t:slim X2 Insulin Pump with Basal-IQ 1006373Recalls & reported problems

Tandem Diabetes Care, Inc. · t:slim X2 Insulin Pump with Basal-IQ, 6.4.x · Class 2

All models of t:slim X2 Insulin Pump (45) · t:slim X2 Insulin Pump with Basal-IQ (14)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Tandem t:slim X2 insulin pump is worn to deliver insulin. Reports mostly describe pumps that stop, batteries that cannot charge or drain quickly, incorrect measurements, and blockage alarms. There are 7,490 reports, with no deaths and 790 injuries. One open recall, begun in May 2022, covers software and display problems that could cause high or low blood sugar. Reports fell to 183 in the last 12 months from 1,074 in the 12 months before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
7,488
Deaths reported
0
Injuries reported
791
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7,488 adverse event reports that list this exact device (DI 00850006613700), received between Sep 10, 2020 and Aug 20, 2026 — classified as 791 injuries, 6,697 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (1,900 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 104 reportsDec 2023: 102 reportsJan 2024: 105 reportsFeb 2024: 87 reportsMar 2024: 142 reportsApr 2024: 147 reportsMay 2024: 115 reportsJun 2024: 120 reportsJul 2024: 138 reportsAug 2024: 150 reportsSep 2024: 143 reportsOct 2024: 156 reportsNov 2024: 116 reportsDec 2024: 104 reportsJan 2025: 153 reportsFeb 2025: 70 reportsMar 2025: 93 reportsApr 2025: 67 reportsMay 2025: 66 reportsJun 2025: 51 reportsJul 2025: 128 reportsAug 2025: 91 reportsSep 2025: 65 reportsOct 2025: 70 reportsNov 2025: 27 reportsDec 2025: 37 reportsJan 2026: 24 reportsFeb 2026: 15 reportsMar 2026: 14 reportsApr 2026: 13 reportsMay 2026: 17 reportsJun 2026: 17 reportsJul 2026: 12 reportsAug 2026: 5 reportsSep 2026: 0 reportsOct 2026: 0 reports202420252026156
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 24, 2022Class IIOpen, Classified

    Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.

    Z-1811-2022 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Pumping Stopped1,900
  • Charging Problem825
  • Incorrect Measurement783
  • Battery Problem621
  • Adverse Event Without Identified Device or Use Problem505
  • Obstruction of Flow487
  • Failure to Sense456
  • Crack439

Patient problems

  • No Clinical Signs, Symptoms or Conditions6,487
  • Hyperglycemia437
  • Hypoglycemia306
  • No Consequences Or Impact To Patient199
  • Elevated Ketones/ Diabetic Ketoacidosis59
  • Insufficient Information34
  • No Known Impact Or Consequence To Patient28
  • Loss of consciousness14

Recent adverse event reports

The 25 most recent of 7,488 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 20, 2026Key or Button Unresponsive/not Working

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50355033 ↗
  • MalfunctionAug 19, 2026Incorrect Measurement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50332352 ↗
  • MalfunctionAug 19, 2026Key or Button Unresponsive/not Working

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50325574 ↗
  • MalfunctionAug 7, 2026Incorrect Measurement

    It was reported that a minimum fill notification occurred after the user filled the cartridge with 200 units of insulin during the load sequence. A new cartridge was loaded to resolve the issue. There was no adverse impact to the customer¿s blood glucose level.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50181696 ↗
  • InjuryAug 6, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the customer experienced an elevated blood glucose (BG) level displayed as "High"; although specific value was not provided. Cause was not known. Customer performed a supply change to address the BG. Recommendation was made to consult a healthcare provider in regards to diabetes management.

    Patient problems: Hyperglycemia

    MAUDE report 50160745 ↗
  • MalfunctionJul 10, 2026Failure to Charge

    IT WAS REPORTED THAT THE PUMP BATTERY COULD NOT BE CHARGED. CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED ADVERSE IMPACT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25782505 ↗
  • MalfunctionJul 10, 2026Key or Button Unresponsive/not Working

    IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS UNRESPONSIVE. CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25781017 ↗
  • MalfunctionJul 9, 2026Key or Button Unresponsive/not Working

    IT WAS REPORTED THAT PUMP HAD UNRESPONSIVE WAKE BUTTON, AND PATIENT DID NOT WANT TO PERFORM TROUBLESHOOTING. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS RECEIVED REGARDING ANY CORRECTIVE ACTIONS TAKEN BY THE CUSTOMER OR TECHNICAL SUPPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25771675 ↗
  • MalfunctionJul 8, 2026Premature Discharge of Battery

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED ASSISTANCE FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUES. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25757008 ↗
  • MalfunctionJul 8, 2026Premature Discharge of Battery

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED ASSISTANCE FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUES. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25757007 ↗
  • MalfunctionJul 8, 2026Key or Button Unresponsive/not Working

    IT WAS REPORTED THAT THE TOUCHSCREEN "WILL NOT ALWAYS RESPOND". CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25752680 ↗
  • MalfunctionJul 8, 2026Key or Button Unresponsive/not Working

    IT WAS REPORTED THAT THE PUMP TOUCHSCREEN WAS DELAYED IN RESPONDING TO TOUCH INPUTS. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TO TROUBLESHOOT THE ISSUE WITH TANDEM TECHNICAL SUPPORT, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25751606 ↗
  • MalfunctionJul 8, 2026Loose or Intermittent Connection; Charging Problem

    IT WAS REPORTED THAT THE PUMP'S USB PORT WAS LOOSE, AND THE PUMP BATTERY INTERMITTENTLY COULD NOT BE CHARGED PROPERLY WITHOUT THE CUSTOMER MANEUVERING THE CABLE INTO THE CHARGING PORT AT A CERTAIN ANGLE. CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25750277 ↗
  • MalfunctionJul 8, 2026Pumping Stopped

    IT WAS REPORTED THAT PUMP EXPERIENCED RECURRING MONTHLY MALFUNCTIONS THAT REQUIRED PATIENT TO RESTART PUMP MULTIPLE TIMES, BUT NO SPECIFIC MALFUNCTION DETAILS OR DATES WERE PROVIDED. NO INFORMATION WAS RECEIVED REGARDING ANY IMPACT TO CUSTOMER¿S BLOOD GLUCOSE LEVEL. PATIENT DECLINED FURTHER FOLLOW UP WITH TECHNICAL SUPPORT, AND NO CORRECTIVE ACTIONS OR ADDITIONAL TROUBLESHOOTING WERE DOCUMENTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25748075 ↗
  • MalfunctionJul 6, 2026Premature Discharge of Battery

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED ADVERSE IMPACT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25716807 ↗
  • MalfunctionJul 6, 2026Charging Problem

    IT WAS REPORTED THAT CHARGING CORD REQUIRED EXTRA FORCE TO PLUG IN, AND THIS CONCERN WAS COMMUNICATED THROUGH A DOCUMENT-ONLY EMAIL. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. CUSTOMER HAD ALREADY DISCUSSED ISSUE WITH SUPPORT TEAM AND DECLINED FURTHER FOLLOW-UP FROM TECHNICAL SUPPORT, AND NO ADDITIONAL CORRECTIVE ACTIONS OR OUTCOMES WERE DOCUMENTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25716330 ↗
  • MalfunctionJul 1, 2026Key or Button Unresponsive/not Working

    IT WAS REPORTED THAT PUMPTOUCHSCREEN BEING SOMETIMES UNRESPONSIVE. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED FURTHER TROUBLESHOOTING AND FOLLOW-UP, AND CASE WAS CLOSED WITH NO ADDITIONAL ACTIONS TAKEN OR INFORMATION RECEIVED REGARDING THE OUTCOME OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25679744 ↗
  • MalfunctionJun 30, 2026Premature Discharge of Battery

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25667681 ↗
  • MalfunctionJun 29, 2026Failure to Charge

    IT WAS REPORTED THAT THE PUMP BATTERY COULD NOT BE CHARGED. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25656071 ↗
  • MalfunctionJun 26, 2026Key or Button Unresponsive/not Working

    IT WAS REPORTED THAT THE WAKE BUTTON WAS NOT WORKING. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE. NO ADDITIONAL INFORMATION WAS PROVIDED AND THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25640256 ↗
  • MalfunctionJun 26, 2026Incorrect Measurement

    IT WAS REPORTED THAT THE BATTERY GAUGE WAS INACCURATE. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS PROVIDED AND THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25637644 ↗
  • MalfunctionJun 26, 2026Premature Discharge of Battery

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED ADVERSE IMPACT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25633899 ↗
  • MalfunctionJun 26, 2026Key or Button Unresponsive/not Working

    IT WAS REPORTED THAT THE WAKE BUTTON WAS NOT WORKING. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. CUSTOMER DECLINED FURTHER TROUBLESHOOTING WITH TECHNICAL SUPPORT, ADDITIONAL INFORMATION WAS NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25631681 ↗
  • MalfunctionJun 25, 2026Premature Discharge of Battery

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TO TROUBLESHOOT WITH TANDEM TECHNICAL SUPPORT, THEREFORE NO ADDITIONAL INFORMATION COULD BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25627175 ↗
  • MalfunctionJun 17, 2026Pumping Stopped

    IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. DURING TROUBLESHOOTING WITH TECHNICAL SUPPORT, THE MALFUNCTION ALARM WAS CLEARED, AND INSULIN DELIVERY WAS RESUMED SUCCESSFULLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25535672 ↗

Other Alternate Controller Enabled Insulin Infusion Pump devices

All Alternate Controller Enabled Insulin Infusion Pump devices →