVER, A SPECIFIC VALUE WAS NOT PROVIDED. CUSTOMER ADMINISTERED A MANUAL INJECTION TO ADDRESS THEIR BG LEVEL.
Patient problems: Hyperglycemia
MAUDE report 22829334 ↗MalfunctionAug 8, 2025Date/Time-Related Software Problem; Date/Time-Related Software Problem
IT WAS REPORTED THAT THE PUMP DATE WAS INCORRECT, CAUSE WAS UNKNOWN. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. MULTIPLE ATTEMPTS WERE MADE BY TANDEM TECHNICAL SUPPORT TO FOLLOW-UP WITH THE CUSTOMER ON THE REPORTED ISSUE, BUT NO RESPONSE WAS RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22750361 ↗InjuryAug 8, 2025Date/Time-Related Software Problem; Date/Time-Related Software Problem
IT WAS REPORTED THAT THE PUMP TIME WAS INCORRECT, CAUSE WAS UNKNOWN. THE CUSTOMER'S BLOOD GLUCOSE (BG) LEVEL WAS "HIGH", ALTHOUGH A SPECIFIC VALUE WAS NOT GIVEN. THE CUSTOMER ADMINISTERED A MANUAL INJECTIONS TO ADDRESS THEIR BG. TIME ISSUE ALSO CAUSED THE CUSTOMER'S BLOOD GLUCOSE (BG) LEVEL TO BE "LOW", ALTHOUGH A SPECIFIC VALUE WAS NOT GIVEN. THE CUSTOMER CONSUMED CARBOHYDRATES. DURING TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, THE CUSTOMER CORRECTED THE TIME AND RESUMED INSULIN THERAPY.
Patient problems: Hyperglycemia; Hypoglycemia
MAUDE report 22750349 ↗InjuryAug 7, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
IT WAS REPORTED THAT THE WAKE BUTTON WAS STICKING. THE CUSTOMER¿S BLOOD GLUCOSE (BG) LEVEL WAS DISPLAYED AS ¿LOW¿; HOWEVER, A SPECIFIC VALUE WAS NOT PROVIDED. CUSTOMER SUSPENDED PUMP USE TO ADDRESS BG LEVEL. REPLACEMENT PUMP WAS SENT TO CUSTOMER TO RESOLVE THE ISSUE.
Patient problems: Hypoglycemia
MAUDE report 22738235 ↗MalfunctionJul 25, 2025Failure to Charge; Failure to Charge
IT WAS REPORTED THAT THE PUMP BATTERY COULD NOT BE CHARGED. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22610393 ↗MalfunctionJul 22, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS UNRESPONSIVE. CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22571714 ↗MalfunctionJul 8, 2025Break; Break
A CARTRIDGE CHANGE ERROR WAS REPORTED FOR THE PUMP, AND THE O-RING IN THE CARTRIDGE WAS FOUND TO BE OUT OF PLACE AND DAMAGED. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. TECHNICAL SUPPORT WORKED WITH THE CUSTOMER TO TROUBLESHOOT THE ISSUE, INCLUDING INSPECTING AND CLEANING THE PUMP'S PNEUMATIC TAP AREA, ATTACHING THE CARTRIDGE CORRECTLY, AND CONFIRMING IT WAS FULLY SNAPPED INTO PLACE. DESPITE THESE EFFORTS, THE CUSTOMER WAS UNABLE TO SUCCESSFULLY LOAD THE CARTRIDGE ONTO THE CURRENT PUMP, BUT THEY WERE ABLE TO DO SO ON A NEW PUMP. THE CUSTOMER RECEIVED CARTRIDGE REPLACEMENTS, AND THE ISSUE WAS RESOLVED WITH INSULIN DELIVERY RESUMED, REQUIRING NO FURTHER ACTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22437555 ↗MalfunctionJul 1, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS UNRESPONSIVE. THERE WAS NO ADVERSE IMPACT TO CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD INSULIN THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22379532 ↗MalfunctionJul 1, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS UNRESPONSIVE. MULTIPLE CONTACT ATTEMPTS WERE MADE BY TANDEM TECHNICAL SUPPORT TO OBTAIN ADDITIONAL INFORMATION REGARDING THE ISSUE; HOWEVER, THE CUSTOMER DID NOT RESPOND. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22379438 ↗MalfunctionMay 29, 2025Charging Problem; Charging Problem
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A POWER SOURCE ALERT AFTER PLUGGING THE PUMP INTO A WALL OUTLET. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. REPORTEDLY, THE ALERT CLEARED, AND THE PUMP SUCCESSFULLY BEGAN CHARGING AFTER LEAVING THE PUMP PLUGGED INTO THE POWER SOURCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22106886 ↗MalfunctionMar 31, 2025Battery Problem; Battery Problem
IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE. REPORTEDLY, THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21734892 ↗MalfunctionMar 18, 2025Pumping Stopped; Pumping Stopped
IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. DURING TROUBLESHOOTING WITH TECHNICAL SUPPORT, THE MALFUNCTION ALARM WAS CLEARED, AND INSULIN DELIVERY WAS RESUMED SUCCESSFULLY. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21633416 ↗InjuryMar 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED A BLOOD GLUCOSE (BG) LEVEL THAT WAS DISPLAYED AS ¿HIGH¿; HOWEVER, A SPECIFIC VALUE WAS NOT PROVIDED. REPORTEDLY, THE CUSTOMER ALLEGED THE PUMP SOFTWARE DID NOT ACCURATELY ADJUST THE AMOUNT OF INSULIN DELIVERED RESULTING IN THE HIGH BG. CUSTOMER GAVE A MANUAL INJECTION TO ADDRESS BG LEVEL. TANDEM TECHNICAL SUPPORT (CTS) PERFORMED A SYSTEM CHECK AND PERFORMED A REVIEW OF THE PUMP DATA TO DETERMINE A POSSIBLE CAUSE. CTS DETERMINED THAT THE PUMP FUNCTIONED AS INTENDED AND THE CAUSE OF THE HIGH BG WAS UNKNOWN, CUSTOMER ACKNOWLEDGED AND UNDERSTOOD. RECOMMENDATION WAS MADE TO DISCUSS DIABETES MANAGEMENT WITH A HEALTH CARE PROFESSIONAL.
Patient problems: Hyperglycemia
MAUDE report 21504926 ↗InjuryFeb 25, 2025Pumping Stopped; Pumping Stopped
IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. DURING TROUBLESHOOTING WITH TECHNICAL SUPPORT, THE MALFUNCTION ALARM WAS CLEARED, AND INSULIN DELIVERY WAS RESUMED SUCCESSFULLY. THE CUSTOMER¿S BLOOD GLUCOSE (BG) LEVEL WAS DISPLAYED AS ¿HIGH¿; HOWEVER, A SPECIFIC VALUE WAS NOT PROVIDED. BG LEVEL WAS CORRECTED WITH A BOLUS VIA THE PUMP.
Patient problems: Hyperglycemia
MAUDE report 21459734 ↗MalfunctionFeb 13, 2025Failure to Sense; Failure to Sense
IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS UNRESPONSIVE. CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21371218 ↗MalfunctionFeb 12, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE CARTRIDGE O-RING WAS MISSING. THE CUSTOMER CONFIRMED THE O-RING WAS NOT LOCATED ON THE PNEUMATIC TAP. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. A NEW CARTRIDGE WAS LOADED TO RESOLVE THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21367509 ↗MalfunctionJan 3, 2025Failure to Sense; Failure to Sense
IT WAS REPORTED THAT THE PUMP TOUCH SCREEN IS SCRATCHED, LEADING TO SCREEN BEING UNRESPONSIVE. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS PROVIDED AND THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21066761 ↗MalfunctionJan 2, 2025Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED THAT THE WAKE BUTTON WAS NOT WORKING. CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21055150 ↗MalfunctionDec 9, 2024Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED THAT THE WAKE BUTTON WAS STICKING. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20881134 ↗MalfunctionDec 6, 2024Failure to Sense; Date/Time-Related Software Problem; Failure to Sense; Date/Time-Related Software Problem
IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS INTERMITTENTLY UNRESPONSIVE. ADDITIONALLY, IT WAS REPORTED THAT THE PUMP TIME WAS INCORRECT, CAUSE WAS UNKNOWN. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS PROVIDED AND THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20872999 ↗MalfunctionDec 6, 2024Failure to Sense; Battery Problem; Failure to Sense; Battery Problem
IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS UNRESPONSIVE AND THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO ADVERSE IMPACT TO CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20863669 ↗MalfunctionNov 25, 2024Charging Problem; Charging Problem
IT WAS REPORTED THAT THE CUSTOMER WAS UNABLE TO CHARGE THE PUMP USING THE TANDEM-PROVIDED USB CHARGING CABLE AND WALL POWER ADAPTER. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. REPORTEDLY, THE CUSTOMER WAS ABLE TO SUCCESSFULLY CHARGE THE PUMP USING A SECONDARY USB CABLE AND WALL POWER ADAPTER. THE CUSTOMER WAS SENT NEW CHARGING SUPPLIES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20772761 ↗MalfunctionNov 15, 2024Incorrect Measurement; Fitting Problem; Battery Problem; Incorrect Measurement; Fitting Problem; Battery Problem
IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY, A CARTRIDGE DID NOT FIT ONTO THE PUMP AND THE INSULIN GAUGE WAS INACCURATE. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20694147 ↗
Other Alternate Controller Enabled Insulin Infusion Pump devices
- t:slim X2 Insulin Pump with Control-IQ 1006402Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump2 recalls18 deaths159,247 reports
- t:slim X2 Insulin Pump with Control-IQ 1006408Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump2 recalls12 deaths92,526 reports
- t:slim X2 Insulin Pump with Control-IQ 1009637Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall4 deaths83,884 reports
- Omnipod 5 Pod POD-BLE-H1-520Insulet Corporation · Alternate Controller Enabled Insulin Infusion Pump1 recall75 deaths71,459 reports
- Omnipod 5 Pod POD-OMNI-I1-6720Insulet Corporation · Alternate Controller Enabled Insulin Infusion Pump2 recalls11 deaths61,198 reports
- t:slim X2 Insulin Pump with Control-IQ 1009643Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall7 deaths51,110 reports
- t:slim X2 Insulin Pump with Control-IQ Technology 1013198Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall8 deaths47,637 reports
- t:slim X2 Insulin Pump with Control-IQ 1006404Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall18 deaths43,718 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page