Skip to content

t:slim X2 Insulin Pump with Basal-IQ 1006377Recalls & reported problems

Tandem Diabetes Care, Inc. · t:slim X2 Insulin Pump with Basal-IQ, 6.4.x, Replacement · Class 2

All models of t:slim X2 Insulin Pump (45) · t:slim X2 Insulin Pump with Basal-IQ (14)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Tandem t:slim X2 insulin pump has 520 adverse event reports, with no deaths and 39 injuries. Many reports describe the pump stopping, incorrect readings and charging or battery problems, and most describe no effect on the patient. One open recall from May 2022 covers memory faults, wrong battery readings and unexpected switches out of automatic mode, which a software update can fix. Reports fell to 3 in the last 12 months from 20 in the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
520
Deaths reported
0
Injuries reported
39
Recalls naming it
1

Summary

The FDA's MAUDE database contains 520 adverse event reports that list this exact device (DI 00850006613724), received between Sep 9, 2020 and Jun 25, 2026 — classified as 39 injuries, 481 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (120 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 5 reportsJan 2024: 13 reportsFeb 2024: 9 reportsMar 2024: 8 reportsApr 2024: 11 reportsMay 2024: 9 reportsJun 2024: 7 reportsJul 2024: 10 reportsAug 2024: 6 reportsSep 2024: 4 reportsOct 2024: 6 reportsNov 2024: 2 reportsDec 2024: 2 reportsJan 2025: 2 reportsFeb 2025: 1 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 2 reportsJul 2025: 3 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202613
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 24, 2022Class IIOpen, Classified

    Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.

    Z-1811-2022 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Pumping Stopped120
  • Incorrect Measurement66
  • Charging Problem59
  • Obstruction of Flow48
  • Battery Problem45
  • Failure to Sense42
  • False Alarm30
  • Crack28

Patient problems

  • No Clinical Signs, Symptoms or Conditions455
  • Hyperglycemia26
  • No Consequences Or Impact To Patient21
  • Hypoglycemia14
  • No Known Impact Or Consequence To Patient5
  • Elevated Ketones/ Diabetic Ketoacidosis4
  • Insufficient Information1
  • Urinary Retention1

Recent adverse event reports

The 25 most recent of 520 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 25, 2026Break; Charging Problem

    IT WAS REPORTED THAT THE PUMP USB PORT WAS DAMAGED, AND THE PUMP BATTERY INTERMITTENTLY COULD NOT BE CHARGED PROPERLY WITHOUT THE CUSTOMER MANEUVERING THE CABLE INTO THE CHARGING PORT AT A CERTAIN ANGLE. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS PROVIDED AND THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25621992 ↗
  • MalfunctionJun 23, 2026Key or Button Unresponsive/not Working

    IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS UNRESPONSIVE. CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25600357 ↗
  • MalfunctionMay 4, 2026Date/Time-Related Software Problem

    IT WAS REPORTED THAT THE PUMP TIME WAS INCORRECT, CAUSE WAS UNKNOWN. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. DURING TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, THE CUSTOMER CORRECTED THE TIME AND RESUMED INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25074542 ↗
  • MalfunctionOct 23, 2025Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT A PUMP MALFUNCTION OCCURRED. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23375810 ↗
  • MalfunctionOct 21, 2025Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT A MALFUNCTION OCCURRED ON THE CUSTOMER'S PUMP. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. TECHNICAL SUPPORT ASSISTED THE CUSTOMER WITH RESETTING THE PUMP, AND THE CUSTOMER WAS INSTRUCTED TO CONTINUE USING THE PUMP WHILE MONITORING FOR ANY RECURRING ISSUES. THE CUSTOMER ACKNOWLEDGED THIS GUIDANCE AND AGREED TO CALL BACK IF THE MALFUNCTION RECURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23348052 ↗
  • MalfunctionSep 10, 2025Premature Discharge of Battery; Pumping Stopped; Premature Discharge of Battery; Pumping Stopped

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY RESULTING IN THE PUMP SHUTTING DOWN. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23017539 ↗
  • MalfunctionAug 27, 2025Premature Discharge of Battery; Premature Discharge of Battery

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FURTHER TROUBLESHOOTING, AND NO ADDITIONAL INFORMATION WAS PROVIDED REGARDING THE OUTCOME OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22907354 ↗
  • MalfunctionJul 22, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE CARTRIDGE WAS LEAKING FROM THE SEPTUM. THERE WAS NO REPORTED ADVERSE EFFECT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. TECHNICAL SUPPORT INITIATED A RETURN FOR THE FAULTY CARTRIDGE, WHILE THE CALLER DECLINED ADDITIONAL ASSISTANCE WITH LOADING A NEW CARTRIDGE, INDICATING THEY COULD MANAGE ON THEIR OWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22569162 ↗
  • MalfunctionJul 8, 2025Pumping Stopped; Pumping Stopped

    A MALFUNCTION WAS REPORTED ON THE PUMP, SHOWING A MALFUNCTION CODE 12. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL, AS IT DID NOT EXCEED 500 MG/DL. TECHNICAL SUPPORT ASSISTED THE CALLER IN RESETTING THE PUMP AND PROVIDED INSTRUCTIONS ON WHAT TO DO IF THE ISSUE RECURRED. THE CUSTOMER WAS ABLE TO CONTINUE USING THE PUMP AND WAS ADVISED TO CALL BACK IF THE MALFUNCTION REAPPEARED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22438386 ↗
  • MalfunctionJul 3, 2025Incorrect Measurement; Incorrect Measurement

    IT WAS REPORTED THAT MINIMUM FILL NOTIFICATIONS OCCURRED AFTER THE USER FILLED THE CARTRIDGES WITH (B)(4) UNITS OF INSULIN DURING THE LOAD SEQUENCE. A NEW CARTRIDGE WAS LOADED TO RESOLVE THE ISSUE. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22396595 ↗
  • MalfunctionJun 6, 2025Device Sensing Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE WAKE BUTTON WAS UNRESPONSIVE. PUMP DISPLAY TURNED ON WHEN PLUGGED INTO A POWER SOURCE. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE. REPORTEDLY, THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22159402 ↗
  • MalfunctionJun 4, 2025Device Sensing Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE WAKE BUTTON WAS UNRESPONSIVE. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TO TROUBLESHOOT WITH TANDEM TECHNICAL SUPPORT, THEREFORE NO ADDITIONAL INFORMATION COULD BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22145810 ↗
  • MalfunctionMay 27, 2025Filling Problem; Incorrect Measurement; Filling Problem; Incorrect Measurement

    IT WAS REPORTED THAT A MINIMUM FILL NOTIFICATION OCCURRED AFTER THE USER FILLED THE CARTRIDGE WITH 250 UNITS OF INSULIN DURING THE LOAD SEQUENCE. A NEW CARTRIDGE WAS LOADED TO RESOLVE THE ISSUE. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 409 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22092788 ↗
  • MalfunctionMay 15, 2025Incorrect Measurement; Charging Problem; Incorrect Measurement; Charging Problem

    IT WAS REPORTED THAT THE BATTERY GAUGE WAS FLUCTUATING. ADDITIONALLY, THE CUSTOMER HAD DIFFICULTY CHARGING THE PUMP BATTERY. THERE WAS NO ADVERSE IMPACT TO CUSTOMER'S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED ASSISTANCE FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUES. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22028236 ↗
  • MalfunctionMar 24, 2025Battery Problem; Charging Problem; Battery Problem; Charging Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS "NOT HOLDING A CHARGE AND WILL NOT CHARGE ABOVE 85 PERCENT." THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED ASSISTANCE FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUES. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21683550 ↗
  • MalfunctionMar 3, 2025Pumping Stopped; Battery Problem; Pumping Stopped; Battery Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY RESULTING IN THE PUMP SHUTTING OFF. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED ASSISTANCE FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21505797 ↗
  • MalfunctionFeb 19, 2025Incorrect Measurement; Battery Problem; Incorrect Measurement; Battery Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. ADDITIONALLY, IT WAS REPORTED THAT THE BATTERY GAUGE WAS FLUCTUATING. CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED ADVERSE IMPACT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21417300 ↗
  • MalfunctionJan 13, 2025Failure to Power Up; Failure to Power Up

    IT WAS REPORTED THAT THE PUMP BATTERY WAS UNRESPONSIVE. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY. THERE WAS NO ADVERSE IMPACT TO CUSTOMER'S BLOOD GLUCOSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21144277 ↗
  • MalfunctionJan 13, 2025Filling Problem; Incorrect Measurement; Filling Problem; Incorrect Measurement

    IT WAS REPORTED THAT A MINIMUM FILL NOTIFICATION OCCURRED. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, THEREFORE NO ADDITIONAL INFORMATION COULD BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21138521 ↗
  • MalfunctionDec 12, 2024Obstruction of Flow; Battery Problem; Device Sensing Problem; Obstruction of Flow; Battery Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY, AND THE WAKE BUTTON WAS NOT WORKING. ADDITIONALLY, OCCLUSION ALARMS OCCURRED. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED ASSISTANCE FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUES. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20919193 ↗
  • MalfunctionDec 11, 2024Loose or Intermittent Connection; Charging Problem; Device Sensing Problem; Loose or Intermittent Connection; Charging Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY, THE WAKE BUTTON WAS STICKING, AND THE WAKE BUTTON WAS NOT WORKING. IT WAS ALSO REPORTED THAT THE PUMP'S USB PORT WAS LOOSE, AND THE PUMP BATTERY INTERMITTENTLY COULD NOT BE CHARGED PROPERLY WITHOUT THE CUSTOMER MANEUVERING THE CABLE INTO THE CHARGING PORT AT A CERTAIN ANGLE. THERE WAS NO ADVERSE IMPACT TO CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20905120 ↗
  • MalfunctionNov 26, 2024Charging Problem; Device Sensing Problem; Charging Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE WAKE BUTTON WAS STICKING. ADDITIONALLY, IT WAS REPORTED THAT THE PUMP BATTERY COULD NOT BE CHARGED. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. MULTIPLE ATTEMPTS WERE MADE BY TANDEM TECHNICAL SUPPORT TO FOLLOW-UP WITH THE CUSTOMER ON THE REPORTED ISSUE, BUT NO RESPONSE WAS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20784940 ↗
  • MalfunctionNov 19, 2024Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20727027 ↗
  • MalfunctionOct 29, 2024Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT A CARTRIDGE ALARM OCCURRED DURING INSULIN DELIVERY. THE CUSTOMER REVERTED TO MANUAL INJECTIONS FOR INSULIN THERAPY. THERE WAS NO ADVERSE IMPACT TO CUSTOMER¿S BLOOD GLUCOSE LEVEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20561223 ↗
  • MalfunctionOct 25, 2024Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT A CARTRIDGE ALARM OCCURRED DURING INSULIN DELIVERY. CUSTOMER LOADED A NEW CARTRIDGE TO RESOLVE THE ISSUE. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20537615 ↗

FDA clearance and classification

The FDA regulates product code QFG (alternate controller enabled insulin infusion pump) under 21 CFR 880.5730 as a Class II device. 21 CFR 880.5730 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: An alternate controller enabled insulin infusion pump (ACE insulin pump) is a device intended for the infusion of insulin into a patient. The ACE insulin pump may include basal and bolus drug delivery at set or variable rates. ACE insulin pumps are designed to reliably and securely communicate with external devices, such as automated insulin dosing systems, to allow insulin delivery commands to be received, executed, and confirmed. ACE insulin pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of insulin delivery.

Common questions about the t:slim X2 Insulin Pump with Basal-IQ 1006377

Has the t:slim X2 Insulin Pump with Basal-IQ 1006377 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1811-2022 (Class II), began on May 24, 2022: Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the t:slim X2 Insulin Pump with Basal-IQ 1006377?

The FDA's MAUDE database lists 520 adverse event reports for this device, received between Sep 9, 2020 and Jun 25, 2026. In those reports, the most common specific device problems are “Pumping Stopped” (120), “Incorrect Measurement” (66) and “Charging Problem” (59); the most common patient problems are “Hyperglycemia” (26), “Hypoglycemia” (14) and “Elevated Ketones/ Diabetic Ketoacidosis” (4). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the t:slim X2 Insulin Pump with Basal-IQ 1006377?

None of the reports are classified as deaths, 39 as injuries and 481 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the t:slim X2 Insulin Pump with Basal-IQ 1006377 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the t:slim X2 Insulin Pump with Basal-IQ 1006377, and is it still sold?

The FDA's device database lists Tandem Diabetes Care, Inc. as the company responsible for it (brand t:slim X2 Insulin Pump with Basal-IQ, model 1006377). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Alternate Controller Enabled Insulin Infusion Pump devices

All Alternate Controller Enabled Insulin Infusion Pump devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page